TWINFIX Ultra HA 4.5 mm Suture Anchor with 2 ULTRABRAID Suture, White / Blue
Medical device recall Z-2229-2013. Published in FDA’s Enforcement Report on Sep 25, 2013.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Sep 25, 2013. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Terminated its status, in FDA’s file
- Nationwide FDA’s word, and 1 state named
- 8889 US how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
TWINFIX Ultra HA 4.5 mm Suture Anchor with 2 ULTRABRAID Suture, White / Blue; Part Number: 72202597 Biodegradable suture anchor
8889 US, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
50324706 50324707 50324958 50325148 50326624 50336118 50336716 50336759 50337105 50337109 50337112 50339394 50339648 50339653 50339656 50339873 50341613 50343278 50344410 50346095 50350419 50352199 50357080 50358106 50358736 50359998…
Every code shown here, 773 characters
50324706 50324707 50324958 50325148 50326624 50336118 50336716 50336759 50337105 50337109 50337112 50339394 50339648 50339653 50339656 50339873 50341613 50343278 50344410 50346095 50350419 50352199 50357080 50358106 50358736 50359998 50361739 50362786 50363871 50366812 50368582 50370504 50372042 50374257 50376594 50378774 50380528 50382637 50384518 50386924 50388022 50390005 50390988 50392388 50395082 50395975 50396667 50398098 50401255 50402220 50404745 50406626 50406675 50408537 50411606 50414330 50415861 50417004 50418086 50422213 50428616 50429266 50429632 50436702 50436747 50436748 50438319 50438651 50438735 50438739 50438918 50445013 50446067 50446068 50446460 50446995 50448591 50450274 50451213 50453452 50453995 50456041 50456190 50457051 50458600 50462057
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Slovenia, Spain, Sweden, Switzerland, Turkey, Ukraine and United Kingdom, Asia-pacific: Australia, China, Dubai, Hong Kong, India, Japan, Malaysia, New Zealand, Philippines, Sri Lanka, Singapore, South Korea, Taiwan, Thailand, Argentina, Bermuda, Brazil, Chile, Colombia, Costa Rica, Israel, Mexico, Puerto Rico, South Africa, and Venezuela.
The states and territories these words name
- Puerto Rico
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Terminated, in FDA’s file.
“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”
openFDA, on “Terminated”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Aug 6, 2013.
- FDA classified it on Sep 17, 2013.
- FDA’s Enforcement Report published it on Sep 25, 2013.
- FDA terminated it on Apr 20, 2016.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Smith & Nephew, Inc. Endoscopy Division, of Andover, MA.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.