Hearth Record

DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500…

Medical device recall Z-3200-2026. Published in FDA’s Enforcement Report on Sep 23, 2026.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Sep 23, 2026. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8 C-M062-6-800 C-M062-7-500 C-M063-2-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-000 SHOULDER - GLENOID #11 5361-6111 SHOULDER - GLENOID #5 5361-6105 SHOULDER - GLENOID #7 5361-6107 SHOULDER - GLENOID #9 5361-6109 SHOULDER - KEELED GLENOID #11 5361-6211 SHOULDER - KEELED GLENOID #5 5361-6205 SHOULDER - KEELED GLENOID #7 5361-6207 SHOULDER - KEELED GLENOID #9 5361-6209 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

12,156 US, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

C-M064-1-300 No UDI Lot RX06413A C-M062-6-800 No UDI Lot RX06268A C-M062-7-500 No UDI Lot RX06275A C-M063-2-600 No UDI Lot RX06326A C-M063-5-000 No UDI Lots RX06350A RX06350AA 5361-6111 UDI-DI 07613327138191 Lots AFZT00U AFZT00Z P2N04B…

Every code shown here, 1,281 characters

C-M064-1-300 No UDI Lot RX06413A C-M062-6-800 No UDI Lot RX06268A C-M062-7-500 No UDI Lot RX06275A C-M063-2-600 No UDI Lot RX06326A C-M063-5-000 No UDI Lots RX06350A RX06350AA 5361-6111 UDI-DI 07613327138191 Lots AFZT00U AFZT00Z P2N04B P3K57 P3K57B P3K57F PYE43B PZA70F PZA70U 5361-6105 UDI-DI 07613327138214 Lots AF2S09U AF2S09Z P1H41 P4T02 P4T02X1 P4T02X2 P4T02X3 P4T02X4 P4T02X5 P4T02X6 5361-6107 UDI-DI 07613327138207 Lots AF3E00 AF3E00B AF3E00F AF3E00G AF3E00Q AF3E00S P4A238 P4A238X1 P4A238X2 P4A238X3 P4A238X4 P4A238X5 P4A238X6 P4A238X7 P4A238X8 P4L09 P4L09X1 P4L09X10 P4L09X2 P4L09X3 P4L09X4 P4L09X5 P4L09X6 P4L09X7 P4L09X8 P4L09X9 P4N06 P4N06X1 P4N06X2 P4N06X3 P4N06X4 P4N06X5 P4N06X6 P4N06X7 P4T03 P4T03X1 P4T03X2 P4T03X3 P4T03X4 P4T03X5 P4T03X6 P4T03X7 P4T03X8 P4V01 P4V01X1 P4V01X2 P4V01X3 P4V01X4 P4V01X5 P4V01X6 P4V01X7 P4V01X8 5361-6109 UDI-DI 07613327138153 Lots AF3E02I AF3E02S AF3E02Z P1H39Q 5361-6211 UDI-DI 07613327138283 Lots AFZS09F AFZS09G AFZS09S P2N08 PYC30X1 PYW24X1 PZKDHX1 5361-6205 UDI-DI 07613327138177 Lots 1NMMPE 1NNMPE P3K58 P3K58B P3K58F 5361-6207 UDI-DI 07613327138269 Lots AF2S10B AF2S10F AF2S10G P4A241 P4A241X1 P4A241X2 P4A241X3 P4V03 P4V03X1 P4V03X2 P4V03X3 P4V03X4 P4V03X5 PZKDFX1 5361-6209 UDI-DI 07613327138238 Lots AF2K03Z AFZS10S AFZS10U

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Jul 30, 2026.
  • FDA classified it on Sep 16, 2026.
  • FDA’s Enforcement Report published it on Sep 23, 2026.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Howmedica Osteonics Corp., of Mahwah, NJ.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.