Hearth Record

Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and…

Medical device recall Z-1981-2021. Published in FDA’s Enforcement Report on Jul 7, 2021.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Jul 7, 2021. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems. Brilliance 16 Power, Brilliance 16/10/6 Water/Air 728240/728245/728246 - Brilliance 16 728255/728256/728266 - Brilliance 6 728250/728251 - Brilliance 10 728260 - Extended Brilliance Workspace

1,301 Monitor Bracket Assemblies total, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Model/ Serial Number/UDI: N/A 728240 60005 728240 60003 728240 6017 728240 60012 728240 60007 728240 6105 728240 6136 728240 60016 728240 60014 728240 60006 728240 6156 728240 5080 728240 60011 728240 6109 728240 5413 728245 3371 728245…

Every code shown here, 1,569 characters

Model/ Serial Number/UDI: N/A 728240 60005 728240 60003 728240 6017 728240 60012 728240 60007 728240 6105 728240 6136 728240 60016 728240 60014 728240 60006 728240 6156 728240 5080 728240 60011 728240 6109 728240 5413 728245 3371 728245 3394 728245 3484_ 728245 3500 728245 3377 728245 . 728245 3448 728245 3483 728245 3443 728245 3360 728245 9732 728245 3426 728245 3429 728245 6109 728245 3408 728245 3473 728245 3419 728246 260032 728246 6444 728246 6265 728246 5452 728246 40018 728246 6618 728246 50081 728246 50014 728246 50026 728246 6556 728246 5449 728246 50029 728246 5485 728246 5942 728246 50046 728246 30019 728246 5313 728246 5013 728246 50096 728246 5207 728246 5916 728246 50094 728246 30033 728246 50032 728246 5174 728246 6503 728246 6519 728246 6511 728246 5773 728246 5816 728246 260015 728246 6491 728246 5296 728246 50188 728246 30103 728246 5041 728246 4017 728246 5632 728246 50010 728246 5658 728246 3382 728246 30063 728246 5515 728246 5343 728246 50053 728246 5621 728246 6141 728246 50209 728246 5431 728246 9374 728246 50005 728246 6632 728246 50025 728246 30109 728246 5754 728246 50211 728246 5078 728246 3170 728246 6255 728246 3014 728246 5097 728246 6612 728246 5812 728246 60013 728246 6533 728246 50134 728255 9802 728255 9728 728255 9795 728256 3382 728256 30103 728256 30100 728256 3014 728256 3296 728256 3463 728256 3344 728256 30029 728256 3113 728256 3063 728256 30110 728256 30063 728266 3635 728250 2579 728251 40008 728251 40004 728251 40006 728251 4005 728251 3008 728251 40021 728251 40019 728260 7330 N/A 728260 12323 N/A

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE, AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KG, KH, KR, LA, LB, LK, LT, LY, MM, MN, MU, MV, MY, NI, NL, NO, NZ, OM, PA, PF, PH, PK, PL, PS, PT, QA, RE, RU, RW, SA, SE, SG , SI, SK, TH, TN, TR, UZ, VN, XK, YE, ZA, ZW.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on May 1, 2021.
  • FDA classified it on Jun 30, 2021.
  • FDA’s Enforcement Report published it on Jul 7, 2021.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Philips North America Llc, of Cambridge, MA.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.