UniCel DxI 800 Access Immunoassay Analyzer
Medical device recall Z-2266-2025. Published in FDA’s Enforcement Report on Aug 13, 2025.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Aug 13, 2025. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Ongoing its status, in FDA’s file
- Nationwide FDA’s word for where it went
- 2203 systems how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
UniCel DxI 800 Access Immunoassay Analyzer, Part Number 973100
2203 systems, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
UDI-DI: 15099590265373; Serial Numbers: 608114 608209 609132 609163 608344 608113 608233 609224 609218 609221 609070 609071 609117 608164 609591 609692 608197 609534 609455 609524 609319 609062 609094 609452 609750 609801 609686 609198…
Every code shown here, 2,000 characters
UDI-DI: 15099590265373; Serial Numbers: 608114 608209 609132 609163 608344 608113 608233 609224 609218 609221 609070 609071 609117 608164 609591 609692 608197 609534 609455 609524 609319 609062 609094 609452 609750 609801 609686 609198 609644 609518 609745 609496 608387 609183 609188 609699 609735 609618 609492 609028 609583 609765 609790 609187 605528 609137 609157 609226 608242 608243 609191 609527 609256 608996 609003 609327 609262 609264 609294 609321 609329 609265 609290 609397 609521 609320 608106 609373 608386 608389 609372 609846 609153 609241 609498 609519 609520 609744 609704 609491 609802 609810 609807 609806 608234 609286 609739 609824 609270 609103 609643 609641 609123 609192 609093 609499 609497 609429 609410 608994 609122 608247 609334 607653 609849 609100 609101 609370 608391 608362 609371 609136 609195 608264 608265 607196 609784 609494 609223 609526 609220 608347 609197 609099 609098 609097 609576 609285 609225 609626 609640 609150 609233 609803 609283 609322 609537 609318 609827 609462 609180 609408 609244 609754 609795 609791 609797 609800 609789 609532 609531 609538 609539 609543 609541 609599 609338 608256 609323 609536 608346 609833 606661 608388 609001 609246 608393 609369 609535 609031 609092 609582 609834 604083 609740 609513 609210 609378 608188 609017 609018 609005 608172 609030 609385 609649 609053 609263 609184 608176 608238 609209 609020 608385 608384 608382 608276 608202 608997 609023 608205 609004 609102 609016 608175 609015 608171 609019 608246 609002 608379 608235 608381 608170 609051 609193 609196 609194 609135 609151 609050 609014 609203 608165 608225 608983 609069 609068 608093 608136 608984 609011 608231 608227 609083 609215 609108 608365 608110 609272 609025 609024 609033 608990 608098 608223 609284 609255 607108 608267 609143 608217 608190 609276 608988 608987 608169 609054 608221 608133 608135 609277 608210 608383 608995 608986 608121 608241 608184 608204 609104 609106 608157 609280 608350 609251 608219 608159 608168 608122…
FDA’s list runs to 15,460 characters. The first part is shown; the rest is on FDA’s record.
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide distribution - US Nationwide.
These words name no state.
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Jun 25, 2025.
- FDA classified it on Aug 5, 2025.
- FDA’s Enforcement Report published it on Aug 13, 2025.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Beckman Coulter, Inc., of Chaska, MN.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- FDA’s list of codes runs to 15,460 characters; the first part is shown here, and the rest is on FDA’s record.
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.