-Outer Kit Box (P/N RFIT-LBL-0140) - FilmArray¿ Meningitis/Encephalitis (ME) Panel
Medical device recall Z-0446-2019. Published in FDA’s Enforcement Report on Nov 21, 2018.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Nov 21, 2018. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Terminated its status, in FDA’s file
- Nationwide FDA’s word, and 11 states named
- 1350 pouches how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Firm identified an issue in the manufacturing process which might contribute to elevated rates of false positive results for HSV-1 while using the ME Panel. Only specific kits are affected.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
-Outer Kit Box (P/N RFIT-LBL-0140) - FilmArray¿ Meningitis/Encephalitis (ME) Panel REF RFIT-ASY-0118 30 For FilmArray systems LOT KIT UDI (01)00815381020123 CE 0086 IVD -Fitment Label (P/N RFIT-LBL-0138) which identifies the pouch label - FilmArray¿ ME Panel v1.4 LOT SN CE 0086 IVD Product Usage: The FilmArray Meningitis/Encephalitis (ME) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with FilmArray systems.
1350 pouches, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
ME Panel Kit Lot No. ME Pouch Lot. No. Kit Expiration Date 580818 892018 4/2/2019 584418 896018 4/5/2019 Note - Kits contain a Primer Pill Lot 210318. The Primer Pill is a freeze dried/lyophilized “pellet” that contains all of the primers and buffer used in the PCR1 reaction.
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide Distribution - US Nationwide in the states of: CA DC, FL, IN, KY, MN, MO, NB, NJ, OH, TX and WI. International: Canada, Columbia and Mexico
The states and territories these words name
- California
- District of Columbia
- Florida
- Indiana
- Kentucky
- Minnesota
- Missouri
- New Jersey
- Ohio
- Texas
- Wisconsin
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Terminated, in FDA’s file.
“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”
openFDA, on “Terminated”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Sep 10, 2018.
- FDA classified it on Nov 9, 2018.
- FDA’s Enforcement Report published it on Nov 21, 2018.
- FDA terminated it on Apr 28, 2021.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- BioFire Diagnostics, LLC, of Salt Lake City, UT.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.