Fmc - Portage Dialysis Center

A dialysis facility in Portage, Indiana 46368

5950 Us Highway 6, Portage, Indiana 46368. (219) 762-5731.

CMS’s own record, on Care Compare

Watch and the shortlist are kept with a free account, on every device. Your watch list. CMS’s own record is on Care Compare.

The short answer

Record last released by CMS on Oct 28, 2026. We couldn’t read Centers for Medicare & Medicaid Services’ list since Nov 2, 2026: qcor.cms.gov now refuses our reader. This record is as of then.

The last 24 months 12 deficiencies in 1 month of the last 36. Last cited Dec 2025. Most serious: Standard-level. Every month

Each figure’s source sits beside it, in its section below.

What these words mean3
Star ratings
CMS rates some kinds of place from 1 to 5 stars, more stars meaning better on CMS’s own measures for that kind: for a nursing home, its health inspections, staffing and quality measures. A rating is CMS’s summary of those measures; it doesn’t promise good care. Source: CMS, Care Compare and the Five-Star Quality Rating System.
Inspection (survey)
A visit by state inspectors, often unannounced, who check a place against the rules it must meet. For nursing homes and other places Medicare pays, CMS calls these visits surveys: a standard survey of a nursing home comes at least every 15 months, and a complaint survey whenever a complaint calls for one. Source: CMS; 42 CFR 488.308.
Deficiency, citation, violation
What inspectors write up when a place didn’t meet a rule: the rule, what they saw, and when. CMS calls it a deficiency; states may say citation or violation. It is the inspectors’ finding, not a court’s, and the place is usually asked for a plan to correct it. Source: CMS, statement of deficiencies (Form CMS-2567); each state’s licensing agency.

What did inspectors find?

The last inspection on record was on Dec 8, 2025: a standard inspection.

12 deficiencies cited in the last three years.

Source:

CMS, QCOR Dialysis Facility (ESRD) 2567 statements of deficiencies. Released Oct 28, 2026. Updated monthly.

When those deficiencies happened

12 deficiencies in 1 month of the last 36. Last cited Dec 2025. Most serious: Standard-level. The last 3 years, to Oct 30, 2026.

12 deficiencies by month over the last 3 years. Each cell is one month.
JanFebMarAprMayJunJulAugSepOctNovDec
2023 NovDec
2024 JanFebMarAprMayJunJulAugSepOctNovDec
2025 JanFebMarAprMayJunJulAugSepOctNovDec
2026 JanFebMarAprMayJunJulAugSepOct

Hover or focus a month to read it.

No visit on record Visited, nothing cited A stronger orange is more serious, on its source’s own levels, read onto four bands: from the least serious it publishes to the most
Source:

Hearth Record drew this from the dates and levels in the source’s own records.

Each deficiency, as cited

Rule titles as CMS’s inspectors cite them, set in sentence case.

Source:

CMS, QCOR Dialysis Facility (ESRD) 2567 statements of deficiencies. Released Oct 28, 2026. Updated monthly.

Dec 8, 2025: standard inspection12 deficiencies
  • Esrd Patient Orientation Training

    The inspectors’ words, 156 words

    “ESRD Patient Orientation Training CFR(s): The dialysis facility must provide appropriate orientation and training to patients, including the areas specified in paragraph (d)(1) of this section. This Standard is not met as evidenced by: 51199 Based on record review and interview, the dialysis facility failed to provide appropriate emergency preparedness emergency orientation and training to a new admission patient in 1 of 1 clinical record review of new admission (Patient 5). Findings include: A policy, dated 09/02/2025, titled “Guidelines for Emergency Preparedness,” indicated the facility staff must educate patients about evacuation procedures and emergency disconnection procedures. The clinical record for Patient 5 indicated Patient was admitted on 10/05/2025. The record failed to evidence documentation Patient was trained to emergency evacuation procedures nor emergency disconnection procedures. During an interview on 12/08/2025, beginning at 12:00 PM, the facility administrator indicated there was no documentation that emergency preparedness training or emergency disconnect training had been completed with Patient 5.”

    Rule E0040. Standard-level.

