Hearth Record

P.R.O. Matt: (a) PROMATT PLUS 36 X 76 MATRESS, NYLON, MATERIAL NUMBER PMP3676NZBA (b) PROMATT PLUS 84”…

Medical device recall Z-2033-2018. Published in FDA’s Enforcement Report on Jun 6, 2018.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Jun 6, 2018. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Joerns Healthcare has identified a potential issue with some of the P.R.O. Matt and ProCair Mattresses. For the identified serial number range of P.R.O. Matt and ProCair Mattresses incorrect connectors (part #s 206-0401 & 206-0405) were provided by a Joerns supplier without the required internal shutoff valves, these shutoff valves prevent air from escaping from the mattress air cells.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

P.R.O. Matt: (a) PROMATT PLUS 36 X 76 MATRESS, NYLON, MATERIAL NUMBER PMP3676NZBA (b) PROMATT PLUS 84” MATRESS, NYLON, MATERIAL NUMBER PMP3684NZBA (c) PROMATT PLUS W/SIDEWALLS, 42X80, MATERIAL NUMBER PMS4280NZBA (d) PROMATT PLUS 36IN MATRESS, MATERIAL NUMBER PMP3680NZBA (e) PROMATT WITH SIDEWALLS ASSEMBLY, MATERIAL NUMBER PMS3680NZBA (f) PROMATT PLUS 80IN MATRESS, STRETCH, MATERIAL NUMBER PMP3680SZBA (g) PROMATT PLUS 42IN MATRESS, MATERIAL NUMBER PMP4280NZBA (h) PROMATT PLUS 85cm MATRESS, DURATUB, MATERIAL NUMBER PMPE3480SZBADT Product Usage: Non powered flotation therapy mattress. The device is intended to treat or prevent decubitus ulcers (bed sores).

915 units, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

P.R.O. Matt: (a) PROMATT PLUS 36 X 76 MATRESS, NYLON, MATERIAL NUMBER PMP3676NZBA; Serial Numbers: 1000057154 1000057153 (b) PROMATT PLUS 84” MATRESS, NYLON, MATERIAL NUMBER PMP3684NZBA; Serial Numbers: 1000051430 1000051425 1000051366…

Every code shown here, 1,990 characters

P.R.O. Matt: (a) PROMATT PLUS 36 X 76 MATRESS, NYLON, MATERIAL NUMBER PMP3676NZBA; Serial Numbers: 1000057154 1000057153 (b) PROMATT PLUS 84” MATRESS, NYLON, MATERIAL NUMBER PMP3684NZBA; Serial Numbers: 1000051430 1000051425 1000051366 1000051422 1000056411 1000058530 1000056414 (c) PROMATT PLUS W/SIDEWALLS, 42X80, MATERIAL NUMBER PMS4280NZBA; Serial Numbers: 1000049130 1000054638 1000054639 (d) PROMATT PLUS 36IN MATRESS, MATERIAL NUMBER PMP3680NZBA; Serial Numbers: 1000057195 1000055122 1000055401 1000057190 1000055328 1000054681 1000057193 1000057171 1000051436 1000055326 1000054386 1000054387 1000054385 1000051441 1000051432 1000051439 1000054936 1000054144 1000054935 1000054390 1000054368 1000051440 1000054932 1000055059 1000055134 1000055070 1000055058 1000055118 1000054872 1000055317 1000055130 1000055074 1000055313 1000055093 1000054906 1000055091 1000055132 1000055068 1000054926 1000054922 1000055065 1000054373 1000055062 1000055329 1000054140 1000054868 1000055319 1000054683 1000051437 1000055111 1000049127 1000054927 1000054380 1000054375 1000052270 1000054383 1000054940 1000053404 1000054143 1000054915 1000054917 1000054379 1000054389 1000054924 1000054941 1000055114 1000054378 1000055316 1000054377 1000055337 1000054933 1000051367 1000054371 1000054388 1000052267 1000054142 1000055332 1000057169 1000057860 1000054907 (e) PROMATT WITH SIDEWALLS ASSEMBLY, MATERIAL NUMBER PMS3680NZBA; Serial Numbers: PMS3680NZBA PMS3680NZBA PMS3680NZBA PMS3680NZBA PMS3680NZBA PMS3680NZBA PMS3680NZBA PMS3680NZBA PMS3680NZBA (f) PROMATT PLUS 80IN MATRESS, STRETCH, MATERIAL NUMBER PMP3680SZBA; Serial Numbers: PMP3680SZBA PMP3680SZBA PMP3680SZBA PMP3680SZBA PMP3680SZBA PMP3680SZBA PMP3680SZBA PMP3680SZBA PMP3680SZBA PMP3680SZBA PMP3680SZBA (g) PROMATT PLUS 42IN MATRESS, MATERIAL NUMBER PMP4280NZBA; Serial Numbers: PMP4280NZBA PMP4280NZBA PMP4280NZBA PMP4280NZBA PMP4280NZBA PMP4280NZBA PMP4280NZBA PMP4280NZBA PMP4280NZBA PMP4280NZBA PMP4280NZBA PMP4280NZBA PMP4280NZBA…

FDA’s list runs to 2,350 characters. The first part is shown; the rest is on FDA’s record.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide Distribution - US Nationwide in the states of UK, Canada, Holland

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on May 3, 2018.
  • FDA classified it on May 30, 2018.
  • FDA’s Enforcement Report published it on Jun 6, 2018.
  • FDA terminated it on Sep 28, 2023.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Joerns Healthcare, of Arlington, TX.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • FDA’s list of codes runs to 2,350 characters; the first part is shown here, and the rest is on FDA’s record.
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.