BrightView X, Gamma Camera
Medical device recall Z-0678-2024. Published in FDA’s Enforcement Report on Jan 17, 2024.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Jan 17, 2024. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Ongoing its status, in FDA’s file
- Nationwide FDA’s word for where it went
- 59 US; 59 OUS how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
While using Pre-Programmed Motion during an extrinsic quality assurance scan, a gap is created between the patient support and the detector, which may present a potential extremity entrapment hazard. Risk to patients may include fracture, body part loss of function/debilitation, muscle or ligament sprain or strain, laceration, crush injury, abrasion, or contusion.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
BrightView X, Gamma Camera, Product Code 882478
59 US; 59 OUS, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
No UDI Serial numbers 11000072 11000116 11000007 11000010 11340001 11000027 11000032 4000570 11000041 11000092 11000113 11340003 11000044 11000006 11000033 11000025 11000105 11000016 11000003 11000018 11000065 11000047 11000066 11000101…
Every code shown here, 1,082 characters
No UDI Serial numbers 11000072 11000116 11000007 11000010 11340001 11000027 11000032 4000570 11000041 11000092 11000113 11340003 11000044 11000006 11000033 11000025 11000105 11000016 11000003 11000018 11000065 11000047 11000066 11000101 11000107 11000090 11000029 11000056 11000026 11000046 11000051 11000078 11000114 11000067 11000050 11000084 11000068 11000052 11000057 11000055 11000070 11000088 11000023 11000061 11000060 11000095 11000094 11000102 11000054 11340004 11000091 11000083 11000098 11000063 11000013 11000015 11000062 11000035 11000081 11000009 11000017 11000110 11000004 11000008 11000005 11000020 11000011 11000014 11000012 11000019 11000022 11000031 11000024 11000034 11000036 11000028 11000030 11340002 11000042 11000037 11000080 11000079 11000039 11000040 11000069 11000038 11000045 11000086 11000049 11000043 11000048 11000059 11000058 11000097 11000096 11000073 11000074 11000071 11000064 11000075 11340005 11000077 11000076 11000089 11000087 11000085 11000106 11000100 11000093 11000104 11000099 11000103 11000115 11000112 11000109 11000117 11000108 11000118
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Domestic distribution nationwide. Foreign distribution worldwide.
These words name no state.
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Nov 30, 2023.
- FDA classified it on Jan 5, 2024.
- FDA’s Enforcement Report published it on Jan 17, 2024.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Philips North America, of Cambridge, MA.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.