Hearth Record

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO KIT, Pack Number…

Medical device recall Z-3114-2024. Published in FDA’s Enforcement Report on Sep 18, 2024.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Sep 18, 2024. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Medline convenience kits containing semi-rigid suction liners labeled as follows: a) ENDO KIT, Pack Number DYKE1092B; b) VIDEO BRONCHOSCOPY TRAY, Pack Number DYKE1666; c) SUCTION CANISTER KIT, Pack Number DYKM2196; d) SUCTION CANISTER KIT, Pack Number DYKM2196H; e) BRONCH LAB PACK-LF, Pack Number DYNJ24810B; f) THORACIC, Pack Number DYNJ907991F; g) RECTAL, Pack Number DYNJ908108A; h) ROBOTIC, Pack Number DYNJ908109C; i) CERVICAL, Pack Number DYNJ908748D; j) LAMI, Pack Number DYNJ908750D; k) MERCY TIFFIN BASIC-LF, Pack Number DYNJ909731; l) MERCY TIFFIN LITHOTOMY, Pack Number DYNJ909735; m) BASIC GENERAL SET UP, Pack Number DYNJ909945; n) BASIC GENERAL SET UP, Pack Number DYNJ909945A; o) AMB OCULOPLASTICS MUSC, Pack Number DYNJ910289; p) AMB OCULOPLASTICS MUSC, Pack Number DYNJ910289H; q) MTHZ TOTAL HIP, Pack Number DYNJC9311M; r) BRONCH KIT, Pack Number P906877

3117 units, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

a) DYKE1092B: UDI/DI case 40889942777366, UDI/DI each10889942777365, Lot Numbers: 24BBK418, 24DBH210; b) DYKE1666: UDI/DI case 40193489708487, UDI/DI each10193489708486, Lot Numbers: 24ADB503, 24CDA282, 24CDB279, 24DDB221, 24EDA406; c)…

Every code shown here, 1,850 characters

a) DYKE1092B: UDI/DI case 40889942777366, UDI/DI each10889942777365, Lot Numbers: 24BBK418, 24DBH210; b) DYKE1666: UDI/DI case 40193489708487, UDI/DI each10193489708486, Lot Numbers: 24ADB503, 24CDA282, 24CDB279, 24DDB221, 24EDA406; c) DYKM2196: UDI/DI case 40195327236916, UDI/DI each10195327236915, Lot Numbers: 23LBF762, 24BBB541, 24CBI973, 24DBH448; d) DYKM2196H: UDI/DI case 40195327236916, UDI/DI each10195327236915, Lot Numbers: 23LBF762, 24BBB541, 24CBI973, 24DBH448; e) DYNJ24810B: UDI/DI case 40889942440611, UDI/DI each10889942440610, Lot Numbers: 24CBE214; f) DYNJ907991F: UDI/DI case 40195327641468, UDI/DI each10195327641467, Lot Numbers: 24DLA042, 24DLA594; g) DYNJ908108A: UDI/DI case 40195327507153, UDI/DI each10195327507152, Lot Numbers: 24BBK532; h) DYNJ908109C: UDI/DI case 40195327507238, UDI/DI each10195327507237, Lot Numbers: 23LBT867, 24BBA187, 24DBI723, 24FBA058; i) DYNJ908748D: UDI/DI case 40195327549207, UDI/DI each10195327549206, Lot Numbers: 23LBE394, 24CBD863; j) DYNJ908750D: UDI/DI case 40195327549214, UDI/DI each10195327549213, Lot Numbers: 24FBC284; k) DYNJ909731: UDI/DI case 40195327464524, UDI/DI each10195327464523, Lot Numbers: 24ABU492; l) DYNJ909735: UDI/DI case 40195327464685, UDI/DI each10195327464684, Lot Numbers: 23LBH989, 24ABV099, 24CBG700; m) DYNJ909945: UDI/DI case 40195327512669, UDI/DI each10195327512668, Lot Numbers: 24ABV895; n) DYNJ909945A: UDI/DI case 40195327615803, UDI/DI each10195327615802, Lot Numbers: 24DBA530; o) DYNJ910289: UDI/DI case 40195327618675, UDI/DI each10195327618674, Lot Numbers: 24CBC847; p) DYNJ910289H: UDI/DI case 40198459012489, UDI/DI each10198459012488, Lot Numbers: 24CBC847; q) DYNJC9311M: UDI/DI case 40198459012489, UDI/DI each10198459012488, Lot Numbers: 24FDA111; r) P906877: UDI/DI case 40889942002581, UDI/DI each10889942002580, Lot Numbers: 24FDA021

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide distribution - US Nationwide and the countries of Canada, Panama.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Jun 27, 2024.
  • FDA classified it on Sep 11, 2024.
  • FDA’s Enforcement Report published it on Sep 18, 2024.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • MEDLINE INDUSTRIES, LP - Northfield, of Northfield, IL.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.