Medline Kits, trays, and packs labeled as follows: a) LACERATION KIT
Medical device recall Z-1112-2024. Published in FDA’s Enforcement Report on Mar 6, 2024.
- FDA Class I
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Mar 6, 2024. Copied from openFDA on Oct 24, 2026.
- Class I FDA’s most serious class (needs attention)
- Ongoing its status, in FDA’s file
- Nationwide FDA’s word, and 1 state named
- 21154 units how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class I, the most serious of FDA’s three classes.
“Dangerous or defective products that predictably could cause serious health problems or death. Examples include: food found to contain botulinum toxin, food with undeclared allergens, a label mix-up on a lifesaving drug, or a defective artificial heart valve.”
openFDA, in plain words, its enforcement report fields.
“Class I is a situation in which there is a reasonable probability that the use of, or exposure to, a violative product will cause serious adverse health consequences or death.”
FDA’s rule, 21 CFR 7.3(m)(1), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
FDA, Device recall enforcement reports. Released Oct 23, 2026. Updated weekly.
Which ones are recalled?
Medline Kits, trays, and packs labeled as follows: a) LACERATION KIT, REF SUT13935; b) LACERATION TRAY, REF SUT21215; c) LACERATION TRAY, REF SUT15380; d) LACERATION TRAY W/SNAGFREE NEE, REF SUT19005; e) PLASTICS SUTURE TRAY, REF SUT13535; f) UVC INSERTION TRAY, REF UVT840A g) BRONCHOALVEOLAR LAVAGE PACK, REF DYNDA2369A; h) CENTRAL LINE NEEDLELESS CONNECTOR KIT, REF DYNDC2869B; i) CVC STERILE CAP CHANGE KIT, REF DT20900A; j) DOUBLE VALVE CHANGE KIT, REF DYNDA2660A; k) KIDNEY BIOPSY PACK, REF DYNDH1650; l) PORT ACCESS INFUSION KIT - 0.75IN NEEDLE, REF DYNDC2816C; m) PORT ACCESS KIT, REF EBSI1302A; n) SHORT TERM INFUSAPORT ACCESS KIT, REF DYNDC3188; o) SINGLE VALVE CHANGE KIT, REF DYNDA2659A; p) SUTURE TRAY, REF DYNDS1085A;
21154 units, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 23, 2026. Updated weekly.
Codes to check against yours
a) REF SUT13935, UDI/DI 40193489440714 (case), 10193489440713 (unit), Lot Numbers: 23GBO782, 23GBP869, 23HBA879, 23HBX374; b) REF SUT21215, UDI/DI 653160325645 (case), 10653160325642 (unit), Lot Numbers: 22EBB078, 22FBE772, 22GBJ917,…
Every code shown here, 1,971 characters
a) REF SUT13935, UDI/DI 40193489440714 (case), 10193489440713 (unit), Lot Numbers: 23GBO782, 23GBP869, 23HBA879, 23HBX374; b) REF SUT21215, UDI/DI 653160325645 (case), 10653160325642 (unit), Lot Numbers: 22EBB078, 22FBE772, 22GBJ917, 22HBN681, 22HBW656, 22JBB835, 23ABL668, 23FBT868, 23GBK697, 23HBU987; c) REF SUT15380, UDI/DI 10653160203698 (case), 653160203691 (unit), Lot Numbers: 22EBI572, 22FBK706, 22GBI140, 22GBZ790, 22JBT097, 22KBW287, 22LBS262, 23DBB365, 23EBB439, 23EBS520, 23FBV416, 23HBV213; d) REF SUT19005, UDI/DI 653160271454 (case), 10653160271451 (unit), Lot Numbers: 23CBV894, 23EBM160, 23FBI594, 23GBK013, 23HBW722; e) REF SUT13535, UDI/DI 40193489440271 (case), 10193489440270 (unit), Lot Numbers: 22FBG265, 22FBU120, 22HBK650, 22LBS268, 23CBH315, 23EBH269, 23HBU983; f) REF UVT840A, UDI/DI 653160292008 (case), 10653160292005 (unit), Lot Numbers: 22ABH461, 22ABJ736; g) REF DYNDA2369A, UDI/DI 00653160352160 (case), 10653160352167 (unit), Lot Numbers: 2022051050, 2022072750; h) REF DYNDC2869B, UDI/DI 20193489111719 (case), 10193489111712 (unit), Lot Numbers: 2021121750, 2022010450; i) REF DT20900A, UDI/DI 00653160353211 (case), 10653160353218 (unit), Lot Numbers: 2022061380, 2022061480, 2022082280; j) REF DYNDA2660A, UDI/DI 00653160351897 (case), 10653160351894 (unit), Lot Numbers: 2022042580, 2022050280, 2022082280; k) REF DYNDH1650, UDI/DI 20193489110866 (case), 10193489110869 (unit), Lot Numbers: 2022012050; l) REF DYNDC2816C, UDI/DI 00653160348309 (case), 10653160348306 (unit), Lot Numbers: 2022041880; m) REF EBSI1302A, UDI/DI 20193489111894 (case), 10193489111897 (unit), Lot Numbers:2022020250, 2022022880, 2022082280; n) REF DYNDC3188, UDI/DI 00653160348828 (case), 10653160348825 (unit), Lot Numbers: 2022012450; o) REF DYNDA2659A, UDI/DI 00653160351880 (case), 10653160351887 (unit), Lot Numbers: 2022042850; p) REF DYNDS1085A, UDI/DI 40080196778691 (case), 10080196778690 (unit), Lot Numbers: 2022041150, 2022051350, 2022053150
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama
The states and territories these words name
- Puerto Rico
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Nov 15, 2023.
- FDA classified it on Feb 24, 2024.
- FDA’s Enforcement Report published it on Mar 6, 2024.
FDA, Device recall enforcement reports. Released Oct 23, 2026. Updated weekly.
Who is recalling it?
- MEDLINE INDUSTRIES, LP - Northfield, of Northfield, IL.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 23, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 23, 2026 and checked Oct 23, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 24, 2026.