  • Compliance With Fed/State/Local Laws

    The inspectors’ words, 152 words

    “COMPLIANCE WITH FED/STATE/LOCAL LAWS CFR(s): 494.20 The facility and its staff must operate and furnish services in compliance with applicable Federal, State, and local laws and regulations pertaining to licensure and any other relevant health and safety requirements. This Standard is not met as evidenced by: 51199 Based on record review and interview, the dialysis facility failed to notify the Indiana Department of Health of an allegation of abuse, by an in-center hemodialysis patient, in 1 of 1 dialysis facility. Findings include: A review of the facility grievance log evidenced a grievance, dated 05/06/2025, in which a patient complained that they believed they were punched in the head by a staff member, on 05/05/2025. During an interview on 12/08/2025 beginning at 12:00, the FA indicated the Indiana Department of Health had not been notified of the abuse allegation. She indicated the PCT involved in the grievance resigned without notice the next day.”

    Rule V0101. Standard-level.

  • Ic - Sanitary Environment

    The inspectors’ words, 281 words

    “IC-SANITARY ENVIRONMENT CFR(s): 494.30 The dialysis facility must provide and monitor a sanitary environment to minimize the transmission of infectious agents within and between the unit and any adjacent hospital or other public areas. This Standard is not met as evidenced by: 51199 Based on record review, observation, and interview, the dialysis facility failed to ensure a sanitary environment in 2 of 2 treatment floor observation periods. Findings include: 1. A policy, dated 02/03/2025, titled “Mixing Bleach,” indicated that bleach solution containers should be covered with a lid. 2. During an observation of the treatment floor on 12/03/2025, from 11:00 AM to 12:42 PM, 1 container labeled “1:100 bleach” was left uncovered with the container lid located next to the container for the duration of the observation. A different container labeled “1:100 bleach” was left partially covered with the lid laying halfway off of the container for the duration of the observation. 3. During an observation of the treatment floor on 12/04/2025, beginning at 10:32 AM, one container labeled “1:100 bleach solution” was left uncovered with the lid laying next to the container until the end of the observation period 11:47 AM. A different container labeled “1:100 bleach” was left partially uncovered with the lid laying halfway on the container from 10:32 AM until 11:10 AM, and again from 11:21 until the end of the observation period 11:47 AM. 4. During an interview on 12/04/2025 beginning at 2:30 PM, PCT 2 indicated that bleach solutions should be covered at all times. 5. During an interview on 12/04/2025 beginning at 1:56 PM, PCT 1 indicated that bleach solutions need to remain covered at all times and discarded and remade if found uncovered.”

    Rule V0111. Standard-level.

  • Ic - Wear Gloves/Hand Hygiene

    The inspectors’ words, 433 words

    “IC-WEAR GLOVES/HAND HYGIENE CFR(s): 494.30(a)(1) Wear disposable gloves when caring for the patient or touching the patient’s equipment at the dialysis station. Staff must remove gloves and wash hands between each patient or station. This Standard is not met as evidenced by: 51199 Based on record review, observation, and interview the dialysis facility failed to perform hand hygiene and don clean gloves in 1 of 2 observations of initiation of dialysis with a central venous catheter [CVC] (large catheter into vein) (PCT 1), in 1 of 2 observations of discontinuation of dialysis and post dialysis access care for and arteriovenous fistula [AVF] (dialysis access site) or arteriovenous graft [AVG] (dialysis access site) (PCT 4), and 1 of 1 observation of preparation of the hemodialysis [HD] (a process to clean the blood in a patient whose kidney’s do not work normally) machine (PCT 2). Findings include: 1. A policy, dated 11/06/2023, titled “Hand Hygiene”, indicated that hands will be decontaminated with alcohol-based hand rub or by washing hands before performing any invasive procedures such as vascular access or after contact, immediately after removing gloves, after contact with wound dressings, or after contact with objects within the patient station. 2. During an observation, on 12/03/2025 beginning at 12:11 PM at station 12, PCT 1 failed to perform hand hygiene, remove soiled gloves, and don new gloves prior to initiation of hemodialysis on Patient 4’s CVC. 3. During an observation on 12/04/2025 beginning at 10:57 AM, at station 15, PCT 4 failed to remove soiled gloves, perform hand hygiene, and don clean gloves after disconnecting the bloodlines from the HD machine and prior to removing the needles from Patient 11. 4. During an observation on 12/03/2025 beginning at 11:44 AM, PCT 2 was preparing the HD machine by inserting the filter and tubing at station 15 when an alarm at station 16 began to sound. PCT 2 removed gloves and performed hand hygiene, then attended to the patient and alarm at station 16 (Patient 9). After PCT 2 was finished attending to Patient 9 and the sounding alarm, PCT 2 removed soiled gloves and donned new gloves without performing hand hygiene before returning to station 15 to complete the preparation of the HD machine. 5. During an interview on 12/04/2025 beginning at 1:56 PM, PCT 1 indicated that a glove change and hand hygiene should be performed prior to connecting a patient to the HD machine. 6. During an interview on 12/04/2025 beginning at 2:30 PM, PCT 2 indicated that hand hygiene and glove change should occur when switching between the dialysis station and a patient.”

