GE Healthcare Revolution Apex Elite, System, X-ray, Tomography, Computed
Medical device recall Z-1298-2025. Published in FDA’s Enforcement Report on Mar 19, 2025.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Mar 19, 2025. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Ongoing its status, in FDA’s file
- Not named no state, in where it went
- 250 units how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
GE Healthcare Revolution Apex Elite, System, X-ray, Tomography, Computed
250 units, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 256265STCT, UDI (01)00195278460271(11)230600(21)REV2X2300021CN, Mfg. Lot/Serial Number…
Every code shown here, 1,963 characters
System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number 256265STCT, UDI (01)00195278460271(11)230600(21)REV2X2300021CN, Mfg. Lot/Serial Number REV2X2300021CN; System ID Number 480342APEX1, UDI (01)00195278460271(11)240600(21)REV2X2400065CN, Mfg. Lot/Serial Number REV2X2400065CN; System ID Number 480342APEX3, UDI (01)00195278460271(11)240600(21)REV2X2400067CN, Mfg. Lot/Serial Number REV2X2400067CN; System ID Number 949250APEX, UDI (01)00195278460271(11)230500(21)REV2X2300008CN, Mfg. Lot/Serial Number REV2X2300008CN; System ID Number 858824REV, UDI (01)00195278460271(11)241100(21)REV2X2400147CN, Mfg. Lot/Serial Number REV2X2400147CN; System ID Number 707257APEX1, UDI (01)00195278460271(11)231000(21)REV2X2300077CN, Mfg. Lot/Serial Number REV2X2300077CN; System ID Number 714456APEX, UDI (01)00195278460271(11)231200(21)REV2X2300094CN, Mfg. Lot/Serial Number REV2X2300094CN; System ID Number 719227APEX1, UDI (01)00195278460271(11)240200(21)REV2X2400019CN, Mfg. Lot/Serial Number REV2X2400019CN; System ID Number 970945CT, UDI (01)00195278460271(11)240400(21)REV2X2400046CN, Mfg. Lot/Serial Number REV2X2400046CN; System ID Number 203576EDAPEX, UDI (01)00195278460271(11)240300(21)REV2X2400021CN, Mfg. Lot/Serial Number REV2X2400021CN; System ID Number 860224EDAPEX, UDI (01)00195278460271(11)240500(21)REV2X2400050CN, Mfg. Lot/Serial Number REV2X2400050CN; System ID Number 203276APEX1, UDI (01)00195278460271(11)240800(21)REV2X2400095CN, Mfg. Lot/Serial Number REV2X2400095CN; System ID Number 302744APEX, UDI (01)00195278460271(11)240600(21)REV2X2400078CN, Mfg. Lot/Serial Number REV2X2400078CN; System ID Number 561737APEX1, UDI (01)00195278460271(11)240900(21)REV2X2400114CN, Mfg. Lot/Serial Number REV2X2400114CN; System ID Number 561737APEX2, UDI (01)00195278460271(11)240900(21)REV2X2400112CN, Mfg. Lot/Serial Number REV2X2400112CN; System ID Number 786243APEX1, UDI…
FDA’s list runs to 28,564 characters. The first part is shown; the rest is on FDA’s record.
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide distribution.
These words name no state.
The words say it went outside the United States too.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Feb 18, 2025.
- FDA classified it on Mar 7, 2025.
- FDA’s Enforcement Report published it on Mar 19, 2025.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- GE Medical Systems, LLC, of Waukesha, WI.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- FDA’s list of codes runs to 28,564 characters; the first part is shown here, and the rest is on FDA’s record.
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.