Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515) in combination with…
Medical device recall Z-0697-2021. Published in FDA’s Enforcement Report on Dec 30, 2020.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Dec 30, 2020. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Ongoing its status, in FDA’s file
- Nationwide FDA’s word, and 1 state named
- 3 units US how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Software error -Software REF 615 588 versions 2.0 to 3.3, movement of the focusing is triggered by pressing the button of the footswitch it might occur that the movement does not stop when releasing the button, result in potential risk to the patient’s eye
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
Haag Streit Surgical Floor stands FS 2-11 (Ref no. 615H511) and FS 2-15 (Ref no. 615H515) in combination with the operating microscopes HS Hi-R NEO 900 (Ref no. 657 820) and HS Hi-R NEO 900A (Ref no. 657 821), with software REF 615 588 versions 2.0 to 3.3 - Product Usage: use in ophthalmology and in other disciplines with vertical direction of view
3 units US, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
Floor stands FS 2-11, FS 2-15 with the software REF 615 588 in versions 2.0 to 3.3. FS 2-11 Ref. # 615H511 Serial Numbers:From 1001; FS 2-15 Ref. #615H515 Serial Numbers From 1001; HS Hi-R NEO 900 Ref. # 657 Serial Numbers 820 from 101; HS…
Every code shown here, 484 characters
Floor stands FS 2-11, FS 2-15 with the software REF 615 588 in versions 2.0 to 3.3. FS 2-11 Ref. # 615H511 Serial Numbers:From 1001; FS 2-15 Ref. #615H515 Serial Numbers From 1001; HS Hi-R NEO 900 Ref. # 657 Serial Numbers 820 from 101; HS Hi-R NEO 900A Ref # 657 821 Serial Numbers from 101 Unique Device Identifier (UDI) HS Hi-R NEO 900 +EMWM657 820 0/$+serial no. HS Hi-R NEO 900A +EMWM657 821 0/$+serial no. FS 2 11 +EMWM 615H5110/$+serial no. FS 2-15 +EMWM 615H515 0/$+serial no.
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
US Nationwide distribution in the state of OK.
The states and territories these words name
- Oklahoma
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Nov 20, 2020.
- FDA classified it on Dec 23, 2020.
- FDA’s Enforcement Report published it on Dec 30, 2020.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Haag-Streit USA Inc, of Mason, OH.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.