Hearth Record

WIRION EMBOLIC PROTECTION SYSTEM, Delivery Catheter and Filter and Retrieval Catheter, 6Fr

Medical device recall Z-0394-2022. Published in FDA’s Enforcement Report on Dec 29, 2021.

  • FDA Class I
FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Dec 29, 2021. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class I, the most serious of FDA’s three classes.

“Dangerous or defective products that predictably could cause serious health problems or death. Examples include: food found to contain botulinum toxin, food with undeclared allergens, a label mix-up on a lifesaving drug, or a defective artificial heart valve.”

openFDA, in plain words, its enforcement report fields.

“Class I is a situation in which there is a reasonable probability that the use of, or exposure to, a violative product will cause serious adverse health consequences or death.”

FDA’s rule, 21 CFR 7.3(m)(1), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Under certain circumstances, the WIRION filter assembly may become difficult to withdraw. Under such circumstances, alternative means may be necessary for withdrawal and the device may be prone to tearing or separating which may result in a risk for embolization.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

WIRION EMBOLIC PROTECTION SYSTEM, Delivery Catheter and Filter and Retrieval Catheter, 6Fr, REF (Model Number) WRN-D6

711 units, in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Lot Numbers: Lot 67876, UDI 011085000049188217221221100000067876 Lot 70309, UDI 011085000049188217230110100000070309 Lot 68706, UDI 011085000049188217221227100000068706 Lot 69543, UDI 011085000049188217230103100000069543 Lot 75758, UDI…

Every code shown here, 1,105 characters

Lot Numbers: Lot 67876, UDI 011085000049188217221221100000067876 Lot 70309, UDI 011085000049188217230110100000070309 Lot 68706, UDI 011085000049188217221227100000068706 Lot 69543, UDI 011085000049188217230103100000069543 Lot 75758, UDI 011085000049188217230203100000075758 Lot 77278, UDI 011085000049188217230210100000077278 Lot 79994, UDI 011085000049188217230223100000079994 Lot 81252, UDI 011085000049188217230301100000081252 Lot 81986, UDI 011085000049188217230307100000081986 Lot 83080, UDI 011085000049188217230313100000083080 Lot 81998, UDI 011085000049188217230320100000081998 Lot 84796, UDI 011085000049188217230327100000084796 Lot 87909, UDI 011085000049188217230411100000087909 Lot 89035, UDI 011085000049188217230418100000089035 Lot 86622, UDI 011085000049188217230404100000086622 Lot 90096, UDI 011085000049188217230424100000090096 Lot 93949, UDI 011085000049188217230515100000093949 Lot 91603, UDI 011085000049188217230501100000091603 Lot 94895, UDI 011085000049188217230521100000094895 Lot 91939, UDI 011085000049188217230507100000091939 Lot 101930, UDI 011085000049188217230717100000101930

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

United States- AK, AL, AR, AZ, CA, DC, FL, GA, GU, HI, IA, IL, IN, KY, LA, MA, MI, MN, MO, MS, ND, NE, NH, NJ, NM, NV, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV OUS - Netherlands

The states and territories these words name

38 states and territories.

Every one of them
  • Alaska
  • Alabama
  • Arkansas
  • Arizona
  • California
  • District of Columbia
  • Florida
  • Georgia
  • Guam
  • Hawaii
  • Iowa
  • Illinois
  • Indiana
  • Kentucky
  • Louisiana
  • Massachusetts
  • Michigan
  • Minnesota
  • Missouri
  • Mississippi
  • North Dakota
  • Nebraska
  • New Hampshire
  • New Jersey
  • New Mexico
  • Nevada
  • New York
  • Ohio
  • Pennsylvania
  • South Carolina
  • South Dakota
  • Tennessee
  • Texas
  • Utah
  • Virginia
  • Washington
  • Wisconsin
  • West Virginia

The words say it went outside the United States too.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Nov 22, 2021.
  • FDA classified it on Dec 21, 2021.
  • FDA’s Enforcement Report published it on Dec 29, 2021.
  • FDA terminated it on Aug 31, 2023.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Cardiovascular Systems Inc, of Saint Paul, MN.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.