Hearth Record

Fixed Core Wire Guide (Safe-T-J Curved), Guidewire Reference

Medical device recall Z-1833-2022. Published in FDA’s Enforcement Report on Oct 5, 2022.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Oct 5, 2022. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Fixed Core Wire Guide (Safe-T-J Curved), Guidewire Reference Part Number/GPN: TSCF-25-180-3 G01602 TSCF-35-180-3 G00902 TSCF-21-260-3 G00455 THSCF-25-260-3 G02384 TSCF-25-145-3 G00466 THSCF-35-80-3 G04069 THSCF-35-180-15 G01151 TSCF-21-50-3 G00456 THSCF-35-145-3 G00412 TSCF-35-145-15 G00509 TSCF-35-145-1.5 G00507 TSCF-35-145-3 G00511 TSCF-38-80-3 G00552 TSCF-18-180-3 G02160 TSCF-35-80-3 G00529 TSCF-21-180-3 G01793 TSCF-25-260-3 G00471 C-SCF-18-50-3 G02050 C-SCF-18-40-2 G02659 C-SCF-25-50-3 G02180 TSCF-18-80-3 G00441 C-SCF-15-30-2 G06973

9,903 US; 5,185 OUS, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

TSCF-25-180-3 G01602 UDI-DI: 00827002016026 Lots/Expiration Dates: 14769750 03-06-2027 14795053 15-06-2027 14818416 29-06-2027 TSCF-35-180-3 G00902 UDI-DI: 00827002009028 Lots/Expiration Dates: 14774744 07-06-2027 14793143 14-06-2027…

Every code shown here, 1,999 characters

TSCF-25-180-3 G01602 UDI-DI: 00827002016026 Lots/Expiration Dates: 14769750 03-06-2027 14795053 15-06-2027 14818416 29-06-2027 TSCF-35-180-3 G00902 UDI-DI: 00827002009028 Lots/Expiration Dates: 14774744 07-06-2027 14793143 14-06-2027 14793144 14-06-2027 14795155 15-06-2027 14795156 15-06-2027 14795157 15-06-2027 14797503 16-06-2027 14798672 16-06-2027 14805743 21-06-2027 14805744 21-06-2027 14816671 28-06-2027 14821171 30-06-2027 14822357 30-06-2027 14822382 30-06-2027 14822383 30-06-2027 TSCF-21-260-3 G00455 UDI-DI: 00827002004559 Lots/Expiration Dates: 14795051 15-06-2027 14813712 27-06-2027 THSCF-25-260-3 G02384 UDI-DI: 00827002023840 Lots/Expiration Dates: 14795058 15-06-2027 TSCF-25-145-3 G00466 UDI-DI: 00827002004665 Lots/Expiration Dates: 14795071 15-06-2027 THSCF-35-80-3 G04069 UDI-DI: 00827002040694 Lots/Expiration Dates: 14797500 16-06-2027 14803099 20-06-2027 14830991 07-07-2027 14832155 07-07-2027 14837668 12-07-2027 14843270 15-07-2027 THSCF-35-180-15 G01151 UDI-DI: 00827002011519 Lots/Expiration Dates: 14798678 16-06-2027 TSCF-21-50-3 G00456 UDI-DI: 00827002004566 Lots/Expiration Dates: 14802358 20-06-2027 THSCF-35-145-3 G00412 UDI-DI: 00827002004122 Lots/Expiration Dates: 14803096 20-06-2027 TSCF-35-145-15 G00509 UDI-DI: 00827002005099 Lots/Expiration Dates: 14803101 20-06-2027 TSCF-35-145-1.5 G00507 UDI-DI: 00827002005075 Lots/Expiration Dates: 14803102 20-06-2027 TSCF-35-145-3 G00511 UDI-DI: 00827002005112 Lots/Expiration Dates: 14803184 20-06-2027 14811921 24-06-2027 14811922 24-06-2027 14811923 24-06-2027 14811924 24-06-2027 14811925 24-06-2027 14815767 27-06-2027 14815768 27-06-2027 14815769 27-06-2027 14815770 27-06-2027 14826198 05-07-2027 14826199 05-07-2027 14826200 05-07-2027 14826201 05-07-2027 14826202 05-07-2027 14831157 07-07-2027 14831158 07-07-2027 14831159 07-07-2027 14831160 07-07-2027 14831162 07-07-2027 14835711 11-07-2027 14835712 11-07-2027 14835713 11-07-2027 14835715 11-07-2027 14835716 11-07-2027 14841791 14-07-2027 14841792…

FDA’s list runs to 3,278 characters. The first part is shown; the rest is on FDA’s record.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Domestic distribution nationwide. Foreign distribution worldwide.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Aug 16, 2022.
  • FDA classified it on Sep 28, 2022.
  • FDA’s Enforcement Report published it on Oct 5, 2022.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Cook Incorporated, of Bloomington, IN.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • FDA’s list of codes runs to 3,278 characters; the first part is shown here, and the rest is on FDA’s record.
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.