Hearth Record

Mindray DS USA, Inc., DPM Central Monitoring System with ambulatory Telepack

Medical device recall Z-0798-2013. Published in FDA’s Enforcement Report on Feb 20, 2013.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Feb 20, 2013. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Mindray has identified a software anomaly that may where, in the presence of an asystole, the Telepack ECG waveform may produce a unique step pattern (instead of a flat line) if Lead 1 or V is being utilized for detecting cardiac arrhythmia.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Mindray DS USA, Inc., DPM Central Monitoring System with ambulatory Telepack. Part numbers: 300BF-PA00002, and 300BF-PA00003 (DPM Central Monitoring System) 115-006908-01, -03, -04 (DPM software upgrade kit). The DPM Central Monitoring System transfers information between DPM Central station and other networked devices.

51 units, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Part No Serial No 300BF-PA00002 EH-17000037 300BF-PA00002 EH-17000032 300BF-PA00002 EH-17000033 300BF-PA00002 EH-25000099 300BF-PA00002 EH-18000043 300BF-PA00003 EH-1A000072 300BF-PA00002 EH-22000088 300BF-PA00003 EH-1A000073 300BF-PA00002…

Every code shown here, 1,348 characters

Part No Serial No 300BF-PA00002 EH-17000037 300BF-PA00002 EH-17000032 300BF-PA00002 EH-17000033 300BF-PA00002 EH-25000099 300BF-PA00002 EH-18000043 300BF-PA00003 EH-1A000072 300BF-PA00002 EH-22000088 300BF-PA00003 EH-1A000073 300BF-PA00002 EH-12000060 300BF-PA00002 EH-17000038 300BF-PA00003 EH-21000077 300BF-PA00002 EH-18000058 300BF-PA00002 EH-23000096 300BF-PA00003 EH-19000059 300BF-PA00003 EH-19000060 300BF-PA00002 EH-18000051 300BF-PA00002 EH-12000054 300BF-PA00002 EH-16000029 300BF-PA00002 EH-11000052 300BF-PA00002 EH-17000039 300BF-PA00003 EH-1A000064 300BF-PA00002 EH-18000050 300BF-PA00003 EH-13000078 300BF-PA00003 EH-13000085 300BF-PA00002 EH-18000053 300BF-PA00002 EH-17000036 300BF-PA00002 EH-17000031 300BF-PA00003 EH-17000040 300BF-PA00002 EH-18000054 300BF-PA00002 EH-1A000069 300BF-PA00002 EH-12000062 300BF-PA00002 EH-21000076 115-006908-04 Software Kit 300BF-PA00003 EH-13000083 300BF-PA00003 EH-13000084 300BF-PA00003 EH-22000087 300BF-PA00002 EH-18000056 300BF-PA00002 EH-17000035 300BF-PA00002 EH-18000048 300BF-PA00002 EH-18000052 300BF-PA00003 EH-21000081 300BF-PA00002 EH-17000030 115-006908-03 Software Kit 115-006908-03 Software Kit 115-006908-03 Software Kit 300BF-PA00003 EH-19000062 300BF-PA00002 EH-18000057 300BF-PA00003 EH-22000083 300BF-PA00002 EH-11000051 300BF-PA00002 EH-18000041 300BF-PA00002 EH-21000082.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Nationwide Distribution.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Nov 28, 2012.
  • FDA classified it on Feb 11, 2013.
  • FDA’s Enforcement Report published it on Feb 20, 2013.
  • FDA terminated it on Aug 15, 2013.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Mindray DS USA, Inc. d.b.a. Mindray North America, of Mahwah, NJ.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.