Hearth Record

Artis icono biplane

Medical device recall Z-1647-2022. Published in FDA’s Enforcement Report on Sep 7, 2022.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Sep 7, 2022. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Siemens has become aware of a potential hardware issue with Artis Q, Artis Q.zen, Artis pheno and Artis icono systems with a specific lot of X-ray tubes. In rare cases, if the error detection mechanism fails, it will not be possible to release X-ray any longer until system shutdown.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Artis icono biplane- Model No. 11327600 Artis icono floor- Model No. 11327700

1,389 (590 US; 799 OUS) (Total), in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Artis icono biplane- Model No. 11327600 UDI-DI: 04056869063317 Serial Numbers 180062 180054 180019 180421 180444 180034 180072 180446 180367 180438 180382 180060 180428 180369 180370 180371 180372 180080 180301 180305 180067 180399 180442…

Every code shown here, 1,078 characters

Artis icono biplane- Model No. 11327600 UDI-DI: 04056869063317 Serial Numbers 180062 180054 180019 180421 180444 180034 180072 180446 180367 180438 180382 180060 180428 180369 180370 180371 180372 180080 180301 180305 180067 180399 180442 180040 180379 180386 180360 180333 180352 180407 180433 180412 180400 180397 180432 180316 180447 180363 180429 180084 180418 180334 180018 180376 180354 180046 180434 180357 180358 180361 180362 180081 180423 180411 180414 180381 180021 180443 180309 180356 180378 180393 180368 180413 180323 180439 180445 180430 180043 180437 180049 180377 180077 180055 180014 180037 180373 ***Updated 9/23/22*** 180466 180465 180464 180047 180463 ***Updated 1/9/23*** 180470 180534 180468 180472 180467 180469 180066 180082 Artis icono floor- Model No. 11327700 UDI-DI: 04056869149325 Serial Numbers 170319 170391 170392 170397 170026 170388 170343 170367 170368 170360 170349 170361 170383 170322 170385 170386 170025 170384 170346 170051 170358 170389 170357 170365 170387 170390 170393 170394 170378 170396 170053 170364 ***Updated 1/9/23*** 170402

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide Distribution: US (nationwide) and OUS (Foreign) to countries of: United Arab Emirates; Armenia; Angola; Argentina; Austria; Australia; Bangladesh; Belgium; Bulgaria; Bahrain; Brazil; Canada; Switzerland; China; Colombia; Costa Rica; Czech Republic; Germany; Denmark; Dominican Republic; Algeria; Estonia; Egypt; Spain; Finland; France; United Kingdom; Greece; Guatemala; Hong Kong; Croatia; Inhouse Company; Indonesia; Ireland; Israel; India; Iran; Italy; Jordan; Japan; South Korea; Kuwait; Kazakhstan; Lebanon; Luxembourg; Monaco; Mongolia; Martinique; Mexico; Malaysia; Netherlands; Norway; New Zealand; Oman; Philippines; Poland; Portugal; Qatar; Serbia; Russian Federation; Saudi Arabia; Sweden; Singapore; Slovenia; Slovakia; El Salvador; Thailand; Turkey; Taiwan; Ukraine; Vietnam and South Africa.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Aug 10, 2022.
  • FDA classified it on Aug 26, 2022.
  • FDA’s Enforcement Report published it on Sep 7, 2022.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Siemens Medical Solutions USA, Inc, of Malvern, PA.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.