Hearth Record

uMI 550 System - Product Usage: intended to be operated by qualified healthcare professionals to assist in…

Medical device recall Z-2593-2021. Published in FDA’s Enforcement Report on Oct 6, 2021.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Oct 6, 2021. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

The EXAM acquisition workflow could occasionally freeze after canceling the PET scan.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

uMI 550 System - Product Usage: intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

23 systems, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Model: 88000057 GTIN/UDI Codes: 06971576832026 / 01 06971576832026 240 88000057 21 200017 06971576832026 / 01 06971576832026 240 88000057 21 200023 06971576832026 / 01 06971576832026 240 88000057 21 200024 06971576832026 / 01…

Every code shown here, 1,565 characters

Model: 88000057 GTIN/UDI Codes: 06971576832026 / 01 06971576832026 240 88000057 21 200017 06971576832026 / 01 06971576832026 240 88000057 21 200023 06971576832026 / 01 06971576832026 240 88000057 21 200024 06971576832026 / 01 06971576832026 240 88000057 21 200036 06971576832026 / 01 06971576832026 240 88000057 21 200045 06971576832026 / 01 06971576832026 240 88000057 21 200047 06971576832057 / 01 06971576832057 240 88000057 21 230002 06971576832057 / 01 06971576832057 240 88000057 21 230003 06971576832057 / 01 06971576832057 240 88000057 21 230004 06971576832057 / 01 06971576832057 240 88000057 21 200059 06971576832057 / 01 06971576832057 240 88000057 21 230005 06971576832057 / 01 06971576832057 240 88000057 21 230006 06971576832057 / 01 06971576832057 240 88000057 21 230007 06971576832057 / 01 06971576832057 240 88000057 21 230008 06971576832057 / 01 06971576832057 240 88000057 21 230009 06971576832057 / 01 06971576832057 240 88000057 21 230010 06971576832057 / 01 06971576832057 240 88000057 21 230011 06971576832057 / 01 06971576832057 240 88000057 21 230012 06971576832057 / 01 06971576832057 240 88000057 21 230013 06971576832057 / 01 06971576832057 240 88000057 21 230014 06971576832057 / 01 06971576832057 240 88000057 21 230015 06971576832057 / 01 06971576832057 240 88000057 21 230016 06971576832057 / 01 06971576832057 240 88000057 21 230017 Serial Numbers: 200017; 200023; 200024; 200036; 200045; 200047; 230002; 230003; 230004; 200059; 230005; 230006; 230007; 230008; 230009; 230010; 230011; 230012; 230013; 230014; 230015; 230016; 230017;

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

U.S. Nationwide distribution in the states of CA, IL, IN, TX, and WI.

The states and territories these words name

  • California
  • Illinois
  • Indiana
  • Texas
  • Wisconsin

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Aug 2, 2021.
  • FDA classified it on Sep 29, 2021.
  • FDA’s Enforcement Report published it on Oct 6, 2021.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Shanghai United Imaging Healthcare Co., Ltd., of Shanghai, China.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.