Products that contain the Pericare Wipes: Bardex Lubricath Foley Catheter Tray
Medical device recall Z-3114-2026. Published in FDA’s Enforcement Report on Sep 16, 2026.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Sep 16, 2026. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Ongoing its status, in FDA’s file
- Nationwide FDA’s word, and 48 states named
- 12,463 trays/kits how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
Products that contain the Pericare Wipes: Bardex Lubricath Foley Catheter Tray REF: 899914 899916 899918 UDI codes: 00801741024801¿ 10801741024808 00801741024818¿ 10801741024815 00801741024825¿ 10801741024822 SureStep” Foley Tray System with Bardex¿ Lubricath Temp Sensing Foley Catheter REF: A119416M UDI code: 00801741073700 10801741073707 SureStep” Foley Tray System Bard¿ Lubricath Foley REF: A800361 UDI code: 00801741073939 10801741073936 SureStep” Foley Tray System Bard¿ Lubricath¿ Foley Catheter Tray REF: A899616 UDI code: 00801741104534¿ 10801741104531 SureStep” Foley Tray System Bardex¿ Lubricath¿ Foley Catheter REF: A899914 UDI code: 00801741074004 10801741074001 SureStep” Foley Tray System Bardex¿ Lubricath¿ Foley Catheter REF: A899916 UDI code: 00801741074011¿ 10801741074018 SureStep” Foley Tray System Bardex¿ Lubricath¿ Temperature Sensing Catheter REF: A900216 UDI code: 00801741120978¿ 10801741120975 SureStep” Foley Tray System Bard¿Lubricath¿ Foley Catheter REF: A902914 A902916 A902918 UDI code: 00801741074066¿10801741074063 00801741074073 10801741074070 00801741074080¿10801741074087 Bard¿ Urine Meter Foley Catheter Tray With Bardex¿ Lubricath¿ Foley Catheter REF: 304916 UDI code: 00801741019736 01801741019739 Bard¿ Urine Meter Foley Catheter Tray With Bardex¿ Lubricath¿ Foley Catheter REF: 902916 UDI codes: 00801741024863 10801741024860 (416A) SureStep Foley Tray System Bardex IC Complete Care (714A/716A/718A)SureStep Foley Tray System Bardex IC Complete REF: A304914 A304916 A304918 UDI codes: 00801741104350¿ 10801741104357 00801741104367¿ 10801741104364 00801741100499¿ 10801741100496 SureStep” Foley Tray System Bard¿ Lubricath Foley REF: A800365 UDI codes: 00801741073960 10801741073967 SureStep” Foley Tray System Bardex¿ Lubricath¿ Foley Catheter REF: A899918 UDI codes: 00801741074028 10801741074025 SureStep” Foley Tray System Bard¿Lubricath¿ Foley Catheter REF: A902618 UDI codes: 00801741104640 10801741104647 SureStep” Foley Tray System Bard¿Lubricath¿ Foley Catheter REF: A902616 UDI codes: 00801741104633¿ 10801741104630 SureStep Intermittent Catheter Tray REF: INTL16C UDI codes: 00801741183195¿ 10801741183192 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.
12,463 trays/kits, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
Catalog Numbers / UDI codes: 899914 / 00801741024801¿ 10801741024808 Lot Number: NGKU4611 899916 / 00801741024818¿ 10801741024815 Lot Numbers: NGKU2448 NGKU4610 899918 / 00801741024825¿ 10801741024822 Lot Number: NGKU3555 ***** SureStep”…
Every code shown here, 1,998 characters
Catalog Numbers / UDI codes: 899914 / 00801741024801¿ 10801741024808 Lot Number: NGKU4611 899916 / 00801741024818¿ 10801741024815 Lot Numbers: NGKU2448 NGKU4610 899918 / 00801741024825¿ 10801741024822 Lot Number: NGKU3555 ***** SureStep” Foley Tray System with Bardex¿ Lubricath Temp Sensing Foley Catheter REF: A119416M UDI code: 00801741073700 10801741073707 SureStep” Foley Tray System Bard¿ Lubricath Foley REF: A800361 UDI code: 00801741073939 10801741073936 SureStep” Foley Tray System Bard¿ Lubricath¿ Foley Catheter Tray REF: A899616 UDI code: 00801741104534¿ 10801741104531 SureStep” Foley Tray System Bardex¿ Lubricath¿ Foley Catheter REF: A899914 UDI code: 00801741074004 10801741074001 SureStep” Foley Tray System Bardex¿ Lubricath¿ Foley Catheter REF: A899916 UDI code: 00801741074011¿ 10801741074018 SureStep” Foley Tray System Bardex¿ Lubricath¿ Temperature Sensing Catheter REF: A900216 UDI code: 00801741120978¿ 10801741120975 SureStep” Foley Tray System Bard¿Lubricath¿ Foley Catheter REF: A902914 A902916 A902918 UDI code: 00801741074066¿10801741074063 00801741074073 10801741074070 00801741074080¿10801741074087 Bard¿ Urine Meter Foley Catheter Tray With Bardex¿ Lubricath¿ Foley Catheter REF: 304916 UDI code: 00801741019736 01801741019739 Bard¿ Urine Meter Foley Catheter Tray With Bardex¿ Lubricath¿ Foley Catheter REF: 902916 UDI codes: 00801741024863 10801741024860 (416A) SureStep Foley Tray System Bardex IC Complete Care (714A/716A/718A)SureStep Foley Tray System Bardex IC Complete REF: A304914 A304916 A304918 UDI codes: 00801741104350¿ 10801741104357 00801741104367¿ 10801741104364 00801741100499¿ 10801741100496 SureStep” Foley Tray System Bard¿ Lubricath Foley REF: A800365 UDI codes: 00801741073960 10801741073967 SureStep” Foley Tray System Bardex¿ Lubricath¿ Foley Catheter REF: A899918 UDI codes: 00801741074028 10801741074025 SureStep” Foley Tray System Bard¿Lubricath¿ Foley Catheter REF: A902618 UDI codes: 00801741104640 10801741104647 SureStep” Foley Tray…
FDA’s list runs to 2,181 characters. The first part is shown; the rest is on FDA’s record.
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WY. The countries of Bahrain, Canada, Costs Rica, Guatemala, Mexico, Saudi Arabia, and United Arab Emirates.
The states and territories these words name
48 states and territories.
Every one of them
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- Delaware
- Florida
- Georgia
- Hawaii
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- Washington
- Wisconsin
- Wyoming
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Jun 25, 2026.
- FDA classified it on Sep 4, 2026.
- FDA’s Enforcement Report published it on Sep 16, 2026.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- C.R. Bard Inc, of Covington, GA.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- FDA’s list of codes runs to 2,181 characters; the first part is shown here, and the rest is on FDA’s record.
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.