Hearth Record

GE Healthcare Infinia I

Medical device recall Z-0888-2025. Published in FDA’s Enforcement Report on Jan 22, 2025.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Jan 22, 2025. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

GE Healthcare Infinia I, Model/Catalog Numbers: 1) 5178728, 2) H3000WE, 3) H3000WM, 4) H3000YG; System, Tomography, Computed, Emission

36 units, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

1) 5178728: UDI/DI N/A, System ID YN1200, Serial Number: 16179; System ID YN1650, Serial Number: 17339. 2) H3000WE: UDI/DI N/A, System ID 82426060010, Serial Number Not Available; System ID 82426080021, Serial Number Not Available; System…

Every code shown here, 1,778 characters

1) 5178728: UDI/DI N/A, System ID YN1200, Serial Number: 16179; System ID YN1650, Serial Number: 17339. 2) H3000WE: UDI/DI N/A, System ID 82426060010, Serial Number Not Available; System ID 82426080021, Serial Number Not Available; System ID 82426090035, Serial Number 18605; System ID 82426090037, Serial Number 18604; System ID 82426100057, Serial Number 19191; System ID 82426150036, Serial Number 18624; System ID 82426150038, Serial Number Not Available; System ID 82426180015, Serial Number 18611; System ID 082426010021, Serial Number 18644; System ID 082426030032, Serial Number 18627; System ID 082426050012, Serial Number 18629; System ID 082426070041, Serial Number 18639; System ID 082426090044, Serial Number 18652; System ID 082426100067, Serial Number 18648; System ID 082426100069, Serial Number 18641; System ID 082426120031, Serial Number 18656; System ID 082426120033, Serial Number 18650; System ID 082426140029, Serial Number 18628; System ID 082426140030, Serial Number 18647; System ID 082426150039, Serial Number 18651; System ID 082426150040, Serial Number 18657; System ID 082426160040, Serial Number 18442; System ID 082426160054, Serial Number 18645; System ID 082426170004, Serial Number 18649; System ID 082426190031, Serial Number 18646; System ID 082426210028, Serial Number 18655; System ID 082426230028, Serial Number 18653; System ID 082426250006, Serial Number 18637; System ID 671USNHINF, Serial Number 17799; System ID A5128757, Serial Number 17566; System ID 713794INF, Serial Number 17219; System ID 920794inf, Serial Number 16364. 3) H3000WM: UDI/DI N/A, System ID NM16353, Serial Number 16353; System ID YN1540, Serial Number 16335; System ID YN1555, Serial Number 16315. 4) H3000YG: UDI/DI N/A, System ID A5172908, Serial Number 16143.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide

These words name no state.

The words say it went outside the United States too.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Dec 18, 2024.
  • FDA classified it on Jan 10, 2025.
  • FDA’s Enforcement Report published it on Jan 22, 2025.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING, of Tirat Carmel, Israel.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.