Hearth Record

Philips Respironics V60 Plus Ventilator

Medical device recall Z-0663-2022. Published in FDA’s Enforcement Report on Mar 9, 2022.

  • FDA Class I
FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Mar 9, 2022. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class I, the most serious of FDA’s three classes.

“Dangerous or defective products that predictably could cause serious health problems or death. Examples include: food found to contain botulinum toxin, food with undeclared allergens, a label mix-up on a lifesaving drug, or a defective artificial heart valve.”

openFDA, in plain words, its enforcement report fields.

“Class I is a situation in which there is a reasonable probability that the use of, or exposure to, a violative product will cause serious adverse health consequences or death.”

FDA’s rule, 21 CFR 7.3(m)(1), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

A limited number of ventilators were assembled with expired adhesive. If the adhesive fails, a bracket may become loose and potentially damage the capacitors, which could cause the ventilator to stop providing ventilation to the patient. This failure may either activate both visual and audible alarms, or it may not sound or display an alarm (a silent shutdown).

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Philips Respironics V60 Plus Ventilator Part Number 1138747

294 systems, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

The Component of the Ventilators that is impacted is: N 1034132 BRACKET, CAPACITOR,MOTOR DRIVER PCB Model V60 Part Number: 1138747 Serial Numbers/UDI Code: MX0016240/ (01)00884838090996(21)MX0016240 MX0016241/…

Every code shown here, 1,973 characters

The Component of the Ventilators that is impacted is: N 1034132 BRACKET, CAPACITOR,MOTOR DRIVER PCB Model V60 Part Number: 1138747 Serial Numbers/UDI Code: MX0016240/ (01)00884838090996(21)MX0016240 MX0016241/ (01)00884838090996(21)MX0016241 MX0016242/ (01)00884838090996(21)MX0016242 MX0016243/ (01)00884838090996(21)MX0016243 MX0016244/ (01)00884838090996(21)MX0016244 MX0016245/ (01)00884838090996(21)MX0016245 MX0016246/ (01)00884838090996(21)MX0016246 MX0016247/ (01)00884838090996(21)MX0016247 MX0016248/ (01)00884838090996(21)MX0016248 MX0016522/ (01)00884838090996(21)MX0016522 MX0016523/ (01)00884838090996(21)MX0016523 MX0016525/ (01)00884838090996(21)MX0016525 MX0016527/ (01)00884838090996(21)MX0016527 MX0016529/ (01)00884838090996(21)MX0016529 MX0016531/ (01)00884838090996(21)MX0016531 MX0016532/ (01)00884838090996(21)MX0016532 MX0016533/ (01)00884838090996(21)MX0016533 MX0016534/ (01)00884838090996(21)MX0016534 MX0016535/ (01)00884838090996(21)MX0016535 MX0016536/ (01)00884838090996(21)MX0016536 MX0016538/ (01)00884838090996(21)MX0016538 MX0016539/ (01)00884838090996(21)MX0016539 MX0016540/ (01)00884838090996(21)MX0016540 MX0016541/ (01)00884838090996(21)MX0016541 MX0016542/ (01)00884838090996(21)MX0016542 MX0016545/ (01)00884838090996(21)MX0016545 MX0016546/ (01)00884838090996(21)MX0016546 MX0016547/ (01)00884838090996(21)MX0016547 MX0016548/ (01)00884838090996(21)MX0016548 MX0016549/ (01)00884838090996(21)MX0016549 MX0016550/ (01)00884838090996(21)MX0016550 MX0016552/ (01)00884838090996(21)MX0016552 MX0016553/ (01)00884838090996(21)MX0016553 MX0016554/ (01)00884838090996(21)MX0016554 MX0016556/ (01)00884838090996(21)MX0016556 MX0016559/ (01)00884838090996(21)MX0016559 MX0016560/ (01)00884838090996(21)MX0016560 MX0016561/ (01)00884838090996(21)MX0016561 MX0016562/ (01)00884838090996(21)MX0016562 MX0016568/ (01)00884838090996(21)MX0016568 MX0016569/ (01)00884838090996(21)MX0016569 MX0016570/ (01)00884838090996(21)MX0016570 MX0016571/…

FDA’s list runs to 12,798 characters. The first part is shown; the rest is on FDA’s record.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

U.S.: AR, CA, FL, GA, IN, LA, MD, MO, NY, PA, SC, TN and TX. O.U.S. N/A

The states and territories these words name

13 states and territories.

Every one of them
  • Arkansas
  • California
  • Florida
  • Georgia
  • Indiana
  • Louisiana
  • Maryland
  • Missouri
  • New York
  • Pennsylvania
  • South Carolina
  • Tennessee
  • Texas

The words say it went outside the United States too.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Jan 24, 2022.
  • FDA classified it on Feb 25, 2022.
  • FDA’s Enforcement Report published it on Mar 9, 2022.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Respironics California, LLC, of Carlsbad, CA.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • FDA’s list of codes runs to 12,798 characters; the first part is shown here, and the rest is on FDA’s record.
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.