    Rule V0113. Standard-level.

  • Ic - Hbv - Vaccinate Patients & Staff

    The inspectors’ words, 182 words

    “IC-HBV-VACCINATE PTS/STAFF CFR(s): 494.30(a)(1)(i) Hepatitis B Vaccination Vaccinate all susceptible patients and staff members against hepatitis B. This Standard is not met as evidenced by: 51199 Based on record review and interview, the dialysis facility failed evidence they vaccinated a susceptible patient against Hepatitis B in 1 of 1 clinical record review of Hepatitis B susceptible patients. (Patient 17) Findings include: A policy, dated 02/07/2022, titled “Patient Vaccination for Hepatits B,” indicated the Hepatitis B vaccine should be offered to all susceptible patients and that each patient should sign a vaccination consent or declination form. A clinical record review for Patient 17 evidenced an admission date of 11/08/2025 and Hepatitis testing results, dated 11/08/2025, indicating Patient 17 was susceptible to Hepatitis B. The record failed to evidence a Hepatitis B vaccine consent or declination form. During an interview, on 12/08/2025 beginning at 12:00 PM, the FA indicated the Hepatitis B vaccine was not administered, nor offered to Patient 17. The FA indicated the lab results indicated Patient 17 was susceptible; the facility failed to follow up by offering the Hepatitis B vaccine.”

    Rule V0126. Standard-level.

  • Bacterial Control - Disinfect Monthly/Disinfection Dwell

    The inspectors’ words, 213 words

    “BACT CONTROL-DISINFECT 1X/MO/DWELL CFR(s): 494.40(a) 7 Strategies for bacterial control 7.1 General: disinfect monthly/disinfection dwell Routine low-level disinfection of the pipes should be performed to control bacterial contamination of the distribution system. The frequency of disinfection will vary with the design of the system and the extent to which biofilm has already formed in existing systems, but disinfection must be performed at least monthly. A mechanism should be incorporated in the distribution system to ensure that disinfectant does not drain from pipes during the disinfection period. This Standard is not met as evidenced by: 51199 Based on record review and interview, the dialysis facility failed to ensure monitoring of weekly chemical disinfection of the reverse osmosis system in 1 of 1 dialysis facility. Findings include: 1. A review of the reverse osmosis water system logs for October 2025 indicated no chemical disinfection record entries for the month. 2. A review of the reverse osmosis water system logs for November 2025 indicated chemical disinfection record entries for 11/14/2025 and 11/19/2025. There were chemical disinfection record entries from 11/01/2025 to 11/13/2025 or 11/20/2025 to 11/30/2025. 3. During an interview on 12/03/2025 beginning at 4:10 PM, Biomed 1 indicated he was aware of missing entries for chemical disinfection and was reported in the monthly QAPI meetings.”

    Rule V0219. Standard-level.

