Hearth Record

Spacelabs Medical Xprezzon Bedside Monitor

Medical device recall Z-2019-2012. Published in FDA’s Enforcement Report on Jul 25, 2012.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Jul 25, 2012. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display. The parameters return after power cycling the monitor.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhythmia detection or alarms. The Spacelabs Healthcare Xprezzon Bedside Monitor is intended to alert the user to alarm conditions that are reported by Spacelabs Healthcare parameter modules and/or other devices via Flexport interfaces. The patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. The Spacelabs Healthcare Xprezzon Bedside Monitor may also function as a generic display or computer terminal. As a generic display or terminal, the patient monitor allows network based applications to open windows and display information the Xprezzon and other networked monitors. The Spacelabs Healthcare Xprezzon Bedside Monitor is designed to communicate with a variety of external devices such as displays, network devices, serial devices, user input devices, audio systems, and local/remote recorders. The Xprezzon Bedside Monitors is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a hospital environment.

21 units, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

040-1632-00; 1393-000013; 1393-000039; 1393-000040; 1393-000041; 1393-000341; 1393-000382; 1393-000383; 1393-000384; 1393-000385; 1393-000386; 1393-000387; 1393-000388; 1393-000808; 1393-000831; 1393-001374; 1393-001375; 1393-001453; 1393-001990; 1393-001991; and 1393-001992.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide Distribution - USA including Florida, Iowa, Oklahoma, South Caroline and the countries of India, Germany and Panama.

The states and territories these words name

  • Florida
  • Iowa
  • Oklahoma

The words say it went outside the United States too.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Jun 12, 2012.
  • FDA classified it on Jul 16, 2012.
  • FDA’s Enforcement Report published it on Jul 25, 2012.
  • FDA terminated it on Nov 28, 2012.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Spacelabs Healthcare, Llc, of Issaquah, WA.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.