GE 1.5T Signa HDxt, 3.0T Signa HDxt, Signa Vibrant, HD23 Software and the following

Medical device recall Z-2266-2013. Published in FDA’s Enforcement Report on Oct 2, 2013.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Oct 2, 2013. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

GE Healthcare has recently become aware of a potential safety issue due to non-unique series identifiers embedded in image headers of certain GE MR Products. For ClariView and Screen Save from GVTK (Gems Visualization ToolKit) Viewer, the MR Product can generate image headers with non-unique series identifiers, which when exported to some PACS systems, results in the series being associated wit

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

GE 1.5T Signa HDxt, 3.0T Signa HDxt, Signa Vibrant, HD23 Software and the following model numbers: 2395001, 2395001-2, 2395001-3, 2395001-4, 5269997-2, 5269997-4, 2377062-61, 5127452-3, 5127452, 5337894-3, 5337894-4, 5911000-4, 5911000-5, 53085735308573-2. GE Discovery MR750w 3.0T, Discovery MR750 3.0T, Optima MR450w 1.5T, Discovery MR450 1.5T. The 1.5T Signa HDx family and 3.0T Signa HDx family are a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. IT is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including but not limited to head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used

1,246 (521 US and 725 OUS), in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

REF 60850 Mfg Lot or Serial # 00000204882MR5 00000233303MR7 00000241103MR1 00000245171MR4 00000247204MR1 00000173038MR1 00000136104MR7 00000264553MR9 00000265096MR8 00000005532YR7 00000007097YR9 00000007556YR4 00000281841MR7 00000289583MR7…

Every code shown here, 1,988 characters

REF 60850 Mfg Lot or Serial # 00000204882MR5 00000233303MR7 00000241103MR1 00000245171MR4 00000247204MR1 00000173038MR1 00000136104MR7 00000264553MR9 00000265096MR8 00000005532YR7 00000007097YR9 00000007556YR4 00000281841MR7 00000289583MR7 0000SGH919066P 00000013883YR4 0000SGH231PPQK 00002UA2050JKX 00002UA2400LWP 00002UA24315PT 00002UA24315PY 00002UA24315P5 00002UA24907W7 00000266797MRO 00000267189MR0 00000270550MR7 00000283466MR1 00000305180MR2 00000306054MR8 0000267533MR8A Not Available 00000237000MR5 00000226580MR9 00000230105MR9 00000285956MR9 00000305386MR5 00000012562YR5 Not Available 000002UA4907WH 00002UA2400LWZ 00000177845MR5 00000165805MR3 00000836BD4DE2 00000012383YR6 00000269416MR4 Not Available 00002UA2400LWN 00000284254MR0 00000287469MR1 00000304797MR4 00002UA24907WK 00000005220YR9 00002UA2420VVZ 00002UA24907YC 00002UA24907YD 00002UA2510YTL 00000287996MR3 00000934BD5E4F 00000005966YR7 00000919BD5B8F 00000285151MR7 00000275969MR4 00000286019MR5 00000288669MR5 00001008BD6522 00000PH150024A 00000280423MR5 00000283613MR8 00000288416MR1 00000289840MR1 00000291157MR6 00000297685MR0 00000012064YR2 00000012526YR0 00000013660YR6 00000013820YR6 00000015654YR7 00001052XF502C 00001052XF5068 00001105XF5017 0000SGH048S36G 0000SGH105P174 0000SGH150RLYV 0000SGH150RLYX 0000SGH201TZ0V 0000SGH203QRPF 00002UA2050JKD 00002UA2050JL0 00002UA2070RGR 00002UA211170R 00002UA2281GGK 00002UA2281GGL 00002UA22827QF 00002UA2321LKW 00002UA2400LWJ 00002UA2400LWW 00002UA24907WJ 00000026639MR2 00000276261MR5 00000246260MR4 00000273065MR3 00000011594YR9 00000013110YR2 00001048XF5001 0000SGH106PMD7 814 00002UA2050JK7 00002UA2050JL2 00002UA2201G10 00002UA2201G19 00002UA22912Y5 00000264642MR0 00000305822MR9 00000265606MR4 Not Available 00000304307MR2 00000013981YR6 00002UA2400LX2 00000213294MR2 809 811 00000273256MR8 00000221025MR0 00000174670MR0 00000259576MR7 00000288072MR2 00000298886MR3 00000304463MR3 00000002761M65 00000002763M61 00000002766M64 00000003066M68 00000012720M69…

FDA’s list runs to 18,673 characters. The first part is shown; the rest is on FDA’s record.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide Distribution - All states in continental USA including DC except AK, AZ, RI and WY. and the countries of : AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGARY, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, LEANON, MARINIQUE, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE,SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, LUXEMBOURG, PORTUGAL, VIETNAM and URUGUAY.

The states and territories these words name

  • District of Columbia

They leave out Alaska, Arizona, Rhode Island and Wyoming.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Apr 9, 2013.
  • FDA classified it on Sep 23, 2013.
  • FDA’s Enforcement Report published it on Oct 2, 2013.
  • FDA terminated it on May 5, 2014.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • GE Healthcare, LLC, of Waukesha, WI.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • FDA’s list of codes runs to 18,673 characters; the first part is shown here, and the rest is on FDA’s record.
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.