Hearth Record

Medtronic Intersept Custom Tubing Pack with or without coating ( Carmeda BioActive Surfact, Carmeda BioActive…

Medical device recall Z-0618-2013. Published in FDA’s Enforcement Report on Jan 9, 2013.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Jan 9, 2013. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Medtronic is initiating an Urgent Medical Device Customer Notification. We have confirmed that certain Medtronic Intersept Tubing Packs contain packaging trays that are susceptible to damage that can compromise product sterility.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Medtronic Intersept Custom Tubing Pack with or without coating ( Carmeda BioActive Surfact, Carmeda BioActive Surfact and Trillium Biosurface,or Balance Biosurface) with the following Model numbers: 0E27R16, 1A30R6, 2493R24, 2493R25, 5D56R5, 5Z93R4, 5Z93R5, 7E64R2, 7J53R2, 7M13R1, 7M14R1, 7N16R, 7P25R1, 7P93R1, 7Q11R1, BB7L63R2, BB7N26R, CB5174R11, CB175R13, CB5Q03R6, CB5Q03R7, CB6C53R6, CB7C15R2, CB7C59R2, CB7C60R2, CB7C74R2, CB7D91R5, CB7E35R2, CB7E38R1, CB7G21R4, CB7L48R1, CB7L72R1, CB7P82R1, HY6U96R2, HY6Y52R2, HY7E87R1, SS7J91R3, SSCB7L48R, SSTL7G78R1,TL5S33R7, TL6VTTR1, TL7B51R1, TL7G20R3, TL7G78R3, TL7R87R1. Sterilized Using Ethylene Oxide, Nonpyrogenic, Assembled in Mexico, Manufacturer Medtronic, Inc, Minneapolis, Mn 55432 Product is used by perfusionists as part of the extracorporeal circuit during cardiopulmonary bypass procedures, and is configured specifically as designated by each customer.

1552, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Pack Model Pack Lot 0E27R16- 11239917 11252800 11264666; 1A30R6- 12132577 12219267; 2493R24- 11187015 11199707 11224309 11227205 11556735; 2493R25- 11646941 11713200 11745242 11813334 11852673 11886412 11990393 12033630 12114699 12178107…

Every code shown here, 1,933 characters

Pack Model Pack Lot 0E27R16- 11239917 11252800 11264666; 1A30R6- 12132577 12219267; 2493R24- 11187015 11199707 11224309 11227205 11556735; 2493R25- 11646941 11713200 11745242 11813334 11852673 11886412 11990393 12033630 12114699 12178107 12191642 12450856; 5D56R5- 11298229 11357085 11465358 11541270 11601193 11654853 11702415 11745232 11766305 11822320 11838721 11852677 11930380 12024708 12166379 12292156; 5Z93R4- 11713132 11736880 11749958 11766283 11808651 11860604 11878689 11911068 11948473 5Z93R5- 12236493; 12411746 7E64R2- 11933798; 7J53R2- 12172725; 7M13R1- 12168157; 7M14R1- 12349772; 7N16R- 12349786; 7P25R1- 12327801 12449801 12457250; 7P93R1- 12349795 12374420; 7Q11R1- 12344028 12456303; BB7L63R2- 12408737; BB7N26R- 12312571; CB5174R11- 11293051 11315344 11334856 11489271 11745162 11886341 11977682 12019852 12062713 12098852 12139621 12146752 12258941 12277502 12326468 12421465 206110946; CB5175R13- 11292866 11334859 11432873 11442937 12182778 12411655; CB5Q03R6- 11713162 11766245 11787583 11923139 11934952; CB5Q03R7- 12114626; CB6C53R6- 12423767; CB7C15R2- 12411647; CB7C59R2- 11731501; 11813316 11838743 11869767 11908794 11982413 12019907 12146796 12151706 12201727 12292133 12423702; CB7C60R2- 11687047 11724030 11843434 11878672; CB7C74R2- 11895167; CB7D91R5- 11961280 12130493 12146882 12191604 12277455 12411687 12423760; CB7E35R2- 11839101; CB7E38R1- 11843886 11885844 11990360 12105025 12133083 12159400; CB7G21R4- 12032413 12033728 12040524 12083359 12360622; CB7L48R1- 12301694; CB7L72R1- 12155258; CB7P82R1- 12440601; HY6U96R2- 206133995; HY6Y52R2- 11826276; HY7E87R1- 11830948 11898103 11923498 12323794 12435409; SS7J91R3- 12175428; SSCB7L48R- 12120109; SSTL7G78R1 -11986264; TL5S33R7- 12093478; 12139612 12172589 12315375; TL6V77R1- 11217931 11260111; TL7B51R1- 11559097 11622968 11664762 11687063 11712834; TL7G20R3- 12258680 12271138 12360554 TL7G78R3- 12067309; TL7R87R1- 206114578 206145379

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide Distribution--USA (nationwide) including the states of AZ, CA, CT, FL, GA, IL IA, MD, MA, MI, MS, NJ, NY, OH, OK, SC, TN, TX and WI. and the countries of CANADA and MALAYSIA.

The states and territories these words name

19 states and territories.

Every one of them
  • Arizona
  • California
  • Connecticut
  • Florida
  • Georgia
  • Iowa
  • Illinois
  • Massachusetts
  • Maryland
  • Michigan
  • Mississippi
  • New Jersey
  • New York
  • Ohio
  • Oklahoma
  • South Carolina
  • Tennessee
  • Texas
  • Wisconsin

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Nov 19, 2012.
  • FDA classified it on Dec 28, 2012.
  • FDA’s Enforcement Report published it on Jan 9, 2013.
  • FDA terminated it on Apr 4, 2013.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Medtronic Inc. Cardiac Rhythm Disease Management, of Saint Paul, MN.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.