  • Pe - Hemodialysis Patients In View During Treatments

    The inspectors’ words, 361 words

    “PE-HD PTS IN VIEW DURING TREATMENTS CFR(s): 494.60(c)(4) Patients must be in view of staff during hemodialysis treatment to ensure patient safety, (video surveillance will not meet this requirement). This Standard is not met as evidenced by: 51199 Based on record review, observation, and interview, the dialysis facility failed to ensure patients access sites were in view of staff during hemodialysis [HD] (a process to clean the blood of a person whose kidneys do not work normally) treatments to ensure patient safety in 1 of 2 observation periods of the treatment floor. (Patient 6, Patient 10, Patient 7, and Patient 4) Findings include: 1. A policy, dated 05/01/2023, titled “Patient Assessment and Monitoring,” indicated access sites should remain uncovered throughout the treatment. 2. During an observation of the dialysis treatment floor on 12/03/2025, observed the following: Patient 6 was in station 8 and had a blanket covering their HD access site from 11:05 AM until 12:18 PM, when RN 1 removed the blanket and visualized the dialysis access site. At 12:25 PM, Patient 6’s dialysis access site was again covered by a blanket until the conclusion of the treatment floor observation at 12:42 PM. Patient 10 was in station 3 and had a blanket covering their dialysis access site beginning at 11:33 AM until the conclusion of the treatment floor observation at 12:42 PM. Patient 7 was in station 5 and had a blanket covering their HD access site from 11:39 AM until the conclusion of the treatment floor observation at 12:42 PM. Patient 4 was in station 12 and had a blanket covering their HD access site beginning at 12:21 PM until the conclusion of the treatment floor observation at 12:42 PM. 3. During an interview on 12/04/2025 beginning at 2:30 PM, PCT 2 indicated that HD access must remain visible throughout treatment. 4. During an interview on 12/04/2025 beginning at 1:34 PM, RN 1 indicated that HD access must be visible at all times during dialysis. 5. During an interview on 12/04/2025 beginning at 1:56, PCT 1 indicated that HD access must remain visible at all times and if a patient refuses, then there must be documentation stating such.”

    Rule V0407. Standard-level.

  • Immunization/Medication History

    The inspectors’ words, 143 words

    “PA-IMMUNIZATION/MEDICATION HISTORY CFR(s): 494.80(a)(3) The patient’s comprehensive assessment must include, but is not limited to, the following: Immunization history, and medication history. This Standard is not met as evidenced by: 51199 Based on record review and interview, the dialysis facility failed to administer the pneumococcal vaccine to a patient who consented for administration in 1 of 1 clinical record review of patients who had not received the pneumococcal vaccine prior to admission. (Patient 2) Findings include: A clinical record review, for Patient 2, evidenced a Pneumococcal Vaccination Consent form, dated 08/08/2025, that indicated Patient 2 consented to have the pneumococcal vaccine administered. The clinical record review failed to evidence documentation that the pneumococcal vaccine had been administered by the dialysis facility. During an interview on 11/08/2025, beginning at 12:00 PM, the FA indicated that the pneumococcal vaccine was not administered to Patient 2.”

    Rule V0506. Standard-level.

  • Vascular Access - Monitor/Referrals

    The inspectors’ words, 200 words

    “POC-VASCULAR ACCESS-MONITOR/REFERRALS CFR(s): 494.90(a)(5) The interdisciplinary team must provide vascular access monitoring and appropriate, timely referrals to achieve and sustain vascular access. The hemodialysis patient must be evaluated for the appropriate vascular access type, taking into consideration co-morbid conditions, other risk factors, and whether the patient is a potential candidate for arteriovenous fistula placement. This Standard is not met as evidenced by: 51199 Based on record review, observation, and interview, the dialysis facility failed to ensure PCT 2 performed aseptic technique when accessing the arteriovenous fistula [AVF] (dialysis access) for initiation of dialysis in 1 of 2 observations of access of an AVF for initiation of dialysis. (Patient 9) Findings include: A policy, 05/05/2025, titled “Access Assessment and Cannulation,” indicated that once skin asepsis is completed, the cleaned site should not be touched prior to cannulation. During an observation, on 12/03/2025 beginning at 11:35 AM, PCT 2 accessed the AVF of Patient 9 for initiation of dialysis and after skin asepsis, touched and repalpated the site prior to cannulation. During an interview on 12/04/2025 beginning at 2:30 PM, PCT 2 indicated that if repalpation is necessary after skin asepsis, then the cannulation site must be cleaned again prior to cannulation.”

