Hearth Record

Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE

Medical device recall Z-1467-2026. Published in FDA’s Enforcement Report on Mar 4, 2026.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Mar 4, 2026. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 EBSI1596 EBSI1781 EBSI1812 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845

31,848, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 UDI-DI (ea) 10193489111538 UDI-DI (case) 20193489111535 Kit lots 2024052090 2024072690 2024100390 2024121290 2025020490 2025041590 2025062790 2025090890 Medline DIALYSIS DRESSING CHANGE KIT SKU…

Every code shown here, 1,351 characters

Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 UDI-DI (ea) 10193489111538 UDI-DI (case) 20193489111535 Kit lots 2024052090 2024072690 2024100390 2024121290 2025020490 2025041590 2025062790 2025090890 Medline DIALYSIS DRESSING CHANGE KIT SKU EBSI1107 UDI-DI (ea) 10889942723508 UDI-DI (case) 40889942723509 Kit lots 2024051690 2024073190 2024092490 2024121190 2025020790 2025032790 2025052990 EBSI1596 UDI-DI (ea) 10653160357162 UDI-DI (case) 00653160357165 Kit lots 2024070190 2024082390 2024102390 2024121790 2025031090 2025040190 2025052790 2025072890 EBSI1781 UDI-DI (ea) 10653160379638 UDI-DI (case) 00653160379631 Kit lots 2024100490 2024123090 EBSI1812 UDI-DI (ea) 10653160386735 UDI-DI (case) 00653160386738 Kit lots 2025032190 2025052990 Centurion DIALYSIS/PHERESUS CVC DRSG CHANGE KIT SKU DT22630 UDI-DI (ea) 10653160345343 UDI-DI (case) 00653160345346 Kit lots 2024050680 2024072980 2024090380 2024111290 2024121190 2025032890 2025061790 2025072390 Medline DIALYSIS/PHERESIS DRESSING CHANGE KIT SKU EBSI1470 UDI-DI (ea) 10193489199307 UDI-DI (case) 20193489199304 Kit lots 2024052080 2024070880 2024110490 2025010290 2025022090 2025030390 2025040990 2025072990 Centurion VASCATH DRESSING CHANGE TRAY SKU DT22845 UDI-DI (ea) 10653160355151 UDI-DI (case) 00653160355154 Kit lots 2024060790 2024080790 2024100990 2024112590 2025042490 2025082090

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

US Nationwide distribution.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Jan 8, 2026.
  • FDA classified it on Feb 24, 2026.
  • FDA’s Enforcement Report published it on Mar 4, 2026.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Medline Industries, LP, of Northfield, IL.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.