Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Product Usage: The…

Medical device recall Z-1958-2019. Published in FDA’s Enforcement Report on Jul 17, 2019.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Jul 17, 2019. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

software malfunction; It was found when a user performs radiography using the wireless flat panel detector (FPD), a message window displays on the monitor stating image transmission was not completed and there was no image available. It also showed the “OK” button to reacquire image data from the FPD, and the “Cancel” button to cancel the acquisition . When the user selects the “OK” button, the same message window appears. This prompted the user to repeat the same operation several times and finally select the “Cancel” button to quit the reacquistion mode.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Product Usage: The RADREX-I is a general purpose x-ray system that employs Solid State Imager(s), SSXI, which converts x-rays directly into electrical signals which can, after appropriate processing be displayed on LCD monitors or printed to a medical grade image printer. The system console is PC based devise that allows for worklist management, image storage, image processing, image exporting and image printing. The system may be equipped with a table and/or vertical wall unit, is configurable with up to two x-ray tubes, and has an auto switching function. This system is intended for use in conjunction with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck and limbs. The system is used for image acquisition, image display and transmission/output or images to external devices.

46, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

No./Serial No./Manufacture Date: No 1./U8A1312001/JANUARY 2013; No.2/U8A1312002/JANUARY 2013; No.3/U8A1342003/APRIL 2013; No.4/U8A1362004 /JUNE 2013; No. 5 U8B13X2006/OCTOBER 2013; No. 6 U8B13Z2008/DECEMBER 2013; No. 7 U8B1412009/JANUARY…

Every code shown here, 1,488 characters

No./Serial No./Manufacture Date: No 1./U8A1312001/JANUARY 2013; No.2/U8A1312002/JANUARY 2013; No.3/U8A1342003/APRIL 2013; No.4/U8A1362004 /JUNE 2013; No. 5 U8B13X2006/OCTOBER 2013; No. 6 U8B13Z2008/DECEMBER 2013; No. 7 U8B1412009/JANUARY 2014; No. 8 U8B1432010 /MARCH 2014; No. 9 U8B1442011/APRIL 2014; No.10 U8B1462013/JUNE 2014; No.11 U8B1472014/JULY 2014; No.12 U8B1482015/AUGUST 2014; No.13 U8B14Y2016/NOVEMBER 2014; No.14 U8B14Z2017/DECEMBER 2014; No.15 U8B14Z2018/DECEMBER 2014; No.16 U8B14Z2019/DECEMBER 2014; No.17 U8B1512020/JANUARY 2015; No.18 U8B1522021/FEBRUARY 2015; No.19 U8B1522022/FEBRUARY 2015; No.20 U8B1522023/FEBRUARY 2015; No.21 U8B1522024/FEBRUARY 2015; No.22 U8B1612025/JANUARY 2016; No.23 U8B1632026/MARCH 2016; No.24 U8B1642027/APRIL 2016; No.25 U8B1642028/APRIL 2016; No.26 U8B1642029/APRIL 2016; No.27 U8B1662030 /JUNE 2016; No.28 U8B1677001/JULY 2016; No.29 U8B16X2032/OCTOBER 2016; No.30 U8B1722033 /FEBRUARY 2017; No.31 U8B1752034/MAY 2017; No.32 U8B1762035/JUNE 2017; No.33 U8B1782036 /AUGUST 2017; No.34 U8B17Z2037 /DECEMBER 2017; No.35 U8C1812038/JANUARY 2018; No.36 USA1462001/JUNE 2014; No.37 USA1492002/SEPTEMBER 2014; No.38 USA14Z2003/DECEMBER 2014; No.39 USA1512004/JANUARY 2015; No.40 USA1532005/MARCH 2015; No.41 USA15Z2006/DECEMBER 2015; No.42 USA1662007/JUNE 2016; No.43 USA1682008/AUGUST 2016; No.44 USA16X2009/OCTOBER 2016; No.45 USA16Y2011/NOVEMBER 2016; No.46 U8A1372005/JULY 2013; No.47 U8B13X2007/OCTOBER 2013; No.48 USA16X2010/OCTOBER 2016

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

US Nationwide Distribution - PA, UT, FL, NY, AR, LA, WV, NJ, OH, MI, WI, TN, TX, VT, GA, NC, IL

The states and territories these words name

17 states and territories.

Every one of them
  • Arkansas
  • Florida
  • Georgia
  • Illinois
  • Louisiana
  • Michigan
  • North Carolina
  • New Jersey
  • New York
  • Ohio
  • Pennsylvania
  • Tennessee
  • Texas
  • Utah
  • Vermont
  • Wisconsin
  • West Virginia

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Oct 16, 2018.
  • FDA classified it on Jul 11, 2019.
  • FDA’s Enforcement Report published it on Jul 17, 2019.
  • FDA terminated it on Jul 20, 2020.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Canon Medical System, USA, INC., of Tustin, CA.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.