    Rule V0550. Standard-level.

  • Measure, Analyze And Track Quality Indicators

    The inspectors’ words, 146 words

    “QAPI-MEASURE/ANALYZE/TRACK QUAL INDICATORS CFR(s): 494.110(a)(2) The dialysis facility must measure, analyze, and track quality indicators or other aspects of performance that the facility adopts or develops that reflect processes of care and facility operations. These performance components must influence or relate to the desired outcomes or be the outcomes themselves. This Standard is not met as evidenced by: 51199 Based on record review and interview, the facility failed to analyze and track the quality indicators related to personnel and ESRD Network communications in 1 of 1 dialysis facility. Findings include: A review of the facility’s QAPI failed to evidence personnel qualifications and issues nor communication with the ESRD Network were addressed or monitored. During an interview, on 12/08/2025 beginning at 12:00, the FA indicated that personnel qualifications and issues as well as ESRD Network relationship and communication is not an area that is addressed in QAPI.”

    Rule V0628. Standard-level.

  • Continuously Monitor/Take Action/Track/Sustain Improve

    The inspectors’ words, 149 words

    “QAPI-MONITOR/ACT/TRACK/SUSTAIN IMPROVE CFR(s): 494.110(b) The dialysis facility must continuously monitor its performance, take actions that result in performance improvements, and track performance to ensure that improvements are sustained over time. This Standard is not met as evidenced by: 51199 Based on record review and interview, the clinic failed to evidence their interventions or actions taken in QAPI to monitor, improve, and track performance in 1 of 1 dialysis facility. Findings include: A review of the facility QAPI evidenced incomplete water log entries reported by the Technical Department for the months of October 2025, September 2025, August 2025, July 2025, and May 2025; the QAPI documentation failed to evidence documentation of a plan to improve and track performance in this area. During an interview, on 12/08/2025 beginning at 12:00 PM, the FA indicated the incomplete log entries were entered into QAPI, but there was no action plan to improve performance.”

    Rule V0638. Standard-level.

  • Md Resp - Ensure All Adhere To P&P

    The inspectors’ words, 353 words

    “MD RESP-ENSURE ALL ADHERE TO P&P CFR(s): 494.150(c)(2)(i) The medical director must- (2) Ensure that- (i) All policies and procedures relative to patient admissions, patient care, infection control, and safety are adhered to by all individuals who treat patients in the facility, including attending physicians and nonphysician providers; This Standard is not met as evidenced by: 51199 Based on record review and interview, the dialysis facility failed to ensure their policies and procedures were adhered to in regards to accurately administering the dialysis prescription in 1 of 1 clinical record reviewed of a new admission (Patient 5), and in 1 of 1 personnel record reviewed of a Social Worker in regards to TB testing of employees (SW1). Findings include: 1. A policy, dated 05/01/2023, titled “Patient Assessment and Monitoring,” indicated during hemodialysis treatment, dialysate flow rate [DFR] should be checked for accuracy to ensure the prescribed flow rate matches what is being delivered. A clinical record review, for Patient 5, evidenced the prescribed DFR was 800 and on 11/28/2025 the DFR was run at 500 for the duration of the treatment. During an interview on 12/08/2025 beginning at 12:00 PM, the FA indicated there would be no reason a DFR would be administered differently than prescribed, other than error. 2. A policy, dated 05/05/2025, titled “Employee Tuberculosis Testing,” indicated that Tuberculosis skin testing should be performed at the time of hire and the results should be documented in the employee medical record. A personnel record review for SW 1 evidenced a hire date of 12/08/2018, and failed to evidence a Tuberculosis skin test results nor a negative history. During an interview on 12/08/2025 beginning at 12:30, SW 1 indicated that she was originally hired by the company for a different location in 2018 and received a Tuberculosis skin test at that time. SW 1 indicated she has transferred to several different locations over the years and had the skin test repeated when the original test results could not be located. She relayed the test was not repeated when she transferred to this facility, nor evidence that she had previously had a negative test.”

    Rule V0715. Standard-level.

