Invacare TDX SP2
Medical device recall Z-1130-2022. Published in FDA’s Enforcement Report on Jun 1, 2022.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Jun 1, 2022. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Ongoing its status, in FDA’s file
- Nationwide FDA’s word, and 35 states named
- 133 devices how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
The battery wiring harness on affected wheelchairs may become disconnected during shipment from the factory or during wheelchair use traversing large obstacles or after a jolt or impact to the wheelchair. The disconnection of the battery harness results in loss of power to the wheelchair.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
Invacare TDX SP2, Model Nos. TDXSP2 TDXSP2-MCG TDXSP2V
133 devices, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
UDI (DI): Model Number Device Identifier TDXSP2 00841447100911 TDXSP2-MCG 00841447100966 TDXSP2V 00841447101512 Serial Numbers: 22CE000031 22BE004288 22BE002669 22BE003672 22BE004782 22BE004151 22CE000984 22BE004286 22BE004171 22BE004540…
Every code shown here, 1,590 characters
UDI (DI): Model Number Device Identifier TDXSP2 00841447100911 TDXSP2-MCG 00841447100966 TDXSP2V 00841447101512 Serial Numbers: 22CE000031 22BE004288 22BE002669 22BE003672 22BE004782 22BE004151 22CE000984 22BE004286 22BE004171 22BE004540 22CE001086 22BE003661 22CE000184 22CE000878 22CE001449 22CE000032 22CE001421 22BE004933 22CE001002 22CE000180 22CE000725 22BE004544 22BE005009 22BE003671 22BE004284 22BE004159 22BE004285 22BE002652 22BE004282 22BE004798 22BE004281 22BE004170 22CE000182 22CE000183 22CE000876 22CE000873 22CE000981 22CE000870 22CE000871 22CE001248 22CE001004 22CE001084 22BE004776 22BE004165 22BE004931 22BE005010 22CE000875 22CE000728 22CE000727 22BE004910 22BE004904 22BE005007 22BE004160 22BE004169 22BE004175 22BE004166 22BE004905 22CE001420 22CE000496 22BE003645 22BE002651 22CE000999 22CE001078 22BE004545 22BE004763 22BE003662 22BE004929 22BE004937 22BE004176 22BE004903 22CE000181 22BE004167 22BE005006 22BE002666 22BE004177 22BE003962 22BE003957 22BE004173 22BE003664 22BE003659 22BE004767 22BE004940 22BE004777 22BE004538 22CE000186 22CE000867 22BE002935 22BE003959 22CE000493 22CE001077 22BE002934 22CE000726 22BE005004 22CE001251 22BE003961 22BE003673 22BE004902 22BE002936 22BE004801 22BE004287 22BE004995 22BE004539 22CE001075 22CE000490 22CE000879 22BE004168 22CE000847 22CE000033 22CE001005 22CE001006 22BE004765 22BE004768 22BE004769 22BE004770 22BE004911 22BE002932 22BE003960 22CE000874 22CE001253 22BE004930 22CE000178 22CE000179 22CE001080 22BE004174 22BE003963 22BE004935 22BE004775 22BE004172 22CE000177 22BE004543 22BE004764 22BE004766 22CE001076
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide distribution - US Nationwide distribution in the states of AK, AR, CA, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA, and WI. The country of Canada.
The states and territories these words name
35 states and territories.
Every one of them
- Alaska
- Arkansas
- California
- Connecticut
- Florida
- Georgia
- Hawaii
- Iowa
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- New Jersey
- New York
- Ohio
- Oregon
- Pennsylvania
- South Carolina
- Tennessee
- Texas
- Virginia
- Vermont
- Washington
- Wisconsin
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Mar 30, 2022.
- FDA classified it on May 25, 2022.
- FDA’s Enforcement Report published it on Jun 1, 2022.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Invacare Corporation, of Elyria, OH.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.