Hearth Record

NM/CT 870 DR

Medical device recall Z-2998-2020. Published in FDA’s Enforcement Report on Sep 30, 2020.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Sep 30, 2020. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Rotor bearing screws were found loose on detector.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

NM/CT 870 DR Model # H3100AS

31 units, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Serial # 870X64001 System ID: 030020NU06 Serial # 870X64004 System ID: M4018555 Serial # 870X64012 System ID: X634445001 Serial # 870X64007 System ID: V2728113 Serial # 870X64009 System ID: M4016255 Serial # 870X64026 System ID: M40230109…

Every code shown here, 1,004 characters

Serial # 870X64001 System ID: 030020NU06 Serial # 870X64004 System ID: M4018555 Serial # 870X64012 System ID: X634445001 Serial # 870X64007 System ID: V2728113 Serial # 870X64009 System ID: M4016255 Serial # 870X64026 System ID: M40230109 Serial # 870X64017 System ID: M40240121 Serial # 870X64018 System ID:M40240121 Serial # 870X64015 System ID:M40345223 Serial # 870X64016 System ID:499221NU01 Serial # 870X64031 System ID:083049209623319 Serial # 870X64014 System ID:IL1028NM41 Serial # 870X64022 System ID: A5173220 Serial # 870X64019 System ID: 0850260167XE6 Serial # 870X64020 System ID: 0850260169 Serial # 870X64025 System ID: EMM0135 Serial # 870X64010 System ID:NM2882PA Serial # 870X64027 System ID: PL0398NM01 Serial # 870X64029 System ID: PL1031NM01 Serial # 870X64030 System ID: PL2190NM01 Serial # 870X64028 System ID: PL0372NM01 Serial # 870X64023 System ID: 600042NM10 Serial # 870X64013 System ID: NP473631 Serial # 870X64021 System ID: G001NU12 Serial # 870X64008 System ID:00669NUC03

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

USA : AR CT FL GA IL KS KY LA MA MD MI MN MO NE NJ NV NY OH OK RI SC TN TX VA VT WA WI OUS: ARGENTINA AUSTRIA BELGIUM BRAZIL Canada CHINA FRANCE GERMANY GREECE INDIA ISRAEL ITALY JAPAN Korea (Republic of) NETHERLANDS NORWAY PANAMA POLAND PORTUGAL Russia SPAIN SWEDEN SWITZERLAND TAIWAN THAILAND TURKEY UNITED KINGDOM VIETNAM

The states and territories these words name

27 states and territories.

Every one of them
  • Arkansas
  • Connecticut
  • Florida
  • Georgia
  • Illinois
  • Kansas
  • Kentucky
  • Louisiana
  • Massachusetts
  • Maryland
  • Michigan
  • Minnesota
  • Missouri
  • Nebraska
  • New Jersey
  • Nevada
  • New York
  • Ohio
  • Oklahoma
  • Rhode Island
  • South Carolina
  • Tennessee
  • Texas
  • Virginia
  • Vermont
  • Washington
  • Wisconsin

The words say it went outside the United States too.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Aug 14, 2020.
  • FDA classified it on Sep 21, 2020.
  • FDA’s Enforcement Report published it on Sep 30, 2020.
  • FDA terminated it on May 6, 2021.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • GE Healthcare, LLC, of Waukesha, WI.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.