PASCAL and PASCAL Streamline
Medical device recall Z-1774-2013. Published in FDA’s Enforcement Report on Jul 31, 2013.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Jul 31, 2013. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Terminated its status, in FDA’s file
- Not named no state, in where it went
- 7 US, 33 foreign how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
When using the “Favorites” function for saved settings, there is potential for a mismatch between displayed power and the actual power output. This may result in over or under treatment.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
PASCAL and PASCAL Streamline (532 nm.577 nm); Treats the patient’s eye.
7 US, 33 foreign, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
PASCAL Standard 2R/G System v2.x.x and PASCAL Streamline 2R/G Software v2.x.x Serial numbers: United States 33057910 33060410 33063610 66078611 66084911 66085111 66085511 Out of United States OMC-179 11138509 11141209 11147209 11147809…
Every code shown here, 492 characters
PASCAL Standard 2R/G System v2.x.x and PASCAL Streamline 2R/G Software v2.x.x Serial numbers: United States 33057910 33060410 33063610 66078611 66084911 66085111 66085511 Out of United States OMC-179 11138509 11141209 11147209 11147809 11148910 11149010 11149210 11149710 . 11149810 11150110 11150210 11150410 11150610 11150810 11150910 11151010 1 151110 11151211 33054310 33059310 33059510 33062810” 66075211’ 66076711 66077111 66078011 66079911 6608391’1 66084611 66084811 66085011 66085211
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide Distribution-USA and the countries of China, Hong Kong, Brazil, France, Japan, Netherlands, Pakistan, Malaysia, United Kingdom, South Korea, Thailand, Dubai, and South Africa.
These words name no state.
The words say it went outside the United States too.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Is it still going on?
Terminated, in FDA’s file.
“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”
openFDA, on “Terminated”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Mar 27, 2012.
- FDA classified it on Jul 19, 2013.
- FDA’s Enforcement Report published it on Jul 31, 2013.
- FDA terminated it on Jul 19, 2013.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Topcon Medical Laser Systems, Inc, of Santa Clara, CA.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.