Instructions for Use and Patient Manual for HeartWare HVAD System
Medical device recall Z-1426-2021. Published in FDA’s Enforcement Report on Apr 28, 2021.
- FDA Class I
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Apr 28, 2021. Copied from openFDA on Oct 20, 2026.
- Class I FDA’s most serious class (needs attention)
- Ongoing its status, in FDA’s file
- Nationwide FDA’s word for where it went
- 23123 devices (11292 US) how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class I, the most serious of FDA’s three classes.
“Dangerous or defective products that predictably could cause serious health problems or death. Examples include: food found to contain botulinum toxin, food with undeclared allergens, a label mix-up on a lifesaving drug, or a defective artificial heart valve.”
openFDA, in plain words, its enforcement report fields.
“Class I is a situation in which there is a reasonable probability that the use of, or exposure to, a violative product will cause serious adverse health consequences or death.”
FDA’s rule, 21 CFR 7.3(m)(1), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Medtronic HeartWare HVAD System Instructions for Use (IFU) and Patient Manual (PM) updates regarding: Carrying Cases, Driveline Cover Orientation, and Controller Power Up Sequence issues. The PM is being updated to clarify use of the support strap and wear instructions in addition to cleaning instructions of the carrying cases. The IFU and PM are being updated to add a useful life for the carrying cases. The PM is being updated to inform users to keep the driveline cover on when disconnecting and reconnecting the driveline. The IFU and PM are being updated to clarify the power-up sequence that causes the alarm indicator LEDs and both sets of battery LEDs to turn red for 2.5 seconds while the LCD displays the power-on message. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87603 and 87604
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these system components: HVAD Sterile Implant Kit, Model Numbers: 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, MCS1705PU HVAD Controller Kit, Model Numbers: 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1420, 1420JP HVAD AC Adapter Controller, Model Numbers: 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US HVAD DC Adapter Controller: Model Number: 1440 HVAD DATA CABLE 1575- MONITOR, Model Number: 1575 HVAD Battery Pack, Model Numbers: 1650, 1650CA-CLIN, 1650DE HeartWare Patient Pack, Model Number: 1475 HeartWare Waist Pack, Model Numbers: 2050, 2050IL, 2050OUS HeartWare Shoulder Pack, Model Numbers: 2060, 2060IL, 2060OUS The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.
23123 devices (11292 US), in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
All serial numbers
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide distribution: US (nationwide) and countries of: ARGENTINA, Armenia, Australia, AUSTRIA, Bahrain, BELGIUM, Brazil, CANADA, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, FINLAND, France, GERMANY, Greece, Hong Kong, Hungary, INDIA, Israel, Italy, Japan, Kazakhstan, Korea, Kuwait, Latvia, LEBANON, LITHUANIA, Luxembourg, Macedonia, Malaysia, NETHERLANDS, NEW ZEALAND, Norway, POLAND, Qatar, Romania, SAUDI ARABIA, Serbia, Singapore, SLOVAKIA, SOUTH AFRICA Spain, SWEDEN, SWITZERLAND, Taiwan, TURKEY, Ukraine, United Arab, Emirates, United Kingdom, and VIETNAM.
These words name no state.
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Feb 26, 2021.
- FDA classified it on Apr 16, 2021.
- FDA’s Enforcement Report published it on Apr 28, 2021.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Heartware, Inc., of Miami Lakes, FL.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.