5X300 MM Curved Cannula, Arm 1 for use with daVinci SI Surgical System
Medical device recall Z-1332-2014. Published in FDA’s Enforcement Report on Apr 16, 2014.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Apr 16, 2014. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Terminated its status, in FDA’s file
- Not named no state, in where it went
- 5217 total cannulae how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Intuitive Surgical has become aware of the potential for certain lots of the Single-Site 5mm Curved Cannulae to become damaged during use. The second part of this action is to update Single-Site labeling. These labeling changes instruct users how to inspect cannulae prior to use and to appropriately handle cannulae during use.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
5X300 MM Curved Cannula, Arm 1 for use with daVinci SI Surgical System, model number IS3000. Intuitive Surgical 1266 Kifer Road Sunnyvale, CA 94086 The Intuitive Surgical da Vinci Single-Site Instruments and Accessories used with the da Vinci Si Surgical System (IS3000) are indicated for use by trained physicians in an operating room environment for endoscopic manipulation of tissue, grasping, cutting, blunt and sharp dissection, approximation,clip- ligation, electrocautery and suturing during single incision laparoscopic cholecystectomy, benign hysterectomy and salpingo-oophorectomy with the da Vinci Single-Site Instruments and Accessories, including graspers, dissectors, needle drivers, scissors, suction irrigators,monopolar cautery, bipolar cautery, 5mm curved cannulae,5mm and 10 mm straight cannulae, flexible blunt obturators, and the 5mm Single-Site Port.
5217 total cannulae, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
Part number 428061-03: Lot numbers: VE105001 VE120201 VE121301 VE121803 VE122502 VE123106 VE105002 VE120301 VE121304 VE121907 VE122605 VE111201 VE120503 VE121403 VE122002 VE122606 VE114501 VE120703 VE121508 VE122101 VE122902 VE114601 VE121002 VE121605 VE122203 VE123006 VE115201 VE121112 VE121701 VE122401 VE123104.
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide Distribution - USA and the countries of Australia, Belgium, Brazil, Chile, China, Columbia, Denmark, France, Germany, India, Italy, Lebanon, Monaco, Netherlands, Norway, Romania, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom.
These words name no state.
The words say it went outside the United States too.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Is it still going on?
Terminated, in FDA’s file.
“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”
openFDA, on “Terminated”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Mar 13, 2014.
- FDA classified it on Apr 4, 2014.
- FDA’s Enforcement Report published it on Apr 16, 2014.
- FDA terminated it on Sep 3, 2014.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Intuitive Surgical, Inc., of Sunnyvale, CA.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.