Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX¿ Complete…
Medical device recall Z-0831-2015. Published in FDA’s Enforcement Report on Dec 24, 2014.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Dec 24, 2014. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Terminated its status, in FDA’s file
- Nationwide FDA’s word for where it went
- 2,506 units how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX¿ Complete Brain Probe Kit (comprised of IP1, CC1.P1) 2) LICOX¿ Complete Brain Probe Kit (comprised of IP2, CC1.P1) 3) LICOX¿ Complete Brain Tunnelling Probe Kit (comprised of VK5.2, CC1.P1 4) LICOX¿ Complete Brain Probe Kit (comprised of IM1, CC1.SB) Sterile Rx Only Manufacturer CMS -Gesse;schaft for medizinische Sondentechnik mbH Dorfstrasse 2 24247 <eilkendorf Germany Distributed by Integra NeuroSciences 311 Enterprise Drive Plainsboro, NJ 08536, USA
2,506 units, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
CC1 Licox¿ PO2 Microprobe, 200 mm 021213, 040414, 070314, 130114, 240214, 270114, 280314, 300614 CC12 Licox¿ PO2 Microprobe, 300 mm 020614, 040814, 050514, 070414, 070714, 220414 CC1.P1 Licox¿ PMO Combined Probe Containing Oxygen and…
Every code shown here, 881 characters
CC1 Licox¿ PO2 Microprobe, 200 mm 021213, 040414, 070314, 130114, 240214, 270114, 280314, 300614 CC12 Licox¿ PO2 Microprobe, 300 mm 020614, 040814, 050514, 070414, 070714, 220414 CC1.P1 Licox¿ PMO Combined Probe Containing Oxygen and Temperature 050614, 060214, 060314, 100414, 100714, 130214, 130314, 170114, 190614, 210814, 260614, 310714 IM3ST Licox¿ Complete Brain IMC-Probe Kit: 020614, 030314, 110714, 130114, 140314, 160614, 180714, 200114, 280314 IP1.P Licox¿ Single Lumen Bolt CMPS and Temperature Bolt Kit 030714, 060214, 060314, 070814, 080514, 100714, 120614, 130314, 190614, 310714 IP2.P Licox¿ Double Lumen Bolt CMPS, Temperature and ICP Monitoring Kit 030714, 050614, 060214, 060314, 070814, 080514, 090114, 100714, 130314, 140814, 200214, 200414, 260614, 290814 IT2 Licox¿ Tunneling CMPS and Temperature Monitoring Kit 090114, 100414, 190614, 200214, 260614, 310714
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Nationwide, Foreign
These words name no state.
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Terminated, in FDA’s file.
“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”
openFDA, on “Terminated”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Nov 7, 2014.
- FDA classified it on Dec 18, 2014.
- FDA’s Enforcement Report published it on Dec 24, 2014.
- FDA terminated it on Aug 11, 2015.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Integra LifeSciences Corp., of Plainsboro, NJ.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.