Hearth Record

Incisive CT- whole-body computed tomography (CT) X-Ray System

Medical device recall Z-1622-2023. Published in FDA’s Enforcement Report on May 31, 2023.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on May 31, 2023. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Potential for Incorrect Image Orientation resulting images may be flipped or reversed result in misdiagnosis, incorrect treatment of a condition, or additional radiation exposure if a rescan is required

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Incisive CT- whole-body computed tomography (CT) X-Ray System Model Number: 728144

7 units US and 76 unit OUS, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

UDI-Di: (01)00884838105508(21) Serial Numbers: 34053, 34092, 34164, 34168, 34179, 34195, 34200 (US) OUS: System Serial Number 34010 34020 34061 34070 34107 34108 34127 34142 34147 34157 34158 34161 34166 34176 34177 34178 34180 34194 34202…

Every code shown here, 638 characters

UDI-Di: (01)00884838105508(21) Serial Numbers: 34053, 34092, 34164, 34168, 34179, 34195, 34200 (US) OUS: System Serial Number 34010 34020 34061 34070 34107 34108 34127 34142 34147 34157 34158 34161 34166 34176 34177 34178 34180 34194 34202 500453 500499 500501 500504 500506 500508 500511 500518 500529 550004 550023 550025 550034 550041 550052 550065 550069 550077 550079 550084 550086 550093 550103 550107 550114 550115 550119 550121 550128 550130 550140 550141 550150 550162 550164 550169 550174 550176 550181 550187 550188 550191 550194 550195 550203 550212 550234 550241 550248 550249 550251 550256 550266 550270 552038 552040 554023

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

AR, CA, OK, TX Foreign: Country Argentina Australia Austria China Colombia Czech Republic Dominican Rep France Germany Guadeloupe Hong Kong India Japan Kuwait Latvia Lithuania Netherlands Norway Portugal Romania Spain Switzerland Taiwan Thailand United Kingdom Vietnam

The states and territories these words name

  • Arkansas
  • California
  • Oklahoma
  • Texas

The words say it went outside the United States too.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Apr 12, 2023.
  • FDA classified it on May 23, 2023.
  • FDA’s Enforcement Report published it on May 31, 2023.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • PHILIPS HEADQUARTERS CAMBRIDGE, of Cambridge, MA.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.