SOLAR 8000 AND TRANSPORT PRO WITH PATIENT DATA MODULE (Marketed As GE Healthcare CARESCAPE Patient Data…

Medical device recall Z-0807-2014. Published in FDA’s Enforcement Report on Jan 29, 2014.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Jan 29, 2014. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Intermittent noise on ECG and/or Respiration waveforms associated with movement of the ECG cable connector on the DASH, PDM, and/or Tram Module. May result in reduced ECG, Respiration and arrhythmia detection performance.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

SOLAR 8000 AND TRANSPORT PRO WITH PATIENT DATA MODULE (Marketed As GE Healthcare CARESCAPE Patient Data Module) Product Usage: The Patient Data Module (PDM) is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes. The Solar 8000i with Patient Data Module/Transport Pro with Patient Data Module is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional medical facility, such as hospital, clinic, surgical center or doctors office. It can be used in multiple areas such as operating room (OR), post anesthesia care unit (PACU), emergency department (ED), chest pain clinic, general intensive care unit (ICU), critical care unit, surgical intensive care unit (SICU), respiratory intensive care unit, coronary care unit (CCU), medical intensive care unit (MICU), pediatric intensive care unit (PICU), or neonatal intensive care unit (NICU). The Patient Data Module (PDM) is intended to provide uninterrupted acquisition of physiologic parameter data on adult, pediatric and neonatal patients during non-transport/bedside and transport patient care episodes. Physiological parameter data acquired by the PDM includes ECG, invasive pressure, non-invasive blood pressure, pulse oximetry, temperature, cardiac output nad respiration. This device acquires, processes and stores information for all aforementioned parameters and transmits this information to a transport or bedside central processing unit for viewing and alarm surveillance purposes. The Transport Pro patient monitor is intended for use as part of a transport monitoring system for intra-healthcare facility transport. When used with the Patient Data Module (PDM) or the TRAM acquisition module, this device is intended to provide uninterrupted monitoring of physiologic parameter data for adult, pediatric, and neonatal patients during transport from one area of the healthcare facility to another. Physiological parameter data includes ECG, invasive pressure, noninvasive blood pressure, pulse oximetry, temperature and respiration. Both the PDM and TRAM acquisition module acquire, process and store information for all aforementioned parameters. The Solar 8000i patient monitoring system is a multi-parameter physiological patient monitoring system intended for use on adult, pediatric and neonatal patients. It provides uninterrupted monitoring of physiological patient data. The Solar 8000i patient monitoring system is capable of monitoring and analyzing the following parameters for all patient populations: Electrocardiogram, invasive pressure, non-invasive blood pressure, pulse, temperature, cardiac output, respiration, pulse oximetry, venous oxygen saturation, transcutaneous pO2 and pCO2, CO2 and respiratory mechanics. The Solar 8000i patient monitoring system is capable of GE Healthcare Page 3 of 6 monitoring the following parameters for adult and pediatric patient populations: anesthetic agent concentrations, O2, impedance cardiography, electroencephalography and bispectral index. The Solar 8000i patient monitoring system interfaces with a variety of third-party peripheral medical devices that support serial and /or analog data outputs. Information from these devices can be displayed, trended and stored in the monitoring system. The solar 8000i patient monitoring system also provides physiological data over the Unity Network.

2516 (2,321 US, 195 OUS), in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Serial Number SA308044521GR SA308186360GA SA310444859GA SA308186380GA SA312091241GA SA308491207GA SA308186401GA SA308186385GA SA308186374GA SA308155957GA SA308186391GA SA308186366GA SA308186410GA SA308094871GA SA308094818GA SA308186445GA…

Every code shown here, 1,988 characters

Serial Number SA308044521GR SA308186360GA SA310444859GA SA308186380GA SA312091241GA SA308491207GA SA308186401GA SA308186385GA SA308186374GA SA308155957GA SA308186391GA SA308186366GA SA308186410GA SA308094871GA SA308094818GA SA308186445GA SA308186381GA SA312479810GA SA308094940GA SA308094803GA SA308094872GA SA308084770GA SA310528992GA SA310506733GA SA310506696GA SA311153223GA SA311142931GA SA312080647GA SA312080894GA SA308186396GA SA308166121GA SA308186397GA SA308094807GA SA308186404GA SA311153169GA SA308186377GA SA308398894GA SA312253398GA SA308094809GA SA309263954GA SA310048616GA SA309527962GA SA310038386GA SA310048604GA SA310048613GA SA310028273GA SA310038462GA SA308317172GA SA311184814GA SA309365237GR SA311142946GA SA307332233GA SA307231341GA SA312069693GA SA309456714GA SA311163858GA SA312049242GA SA311317628GA SA307423146GA SA310259886GA SA309142960GA SA309334557GA SA307393020GA SA307453540GA SA307210827GA SA312018863GA SA307241567GA SA311163574GA SA312039090GA SA312080635GA SA309263938GA SA309304244GR SA309233517GA SA312039205GA SA312080692GA SA308347813GA SA308398654GA SA307504161GA SA312059549GA SA308491112GA SA312059550GA SA309021966GA SA310351543GA SA309193340GA SA308186334GA SA311101421GA SA310444513GA SA311518717GA SA312122048GA SA312029011GA SA309203388GA SA309344857GA SA312028978GA SA312253251GA SA312121695GA SA311508589GR SA312152342GA SA310517769GA SA310517770GA SA311297355GA SA311163926GA SA312029028GA SA308419235GA SA312069661GA SA311458372GA SA307382560GA SA311428153GA SA308511698GA SA308511500GA SA308368183GA SA309223482GA SA310361769GA SA312039183GA SA312079967GA SA308491092GA SA311458333GA SA310393014GA SA310507045GA SA310517560GA SA310331323GA SA311163587GA SA311184495GA SA307291912GA SA310393006GA SA311060085GA SA309112590GA SA310361759GA SA311194946GA SA309112586GA SA309112577GA SA309112584GA SA309112588GA SA309112583GA SA309112582GA SA309112591GA SA309112576GA SA309112569GA SA308166208GA SA308084740GA SA309112585GA SA309112578GA…

FDA’s list runs to 35,290 characters. The first part is shown; the rest is on FDA’s record.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide Distribution: US (nationwide) including DC and PR except SD, VT and WY; and Internationally to: ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, EGYPT, GERMANY, UNITED KINGDOM, IRELAND, IRAQ, ISRAEL, JAMAICA, JAPAN, JORDAN, KOREA, KUWAIT, LEBANON, LITHUANIA, MARITIUS, MALDOVA, NEW ZEALAND, NIGERIA, PAKISTAN, PANAMA, PARAGUAY, PERU, POLAND, QATAR, RUSSIA, SLOVAKIA, SOUTH AFRICA, SWEDEN, SWITZERLAND, TUNISIA, UNITED ARAB EMIRATES, VENEZUELA, CHILE, CANADA, and BERMUDA.

The states and territories these words name

  • District of Columbia
  • Puerto Rico

They leave out South Dakota, Vermont and Wyoming.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Dec 5, 2013.
  • FDA classified it on Jan 22, 2014.
  • FDA’s Enforcement Report published it on Jan 29, 2014.
  • FDA terminated it on Jul 25, 2016.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • GE Healthcare, LLC, of Waukesha, WI.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • FDA’s list of codes runs to 35,290 characters; the first part is shown here, and the rest is on FDA’s record.
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.