Hearth Record

Omnipod DASH Insulin Management System User Guide (OUS Only) Product

Medical device recall Z-1203-2021. Published in FDA’s Enforcement Report on Mar 17, 2021.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Mar 17, 2021. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Certain foreign user guides include a misprint which could lead to an incorrect bolus delivery and hypoglycemia. Only Outside the US (OUS): Error in the Omnipod DASH User Guide text, page 6 of the User Guide, the text states that the comma (,) is not used for the decimal on the PDM, when in fact it is the period (.) that is not used.If a user were to use the period as a decimal separator when entering a bolus value, this decimal separator would not be recognized by the device and any numbers entered after the period could result in a larger value than intended

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Omnipod DASH Insulin Management System User Guide (OUS Only) Product Catalog Number: PT 000002 XXX XXX XX PT 000318 PT 000319 Distributed to: Canada (French only), the Netherlands, France, Austria, Germany, Belgium, Greece, Turkey, Norway, Sweden, Denmark, and Finland

21,723 guides, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Part Number/Rev./Country: PT 000002 AUT DEU mg 1, 2, 3 Guide, Dash User, German Austria PT 000002 BEL DEU MG 1, 2, 3 Guide, Dash User Guide, Belgium Germany, mgdl PT 000002 BEL FRE MG 1, 2, 3, 4 Guide, Dash User Guide, Belgium French, mgdl…

Every code shown here, 1,017 characters

Part Number/Rev./Country: PT 000002 AUT DEU mg 1, 2, 3 Guide, Dash User, German Austria PT 000002 BEL DEU MG 1, 2, 3 Guide, Dash User Guide, Belgium Germany, mgdl PT 000002 BEL FRE MG 1, 2, 3, 4 Guide, Dash User Guide, Belgium French, mgdl PT 000002 BEL NLD MG 1, 2, 3 Guide, Dash User Guide, Belgium Dutch, mgdl PT 000002 DEU DEU mg 1, 2, 3 Guide, Dash User Guide, German Germany, mgdl PT 000002 DEU DEU mm 1, 2, 3 Guide, Dash User Guide, German Germany, mmoll PT 000002 DNK DANmm 1, 2, 3 Guide, Dash User Guide, Danish Denmark PT 000002 FIN FIN mm 1, 2, 3 Guide, Dash User Guide, Finnish Finland PT 000002 FRA FRE mg 1, 2, 3, 4 Guide, Dash User Guide, French France PT 000002 NLD NLD mm 1, 2, 3, 4 Guide, Dash User Guide, Dutch Netherland PT 000002 NOR NORmm 1, 2, 3 Guide, Dash User Guide, Norwegian Norway PT 000002 SWE SWEmm 1, 2, 3 Guide, Dash User Guide, Swedish Sweden 41333 CAN 1, 2 User Guide, Omnipod, Dash, Canada PT 000318 1, 2 Guide, Dash User Guide, Greece PT 000319 1, 2 Guide, Dash User Guide, Turkey

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Canada (French only), the Netherlands, France, Austria, Germany, Belgium, Greece, Turkey, Norway, Sweden, Denmark, and Finland

These words name no state.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Feb 1, 2021.
  • FDA classified it on Mar 10, 2021.
  • FDA’s Enforcement Report published it on Mar 17, 2021.
  • FDA terminated it on Feb 8, 2023.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Insulet Corporation, of Acton, MA.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: E-Mail.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.