Hearth Record

PERI SCR. 3.5MM X 10MM PERI SCR. 3.5MM X 12MM PERI SCR. 3.5MM X 14MM PERI SCR 3.5MM X 44MM STE PERI SCR.…

Medical device recall Z-1040-2014. Published in FDA’s Enforcement Report on Feb 26, 2014.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Feb 26, 2014. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Zimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the Building 1 packaging operation at this facility were not properly validated. As a result, Zimmer conducted seal integrity testing to evaluate if previously packaged product from this facility met the established requirements for sterile barrier integrity. The acceptance criteria established for this testing was 99% reliability with a 90% confidence interval. These packaging tray configurations met the acceptance criteria for sterile barrier integrity of both the inner and outer trays, however, Zimmer has decided to recall all manufacturing lots that have not expired and that were sterile packaged in the ZMBV facility Building 1 packaging operations.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

PERI SCR. 3.5MM X 10MM PERI SCR. 3.5MM X 12MM PERI SCR. 3.5MM X 14MM PERI SCR 3.5MM X 44MM STE PERI SCR. 3.5MM X 65MM PERI. SCR 4.0MM X24MM PERI. SCR 4.0MM X40MM PERI SCR. 4.0MMX14MM FULL PERI SCR. 4.0MMX32MM FULL PERI SCR. 4.0MMX38MM FULL 3.5MM CORT SCR X 10MM 3.5MM CORT SCR X 12MM 3.5MM CORT SCR X 14MM 3.5MM CORT SCR X 16MM 3.5MM CORT SCR X 18MM 3.5MM CORT SCR X 20MM 3.5MM CORT SCR X 22MM 3.5MM CORT SCR X 24MM 3.5MM CORT SCR X 26MM 3.5MM CORT SCR X 28MM 3.5MM CORT SCR X 30MM 3.5MM CORT SCR X 32MM 3.5MM CORT SCR X 34MM 3.5MM CORT SCR X 36MM 3.5MM CORT SCR X 38MM 3.5MM CORT SCR X 40MM 3.5MM CORT SCR X 42MM 3.5MM CORT SCR X 44MM 3.5MM CORT SCR X 46MM 3.5MM CORT SCR X 48MM 3.5MM CORT SCR X 50MM 3.5MM CORT SCR X 52MM 3.5MM CORT SCR X 54MM 3.5MM CORT SCR X 56MM 3.5MM CORT SCR X 58MM 3.5MM CORT SCR X 60MM 3.5MM CORT SCR X 65MM 3.5MM CORT SCR X 70MM 3.5MM CORT SCR X 75MM 3.5MM CORT SCR X 80MM 3.5MM CORT SCR X 85MM 3.5MM CORT SCR X 90MM 3.5MM CORT SCR X 100MM 3.5MM CORT SCR X 105MM 3.5MM CORT SCR X 110MM CANC SCR 4.0MM X 10MM CANC SCR 4.0MM X 10MM FUL CANC SCR 4.0MM X 12MM CANC SCR 4.0MM X 12MM FUL CANC SCR 4.0MM X 14MM CANC SCR 4.0MM X 14MM FUL CANC SCR 4.0MM X 16MM CANC SCR 4.0MM X 16MM FUL CANC SCR 4.0MM X 18MM CANC SCR 4.0MM X 18MM FUL CANC SCR 4.0MM X 20MM CANC SCR 4.0MM X 20MM FUL CANC SCR 4.0MM X 22MM CANC SCR 4.0MM X 22MM FUL CANC SCR 4.0MM X 24MM CANC SCR 4.0MM X 24MM FUL CANC SCR 4.0MM X 26MM CANC SCR 4.0MM X 26MM FUL CANC SCR 4.0MM X 28MM CANC SCR 4.0MM X 28MM FUL CANC SCR 4.0MM X 30MM CANC SCR 4.0MM X 30MM FUL CANC SCR 4.0MM X 32MM CANC SCR 4.0MM X 32MM FUL CANC SCR 4.0MM X 34MM CANC SCR 4.0MM X 34MM FUL CANC SCR 4.0MM X 36MM CANC SCR 4.0MM X 36MM FUL CANC SCR 4.0MM X 38MM CANC SCR 4.0MM X 38MM FUL CANC SCR 4.0MM X 40MM CANC SCR 4.0MM X 40MM FUL

7,044,680 total, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

61337161 61251498 61251499 61220432 61553288 61966252 61966257 60856802 60871924 60871946 60837236 61168818 61168821 61251340 61168823 61251341 61168825 61216588 61247526 61168827 61216589 61247527 61168829 61216590 61247528 61168830…

Every code shown here, 1,043 characters

61337161 61251498 61251499 61220432 61553288 61966252 61966257 60856802 60871924 60871946 60837236 61168818 61168821 61251340 61168823 61251341 61168825 61216588 61247526 61168827 61216589 61247527 61168829 61216590 61247528 61168830 61216591 61247529 61168831 61247530 61168833 61216593 61247531 61168835 61247532 61168837 61216596 61247533 61168839 61168841 61168842 61168843 61168845 61247536 61168847 61223705 61168849 61168850 61168852 61168854 61168856 61168858 61168859 61168860 61168861 61168862 61567437 61411579 61678383 61678382 61835910 61835909 61214649 61216599 61214650 61216602 61251346 61216605 61591005 61190532 61190535 61740546 61190538 61190539 61401997 61190540 61190543 61560426 61823977 61190544 61190546 61740547 61190547 61190549 61740548 61190551 61190552 61560430 61190554 61190555 61560431 61190618 61190619 61740550 61190620 61190627 61560427 61190628 61190629 61401998 61190630 61190632 61740551 61190633 61190634 61740552 61190635 61190636 61740553 61190638 61190639 61190640 61190641 61190642 61740554 61190643

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide Distribution-USA (nationwide) and the countries of Belgium, Canada, Canary Islands, China, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Korea, Luxemburg, Netherlands, New Zealand, Norway, Portugal Reunion, Russian Federation, Singapore, Slovakia, Spain, Sweden, Switzerland, Turkey, Taiwan, Thailand, and United Kingdom.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Oct 11, 2013.
  • FDA classified it on Feb 20, 2014.
  • FDA’s Enforcement Report published it on Feb 26, 2014.
  • FDA terminated it on Apr 14, 2017.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Zimmer, Inc., of Warsaw, IN.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.