Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1
Medical device recall Z-2037-2026. Published in FDA’s Enforcement Report on May 6, 2026.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on May 6, 2026. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Ongoing its status, in FDA’s file
- Not named no state, in where it went
- 3290 kits how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. CATARACT PACK, Medline Kit Number/SKU DYNJ48725A; 2. OCULAR PACK, Medline Kit Number/SKU DYNJ64138B; 3. OCULOPLASTY PACK, Medline Kit Number/SKU DYNJ64572A; 4. OCULOPLASTY PACK, Medline Kit Number/SKU DYNJ64572B; 5. MICROSURGERY PACK, Medline Kit Number/SKU DYNJ67284A; 6. BASIC EYE PACK, Medline Kit Number/SKU DYNJ81411A; 7. SMR SINUS PACK, Medline Kit Number/SKU DYNJ84478; 8. FESS/SEPTO PACK, Medline Kit Number/SKU DYNJ85025; 9. PLASTIC EYE PACK, Medline Kit Number/SKU DYNJ89030; 10. NEURO, Medline Kit Number/SKU DYNJ902935K; 11. OCULAR, Medline Kit Number/SKU DYNJ909065; 12. OCULAR, Medline Kit Number/SKU DYNJ909065A.
3290 kits, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
Medline Kit Number/SKU DYNJ48725A: UDI/DI each 10193489853896, UDI/DI case 40193489853897, Lot Number: 24IMG757; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 22FME241; Medline Kit…
Every code shown here, 1,996 characters
Medline Kit Number/SKU DYNJ48725A: UDI/DI each 10193489853896, UDI/DI case 40193489853897, Lot Number: 24IMG757; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 22FME241; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 22DMC340; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21JMC829; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21IMC497; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21HMD139; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21GME747; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21FMA440; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21DBU602; Medline Kit Number/SKU DYNJ64138B: UDI/DI each 10193489355284, UDI/DI case 40193489355285, Lot Number: 21CBJ936; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23IBC227; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23FBA458; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23DBC239; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 23ABI841; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 22IBP260; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 22IBO842; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case 40195327112555, Lot Number: 22IBN801; Medline Kit Number/SKU DYNJ64572A: UDI/DI each 10195327112554, UDI/DI case…
FDA’s list runs to 10,516 characters. The first part is shown; the rest is on FDA’s record.
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide distribution.
These words name no state.
The words say it went outside the United States too.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Mar 19, 2026.
- FDA classified it on Apr 30, 2026.
- FDA’s Enforcement Report published it on May 6, 2026.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Medline Industries, LP, of Northfield, IL.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- FDA’s list of codes runs to 10,516 characters; the first part is shown here, and the rest is on FDA’s record.
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.