Hearth Record

Shimadzu Corporation - Mobile DaRt Evolution This device is a mobile X-ray system in order to take general…

Medical device recall Z-1369-2013. Published in FDA’s Enforcement Report on Jun 12, 2013.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Jun 12, 2013. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

The recall was initiated because Shimadzu Corporation has identified a potential issue in certain manufacturing lots of the Shimadzu MobileDaRt Evolution in combination with CXDI-70C, CXDI-55C, and CXDI-55G. As a result of Shimadzu’s investigation, there is a potential risk for screws becoming loose inside the Shimadzu Mobile X-ray systems.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Shimadzu Corporation - Mobile DaRt Evolution This device is a mobile X-ray system in order to take general radiography of patients in the hospital who cannot move and/or of out patients in emergency .

200 units, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

In combination with CXDI-70C: Serial Numbers: 410001F0A004 410001B13012 410001B13016 410001F0A005 410001B1300E 410001B13010 0162S67303 410001B13018 410001B1300A 410001B13017 410001B1301B 410001B1301A 410001B13019 410001B1301D 410001B13011…

Every code shown here, 1,282 characters

In combination with CXDI-70C: Serial Numbers: 410001F0A004 410001B13012 410001B13016 410001F0A005 410001B1300E 410001B13010 0162S67303 410001B13018 410001B1300A 410001B13017 410001B1301B 410001B1301A 410001B13019 410001B1301D 410001B13011 410001B1301C 410001B13001 410001B13015 410001F0B004 410001B13008 410001B13013 410001F0B005 410001F0B006 410001F0A007 410001B13009 410001B1300B 410001B15035 410001B13014 410001B14009 410001B1301F 410001B14002 410001B14001 410001B14006 410001B14004 410001B15005 410001B13002 410001B1301E 410001B15016 410001B15009 410001B15007 410001B15006 410001B1300F 410001B15001 410001B15017 410001B15002 410001B15020 In combination with CXDI-55C: Serial Numbers: 0162S65805 0162S65810 0162S65808 0162S66001 0162S66008 0162S65809 0162S66004 0162S66005 0162S66009 0162S66010 0162S66201 0162S66202 0162S66207 0162S66208 0162S66206 0162S66302 0162S66402 0162S66409 0162S66410 0162S66403 0162S66501 0162S66405 0162S66407 0162S66406 0162S66505 0162S66502 0162S66503 0162S66408 0162S66504 0162S66603 0162S66508 0162S66507 0162S66510 0162S66709 0162S66601 0162S66610 0162S66609 0162S66605 In combination with CXDI-55G Serial Numbers: 0162S66205 0162S66307 0162S66309 0162S65901 0162S66003 0162S66306 0162S66308 0162S66509 0162S65709 0162S66606 0162S66802 0162S66803

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

US Nationwide Distribution

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Mar 18, 2013.
  • FDA classified it on Jun 3, 2013.
  • FDA’s Enforcement Report published it on Jun 12, 2013.
  • FDA terminated it on Feb 13, 2014.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Shimadzu Medical Systems, of Torrance, CA.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.