iCast Covered Stent System
Medical device recall Z-1077-2022. Published in FDA’s Enforcement Report on May 25, 2022.
- FDA Class I
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on May 25, 2022. Copied from openFDA on Oct 20, 2026.
- Class I FDA’s most serious class (needs attention)
- Ongoing its status, in FDA’s file
- Not named no state, in where it went
- 68812 units how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class I, the most serious of FDA’s three classes.
“Dangerous or defective products that predictably could cause serious health problems or death. Examples include: food found to contain botulinum toxin, food with undeclared allergens, a label mix-up on a lifesaving drug, or a defective artificial heart valve.”
openFDA, in plain words, its enforcement report fields.
“Class I is a situation in which there is a reasonable probability that the use of, or exposure to, a violative product will cause serious adverse health consequences or death.”
FDA’s rule, 21 CFR 7.3(m)(1), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Firm has observed increased rate of customer complaints related to separation of the balloon or catheter hub from the delivery catheter during delivery system withdrawal causing procedural delay and possible administration of additional anesthesia and contrast which may impact patients’ renal function. Potential harms include occlusion or embolism and associated response, with specific outcomes such as amputation, embolism, loss of organ function, organ infarction, or tissue infarction. ***Updated June 2023*** Additional surgical stress caused by prolonged interventional surgery has the potential to lead to myocardial infarction or death.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
iCast Covered Stent System Product Code/Part Number: 85400 85401 85402 85403 85404 85405 85406 85407 85408 85409 85410 85411 85412 85413 85414 85415 85416 85417 85418 85419 85420 85424 85440 85441 85442 85443 85444 85445 85450 85451 85452 85453 85454 85455 Product Code/UDI /Description: 85400 00650862854008 iCAST COVERED STENT,5MMX38MMX80CM 85401 00650862854015 iCAST COVERED STENT,5MMX59MMX80CM 85402 00650862854022 iCAST COVERED STENT,6MMX38MMX80CM 85403 00650862854039 iCAST COVERED STENT,6MMX59MMX80CM 85404 00650862854046 iCAST COVERED STENT,7MMX38MMX80CM 85405 00650862854053 iCAST COVERED STENT,7MMX59MMX80CM 85406 00650862854060 iCAST COVERED STENT,8MMX38MMX80CM 85407 00650862854077 iCAST COVERED STENT,8MMX59MMX80CM 85408 00650862854084 iCAST COVERED STENT,9MMX38MMX80CM 85409 00650862854091 iCAST COVERED STENT,9MMX59MMX80CM 85410 00650862854107 iCAST COVERED STENT,5MMX38MMX120CM 85411 00650862854114 iCAST COVERED STENT,5MMX59MMX120CM 85412 iCAST COVERED STENT,6MMX38MMX120CM 85413 00650862854138 iCAST COVERED STENT,6MMX59MMX120CM 85414 00650862854145 iCAST COVERED STENT,7MMX38MMX120CM 85415 00650862854152 iCAST COVERED STENT,7MMX59MMX120CM 85416 00650862854169 iCAST COVERED STENT,8MMX38MMX120CM 85417 00650862854176 iCAST COVERED STENT,8MMX59MMX120CM 85418 00650862854183 iCAST COVERED STENT,9MMX38MMX120CM 85419 00650862854190 iCAST COVERED STENT,9MMX59MMX120CM 85420 00650862854206 iCAST COVERED STENT,10MMX38MMX80CM 85424 00650862854244 iCAST COVERED STENT,10MMX38MMX120CM 85440 00650862854404 iCAST COVERED STENT,5MMX16MMX80CM 85441 00650862854411 iCAST COVERED STENT,5MMX22MMX80CM 85442 00650862854428 iCAST COVERED STENT 6MMX16MMX80CM 85443 00650862854435 iCAST COVERED STENT,6MMX22MMX80CM 85444 00650862854442 iCAST COVERED STENT,7MMX16MMX80CM 85445 00650862854459 iCAST COVERED STENT,7MMX22MMX80CM 85450 00650862854503 iCAST COVERED STENT,5MMX16MMX120CM 85451 00650862854510 iCAST COVERED STENT,5MMX22MMX120CM 85452 00650862854527 iCAST COVERED STENT,6MMX16MMX120CM 85453 00650862854534 iCAST COVERED STENT,6MMX22MMX120CM 85454 00650862854541 iCAST COVERED STENT,7MMX16MMX120CM 85455 00650862854558 iCAST COVERED STENT,7MMX22MMX120CM
68812 units, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
All lots
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Foreign: Australia Austria Bahrain Belgium Bosnia-Herzegovina Brazil Bulgaria Canada Chile Colombia Costa Rica Cyprus Czech Republic Denmark Ecuador Egypt El Salvador Estonia Finland France Germany Great Britain Greece Hong Kong Hungary Iceland Ireland, Republic of Israel Italy Japan Jordan Kuwait Latvia Lebanon Lithuania Macedonia Malaysia Malta Mexico Netherlands New Zealand Norway Panama Paraguay Peru Philippines Poland Portugal Qatar Reunion Russia San Marino Saudi Arabia Singapore Slovak Republic Slovenia South African Republic South Korea Spain Sweden Switzerland Taiwan, Thailand Trinidad and Tobago Turkey United Arab Emirates Uruguay Vietnam
These words name no state.
The words say it went outside the United States too.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Mar 3, 2022.
- FDA classified it on May 18, 2022.
- FDA’s Enforcement Report published it on May 25, 2022.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Atrium Medical Corporation, of Merrimack, NH.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: E-Mail.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.