Hearth Record

The Medtronic StrataMR adjustable valves and shunts are designed to provide continuous cerebrospinal fluid…

Medical device recall Z-1702-2017. Published in FDA’s Enforcement Report on Apr 19, 2017.

  • FDA Class I
FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Apr 19, 2017. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class I, the most serious of FDA’s three classes.

“Dangerous or defective products that predictably could cause serious health problems or death. Examples include: food found to contain botulinum toxin, food with undeclared allergens, a label mix-up on a lifesaving drug, or a defective artificial heart valve.”

openFDA, in plain words, its enforcement report fields.

“Class I is a situation in which there is a reasonable probability that the use of, or exposure to, a violative product will cause serious adverse health consequences or death.”

FDA’s rule, 21 CFR 7.3(m)(1), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Incorrect valve rotor orientation due to misalignment and/or off-center positioning of the StrataMR Locator Tool: When the StrataMR valve is set in the incorrect orientation, higher valve opening pressures result leading to a corresponding underdrainage in CSF fluid performance. This condition may result in the following adverse health consequences: ventriculomegaly, headaches, nausea, vomiting and lethargy. If left untreated, underdrainage has the potential to lead to coma and death.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

The Medtronic StrataMR adjustable valves and shunts are designed to provide continuous cerebrospinal fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The design enables the physician to non-invasive adjust valve pressure/performance level pre- and post-implantation by using magnetic adjustment tools without the need for radiographic confirmation. After surgical implantation, the Medtronic Neurosurgery StrataMR valves provide a non-invasive method to address changing patient needs in the management of hydrocephalus. The valve design enables the physician to adjust the valve pressure/performance level using the Medtronic StrataMR Adjustment Tools. The StrataMR valves are intended to maintain performance level setting under magnetic resonance imaging (MRI) exposure when scanned in accordance with the MRI conditions in the labeling. The StrataMR valves, small and regular, provide the full range of performance levels: 0.5, 1.0, 1.5, 2.0, and 2.5. The currently set performance level is discernible through use of the Indicator Tool or by x-ray (for corresponding settings, refer to figure 6). The dome is designed to allow needle puncture via a 25-gauge or smaller noncoring needle. The StrataMR valve can be flushed in either the distal or proximal direction by percutaneous finger pressure. Occluders are located proximally and distally to the central reservoir on the StrataMR valve regular and only distally on the small model to facilitate selective flushing.

2622 units (US = 1796, outside US =826), in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Worldwide Catalog Number 42955, VALVE 42955 FP-STRATAMR SMALL E03581, E04352, E05120, E05121, E06139, E06476, E09475, E11656, E12556, E12753, E13482, E13483, E13484, E13641, E14518, E14519, E18555, E18556, E18557, E18558, E19670, E19866…

Every code shown here, 827 characters

Worldwide Catalog Number 42955, VALVE 42955 FP-STRATAMR SMALL E03581, E04352, E05120, E05121, E06139, E06476, E09475, E11656, E12556, E12753, E13482, E13483, E13484, E13641, E14518, E14519, E18555, E18556, E18557, E18558, E19670, E19866 Catalog Number 42965, VALVE 42965 FP-STRATAMR REGULAR E03582, E04353, E06015, E06016, E06138, E06412, E09619, E11655, E12041, E12265, E12266, E12950, E12951, E12952, E12953, E13485, E13698, E18559, E18560, E19084, E19540, E19671, E19867, E22208, E22242 Catalog Number 46955, ASSY 46955 FP SHUNT STRATAMR SMALL E12752, E12932, E13573, E16247 Catalog Number 46960, SHUNT 46960 SNAP STRATAMR SMALL E12273, E12780, E12934, E14748 Catalog Number 46965, ASSY 46965 FP SHUNT STRATAMR REGULAR E12779, E12933, E13576, E14270 Catalog Number 46970, SHUNT 46970 SNAP STRATAMR REG E12272, E12620, E12935

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Product was distributed throughout the US and to Australia, Belgium, Canada, Finland, Germany, Hong Kong, Israel, Italy, Japan, New Zealand, Norway, Poland, Portugal, the Russian Federation, Singapore, Spain, Sweden, Switzerland, Turkey, and the UK.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Feb 22, 2017.
  • FDA classified it on Apr 7, 2017.
  • FDA’s Enforcement Report published it on Apr 19, 2017.
  • FDA terminated it on Sep 19, 2017.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Medtronic Neurosurgery, of Goleta, CA.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.