Hearth Record

Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE

Medical device recall Z-1089-2026. Published in FDA’s Enforcement Report on Jan 21, 2026.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Jan 21, 2026. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Medline kits containing Stryker’s Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Medline medical procedure convenience kits labeled as: 1) LAVH PROCEDURE, REF CDS980754T; 2) GYN LAP HYSTERECTOMY CDS, REF CDS982662Q; 3) DAVINCI HYSTERECTOMY CDS, REF CDS985344F; 4) LAVH, REF DYNJQ9041R; 5) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ900244I; 6) ROBOTIC LAVH, REF DYNJ901281I; 7) LAVH-LF, REF DYNJ901829T; 8) LEX LAP HYST, REF DYNJ902030J; 9) LITHOTOMY-SLINGS-LF, REF DYNJ905274D; 10) GYN LAPAROSCOPY, REF DYNJ905485J; 11) ROBOTIC HYSTERECTOMY, REF DYNJ906380C; 12) LAPAROSCOPIC HYSTERECTOMY, REF DYNJ910634; 13) LAVH GYN/ONC, REF DYNJ910927.

1457 units, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

REF CDS980754T: UDI/DI 10198459121814 (EA) 40198459121815 (CS), Lot Number 25EMD795; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25EBR237; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot…

Every code shown here, 1,991 characters

REF CDS980754T: UDI/DI 10198459121814 (EA) 40198459121815 (CS), Lot Number 25EMD795; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25EBR237; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25EBP591; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25DBM353; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25CBL184; REF CDS982662Q: UDI/DI 10195327306793 (EA) 40195327306794 (CS), Lot Number 25BBL633; REF CDS985344F: UDI/DI 10195327312565 (EA) 40195327312566 (CS), Lot Number 25FBK918; REF CDS985344F: UDI/DI 10195327312565 (EA) 40195327312566 (CS), Lot Number 25CBJ496; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25FBF371; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25EBK493; REF DYNJQ9041R: UDI/DI 10198459043086 (EA) 40198459043087 (CS), Lot Number 25DBH895; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25DBC272; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25CBU642; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25CBJ298; REF DYNJ900244I: UDI/DI 10195327464929 (EA) 40195327464920 (CS), Lot Number 25BBU206; REF DYNJ901281I: UDI/DI 10195327699154 (EA) 40195327699155 (CS), Lot Number 25FBA261; REF DYNJ901281I: UDI/DI 10195327699154 (EA) 40195327699155 (CS), Lot Number 25CBN230; REF DYNJ901829T: UDI/DI 10198459231384 (EA) 40198459231385 (CS), Lot Number 25FDB398; REF DYNJ901829T: UDI/DI 10198459231384 (EA) 40198459231385 (CS), Lot Number 25DDA191; REF DYNJ901829T: UDI/DI 10198459231384 (EA) 40198459231385 (CS), Lot Number 25BDB095; REF DYNJ902030J: UDI/DI 10195327391249 (EA) 40195327391240 (CS), Lot Number 25DBJ371; REF DYNJ905274D: UDI/DI 10195327393014 (EA) 40195327393015 (CS), Lot Number 25BBS130; REF DYNJ905485J: UDI/DI 10198459202025 (EA) 40198459202026 (CS), Lot Number 25FBJ689; REF DYNJ905485J: UDI/DI…

FDA’s list runs to 2,737 characters. The first part is shown; the rest is on FDA’s record.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

US Nationwide distribution.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Dec 16, 2025.
  • FDA classified it on Jan 14, 2026.
  • FDA’s Enforcement Report published it on Jan 21, 2026.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Medline Industries, LP, of Northfield, IL.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • FDA’s list of codes runs to 2,737 characters; the first part is shown here, and the rest is on FDA’s record.
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.