The one inspection in the last three years
Inspections on record
DateVisitWhat the record says
Dec 8, 2025Standard inspection12 deficiencies cited

Is it certified?

  • Medicare-certified dialysis facility, certified or last recertified on Jan 1, 2000.
  • Run for profit.
  • No deficiency on record in the last three years.
  • Chain: Fresenius Medical Care.
Source:

CMS, Dialysis Facility - Listing by Facility. Released Jul 15, 2026. Updated quarterly.

CMS, QCOR Dialysis Facility (ESRD) 2567 statements of deficiencies. Released Oct 28, 2026. Updated monthly.

Has it been penalized?

  • CMS doesn’t publish fines or other penalties for dialysis facilities.
  • None of the deficiencies cited in the last three years came from an inspection that included a complaint investigation.
Source:

CMS, QCOR Dialysis Facility (ESRD) 2567 statements of deficiencies. Released Oct 28, 2026. Updated monthly.

What does it offer?

  • Offers in-center hemodialysis.
  • Doesn’t offer peritoneal dialysis, home hemodialysis training and a shift starting at 5 p.m. or later.
  • 17 dialysis stations.
  • CMS doesn’t publish staffing levels for dialysis facilities.
Source:

CMS, Dialysis Facility - Listing by Facility and Patient survey (ICH CAHPS) - Facility. Released Jul 15, 2026. Updated quarterly.

Who owns it?

  • Part of the Fresenius Medical Care chain, as CMS lists it.
  • CMS publishes no owners file for dialysis facilities, so no other owner is named.
Source:

CMS, Dialysis Facility - Listing by Facility. Released Jul 15, 2026. Updated quarterly.

How do patients fare?

  • CMS rates it 3 of 5 stars (Jan 1, 2021 to Dec 31, 2024).
  • Compared by CMS with what it expects for the facility’s patients: patient survival, as expected; hospital admissions, as expected; readmissions to hospital, as expected; blood transfusions, as expected; emergency department visits, as expected; emergency department visits within 30 days of a hospital stay, as expected; bloodstream infections, as expected; patients dialysing through a fistula, as expected; patients on a kidney transplant waitlist, as expected.
  • CMS gives no : “Too few completed survey responses to report.”
Source:

CMS, Dialysis Facility - Listing by Facility and Patient survey (ICH CAHPS) - Facility. Released Jul 15, 2026. Updated quarterly.

Ratings and measures, as published
MeasureRating
Star ratingCMS’s star rating for dialysis facilities, 1 to 53 of 5
Patient survey ratingCMS star rating from its survey of the facility’s in-center hemodialysis patients (ICH CAHPS), 1 to 5Not rated

How does it compare nearby?

  • Of the 6 dialysis facilities CMS lists in Porter County, 6 have a star rating from CMS: 1 rated 5 of 5, 4 rated 3 of 5, 1 rated 2 of 5.
  • CMS rates this one 3 of 5.
Source:

CMS, Dialysis Facility - Listing by Facility. Released Jul 15, 2026. Updated quarterly.

What the record doesn’t say

  • CMS doesn’t publish fines or other penalties for dialysis facilities, so none are shown.
  • Only complaints that led to a citation appear; complaints that didn’t aren’t published.
  • QCOR says it has no dialysis inspection reports after February 19, 2026, while it moves to a new system: a later inspection is not shown yet.
  • CMS doesn’t publish staffing levels for dialysis facilities.
  • CMS publishes no owners file for dialysis facilities: the chain it lists is the only owner named.

Sources

If you believe something here is wrong, a correction starts with the agency whose record this copies: CMS. CMS’s own record is on Care Compare. If Hearth Record copied something wrongly, it will fix it and log the correction publicly, with its date: how corrections work.

Source:

This copy of the record was made on Oct 30, 2026.

From the facility

The public record above comes first. Anything the facility adds will appear here, below it, and say who wrote it.

Replies

None. Replies aren’t open yet. When they open, the facility will be able to answer any finding for free, and the reply will show beside that finding, unedited and dated.

Paid listing

None. Paid listings aren’t open yet. When they open, anything paid for will say “Paid by the facility”, and paying will never change a finding, a rating or the order of results.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.