VersaCross Access Solution (VXSK), Product Description and ID: VXSK
Medical device recall Z-2103-2025. Published in FDA’s Enforcement Report on Jul 16, 2025.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Jul 16, 2025. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Ongoing its status, in FDA’s file
- Nationwide FDA’s word for where it went
- 14,724 how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
VersaCross Access Solution (VXSK), Product Description and ID: VXSK (180J RF WIRE 63CM/45DEG/D0) VXSK0101 VXSK (180J RF WIRE 63CM/45DEG/D1) VXSK0102 VXSK (180J RF WIRE 63CM/55DEG/D0) VXSK0103 VXSK (180P RF WIRE 63CM/45DEG/D0) VXSK0121 VXSK (180P RF WIRE 63CM/45DEG/D1) VXSK0122 VXSK (180P RF WIRE 63CM/55DEG/D0) VXSK0123 VXSK (180P RF WIRE 63CM/55DEG/D1) VXSK0124 VXSK (180P RF WIRE 63CM/90DEG/D0) VXSK0125 VXSK (230J RF WIRE 63CM/45DEG/D0) VXSK0111 VXSK (230J RF WIRE 63CM/45DEG/D1) VXSK0112 VXSK (230P RF WIRE 63CM/45DEG/D0) VXSK0131 VXSK (230P RF WIRE 63CM/45DEG/D1) VXSK0132 VXSK (230P RF WIRE 63CM/55DEG/D0) VXSK0133 VXSK (230P RF WIRE 63CM/55DEG/D1) VXSK0134 VXSK (230P RF WIRE 81CM/45DEG/D0) VXSK0137 VXSK-180P RF WIRE 63CM/45DEG/D0 SHEATH VXSK0021 VXSK-180P RF WIRE 63CM/45DEG/D1 SHEATH VXSK0022 VXSK-180P RF WIRE 63CM/55DEG/D1 SHEATH VXSK0024 VXSK-180P RF WIRE 63CM/90DEG/D0 SHEATH VXSK0025 VXSK-230P RF WIRE 63CM/45DEG/D0 SHEATH VXSK0031 VXSK-230P RF WIRE 63CM/45DEG/D1 SHEATH VXSK0032 VXSK-230P RF WIRE 63CM/55DEG/D1 SHEATH VXSK0034
14,724, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
Product ID/UDI-DI/Batch Numbers: VXSK0021 UDI-DI 00685447011729 Lots 35590858 35607785 35627540; VXSK0022 UDI-DI 00685447011736 Lots 35487923 35544014 35552652 35566071 35595443 35604160 35616857 VXSK0024 UDI-DI 00685447011750 Lot…
Every code shown here, 1,282 characters
Product ID/UDI-DI/Batch Numbers: VXSK0021 UDI-DI 00685447011729 Lots 35590858 35607785 35627540; VXSK0022 UDI-DI 00685447011736 Lots 35487923 35544014 35552652 35566071 35595443 35604160 35616857 VXSK0024 UDI-DI 00685447011750 Lot 35576278; VXSK0025 UDI-DI 00685447011767 Lot 35582041; VXSK0031 UDI-DI 00685447011828 Lots 35499536 35627156; VXSK0032 UDI-DI 00685447011835 Lots 35508294 35518904 35563958 35582043 35635451; VXSK0034 UDI-DI 00685447011859 Lots 35573519 35593310; VXSK0101 UDI-DI 00685447019749 Lot 35893570; VXSK0102 UDI-DI 00685447019756 Lot 35720571; VXSK0103 UDI-DI 00685447019763 Lot 36042383; VXSK0111 UDI-DI 00685447019794 Lot 35944186; VXSK0112 UDI-DI 00685447019800 Lot 36061050; VXSK0121 UDI-DI 00685447019817 Lots 35745702 35787463; VXSK0122 UDI-DI 00685447019824 Lots 35699501 35720572 35768613 35780309 35790937 35830237 35833400; VXSK0123 UDI-DI 00685447019831 Lot 35843757; VXSK0124 UDI-DI 00685447019848 Lots 35697863 35883199; VXSK0125 UDI-DI 00685447019855 Lot 35859706; VXSK0131 UDI-DI 00685447019862 Lots 35729751 35807530 35900608; VXSK0132 UDI-DI 00685447019879 Lots 35695371 35729750 35807531 35910367 35955008; VXSK0133 UDI-DI 00685447019886 Lot 35839894; VXSK0134 UDI-DI 00685447019893 Lot 35666909; VXSK0137 UDI-DI 00685447019909 Lot 35985522
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.
These words name no state.
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Jun 10, 2025.
- FDA classified it on Jul 10, 2025.
- FDA’s Enforcement Report published it on Jul 16, 2025.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Boston Scientific Corporation, of Saint Paul, MN.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: N/A.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.