DP-30 Digital Ultrasonic Diagnostic Imaging System
Medical device recall Z-1330-2020. Published in FDA’s Enforcement Report on Feb 26, 2020.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Feb 26, 2020. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Terminated its status, in FDA’s file
- Nationwide FDA’s word, and 7 states named
- 53 DP-30 systems and 18 USB’s containing the affected software how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
The DP-30 displays an incorrect needle-guide bracket when used with the 65EC10EA model transducer.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
DP-30 Digital Ultrasonic Diagnostic Imaging System, Part No. 1155E-PA00004 - Product Usage: The Digital Ultrasonic Diagnostic Imaging System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, muscular-skeletal (conventional, superficial), cardiac (adult, pediatric), peripheral vascular and urology exams.
53 DP-30 systems and 18 USB’s containing the affected software, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
DP-30 Ultrasound with software version 03.00.00 and 03.01.00
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
US Nationwide distribution in the states of CA, GA, IL, MD, NY, SC, and WA.
The states and territories these words name
- California
- Georgia
- Illinois
- Maryland
- New York
- South Carolina
- Washington
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Terminated, in FDA’s file.
“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”
openFDA, on “Terminated”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Jan 13, 2020.
- FDA classified it on Feb 20, 2020.
- FDA’s Enforcement Report published it on Feb 26, 2020.
- FDA terminated it on Aug 21, 2020.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Mindray DS USA, Inc. dba Mindray North America, of Mahwah, NJ.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.