Centricity Universal Viewer 6.0
Medical device recall Z-0722-2023. Published in FDA’s Enforcement Report on Dec 28, 2022.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Dec 28, 2022. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Ongoing its status, in FDA’s file
- Nationwide FDA’s word, and 22 states named
- 236 units how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
There is the potential that after installation of the IAS tool, some studies can remain hidden without user notification. Access to those hidden studies requires GE Healthcare assistance, which can result in a delay in reading the study.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
Centricity Universal Viewer 6.0. Used to display medical images (Including mammograms) and data from various imaging sources.
236 units, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
Installed Product ID: 100666-1-Centricity Universal Viewer-14194768, 100666-1-Centricity Universal Viewer-20084199, 4405-1-Centricity Universal Viewer-00558662, 4405-1-Centricity Universal Viewer-19508102, 405-1-Centricity Universal…
Every code shown here, 1,986 characters
Installed Product ID: 100666-1-Centricity Universal Viewer-14194768, 100666-1-Centricity Universal Viewer-20084199, 4405-1-Centricity Universal Viewer-00558662, 4405-1-Centricity Universal Viewer-19508102, 405-1-Centricity Universal Viewer-19050922, 4405-1-Centricity Universal Viewer-19050921, 104535-4-Centricity Universal Viewer-02396030, 104535-3-Centricity Universal Viewer-02396026, 104535-3-Centricity Universal Viewer-02396027, 104535-3-Centricity Universal Viewer-02396028, 104535-3-Centricity Universal Viewer-02396029, 104535-4-Centricity Universal Viewer-01806096, 4406-2-Centricity Universal Viewer-00558683, 4446-2-Centricity Universal Viewer-00558643, 4407-1-Centricity Universal Viewer-00558657, 4407-2-Centricity Universal Viewer-02315681, 4406-2-Centricity Universal Viewer-20501355, 4318-1-Centricity Universal Viewer-00558703, 4318-1-Centricity Universal Viewer-20080694, 10227-2-Centricity Universal Viewer-02349137, 4048-1-Centricity Universal Viewer-00558695, 4048-1-Centricity Universal Viewer-14767809, 4048-1-Centricity Universal Viewer-14767808, 4048-1-Centricity Universal Viewer-14767807, 4048-1-Centricity Universal Viewer-14767806, 4048-1-Centricity Universal Viewer-02256595, 9768-1-Centricity Universal Viewer-00558661, 9768-1-Centricity Universal Viewer-01972254, 4353-2-Centricity Universal Viewer-01589220, 13162-8-Centricity Universal Viewer-01452133, 13162-8-Centricity Universal Viewer-14776447, 13162-8-Centricity Universal Viewer-14776449, 4687-1-Centricity Universal Viewer-00558334, 100489-1-Centricity Universal Viewer-01819402, 100489-1-Centricity Universal Viewer-20791988, 104287-2-Centricity Universal Viewer-00555505, 4451-1-Centricity Universal Viewer-00558107, 100366-1-Centricity Universal Viewer-14777071, 100366-1-Centricity Universal Viewer-00558522, 4583-1-Centricity Universal Viewer-01367198, 2337-5-Centricity Universal Viewer-01810490, 2337-5-Centricity Universal Viewer-19670358, 4723-1-Centricity Universal Viewer-02303576,…
FDA’s list runs to 6,399 characters. The first part is shown; the rest is on FDA’s record.
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Nationwide distribution to CA, CO, DE, FL, IA, IL, MA, MI, MN, MO, MS, MT, ND, NJ, NY, OK, OR, SD, TX, WI, WV, WY. International distribution to Austria, Brazil, France, Germany, Ghana, Ireland, Italy, Kuwait, Poland, Saudi Arabia, South Africa, Spain, Switzerland, United Kingdom.
The states and territories these words name
22 states and territories.
Every one of them
- California
- Colorado
- Delaware
- Florida
- Iowa
- Illinois
- Massachusetts
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Dakota
- New Jersey
- New York
- Oklahoma
- Oregon
- South Dakota
- Texas
- Wisconsin
- West Virginia
- Wyoming
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Oct 28, 2022.
- FDA classified it on Dec 16, 2022.
- FDA’s Enforcement Report published it on Dec 28, 2022.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- GE Medical Systems, LLC, of Waukesha, WI.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- FDA’s list of codes runs to 6,399 characters; the first part is shown here, and the rest is on FDA’s record.
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.