Superion IDS Kit Superion Indirect Decompression System 8MM, VF Implant 101-9808 00884662000529 Superion…

Medical device recall Z-2485-2025. Published in FDA’s Enforcement Report on Sep 10, 2025.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Sep 10, 2025. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

As a result of a comprehensive product performance review it was determine that the device and instruments are not consistently meeting expected performance levels. Issues include breakage of the instruments and implant/device.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Superion IDS Kit Superion Indirect Decompression System 8MM, VF Implant 101-9808 00884662000529 Superion Indirect Decompression System 10MM, VF Implant 101-9810 00884662000536 Superion Indirect Decompression System 12MM, VF Implant 101-9812 00884662000543 Superion Indirect Decompression System 14MM, VF Implant 101-9814 00884662000550 Superion Indirect Decompression System 16MM, VF Implant 101-9816 00884662000567 Superion Indirect Decompression System Kit, VF Instrument 102-9800 00884662000574 The Vertiflex Procedure Instrument Platform, VF Instrument 140-9800 00884662000611 The Superion IDS Kit includes a set of proprietary instruments to deliver the Superion Implant in a minimally invasive manner. Instruments specifically designed for implanting the Superion Implant are sterile, single-use disposable instruments, consisting of a Dilator Assembly, Cannula Assembly, Interspinous Gauge, Inserter, Reamer, and Driver.

70,516, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Superion Indirect Decompression System 8MM, VF Implant 101-9808 00884662000529 ALL Lot/Batch Number: 40280174 40145632 40117221 40113198 40113255 40113385 40113391 40113482 40106856 40027385 26176977 800276 800301 Superion Indirect…

Every code shown here, 1,999 characters

Superion Indirect Decompression System 8MM, VF Implant 101-9808 00884662000529 ALL Lot/Batch Number: 40280174 40145632 40117221 40113198 40113255 40113385 40113391 40113482 40106856 40027385 26176977 800276 800301 Superion Indirect Decompression System 10MM, VF Implant 101-9810 00884662000536 ALL Lot/Batch Numbers: 40222297 40213592 40199967 40198809 40195249 40186649 40181505 40166300 40157622 40141657 40141770 40136859 40131798 40132036 40112806 40106870 40093564 40080016 40080111 40076497 40070306 40070307 40052009 40045101 40034339 40027338 40027364 40022398 40018296 40018338 40010128 40004038 40004068 40000647 30244690 30178729 29985230 29849515 29849969 29668712 29668713 29578948 29578949 29360265 29360263 29360264 29160559 29160625 28756501 28627193 28627192 28627194 28501592 28501591 28501593 28344375 28344376 28344377 28133122 800279 800345 800269 800278 800268 800280 800261 800270 700089 800259 800260 800258 800250 800317 800318 700085 Superion Indirect Decompression System 12MM, VF Implant 101-9812 00884662000543 ALL Lot/Batch Numbers: 40280071 40245960 40222355 40211234 40211404 40211463 40199117 40199211 40186760 40165937 40157506 40157279 40139759 40140041 40132011 40129328 40114193 40114290 40106757 40106924 40093299 40093565 40080015 40080124 40070323 40051777 40045673 40045706 40044635 40045080 40027314 40027337 40027346 40018297 40018311 40018344 40018360 40010112 40010126 40010144 40004051 40004081 40003228 40000665 30244396 30244397 29985232 29985233 29927752 29850082 29850085 29850083 29788495 29667699 29668637 29668640 29623992 29556093 29547992 29540069 29540067 29540068 29394426 29360815 29361201 29361220 29361221 29361159 29361200 29243822 29244010 29244067 29244068 29170416 29164935 28737193 28627392 28627985 28627195 28620905 28545960 28460945 28460942 28460944 28460941 28460943 28444915 28344378 28344379 28282295 28282296 28133124 28133125 28133123 28123077 28051836 28047778 28047777 28047779 27909032 27844690 27812316 27812317 27809085…

FDA’s list runs to 11,063 characters. The first part is shown; the rest is on FDA’s record.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.

The states and territories these words name

50 states and territories.

Every one of them
  • Alaska
  • Alabama
  • Arkansas
  • Arizona
  • California
  • Colorado
  • Connecticut
  • District of Columbia
  • Delaware
  • Florida
  • Georgia
  • Hawaii
  • Iowa
  • Idaho
  • Illinois
  • Indiana
  • Kansas
  • Kentucky
  • Louisiana
  • Massachusetts
  • Maryland
  • Maine
  • Michigan
  • Minnesota
  • Missouri
  • Mississippi
  • Montana
  • North Carolina
  • North Dakota
  • Nebraska
  • New Hampshire
  • New Jersey
  • New Mexico
  • Nevada
  • New York
  • Ohio
  • Oklahoma
  • Oregon
  • Pennsylvania
  • Puerto Rico
  • South Carolina
  • South Dakota
  • Tennessee
  • Texas
  • Utah
  • Virginia
  • Washington
  • Wisconsin
  • West Virginia
  • Wyoming

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Jun 6, 2025.
  • FDA classified it on Sep 2, 2025.
  • FDA’s Enforcement Report published it on Sep 10, 2025.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Boston Scientific Neuromodulation Corporation, of Valencia, CA.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • FDA’s list of codes runs to 11,063 characters; the first part is shown here, and the rest is on FDA’s record.
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.