Hearth Record

Mindray A3 Anesthesia Delivery System

Medical device recall Z-0290-2013. Published in FDA’s Enforcement Report on Nov 28, 2012.

  • FDA Class I
FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Nov 28, 2012. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class I, the most serious of FDA’s three classes.

“Dangerous or defective products that predictably could cause serious health problems or death. Examples include: food found to contain botulinum toxin, food with undeclared allergens, a label mix-up on a lifesaving drug, or a defective artificial heart valve.”

openFDA, in plain words, its enforcement report fields.

“Class I is a situation in which there is a reasonable probability that the use of, or exposure to, a violative product will cause serious adverse health consequences or death.”

FDA’s rule, 21 CFR 7.3(m)(1), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Mindray has identified the potential for a system leak resulting an issue related to the carbon dioxide absorber canister gasket on A3/A5 Anesthesia Delivery Systems. A potential system leak resulting from the canister gasket will likely present itself during the Automatic Circuit Leak and Compliance Test performed at startup and the Manual Leak Test recommended to be performed before each case.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Mindray A3 Anesthesia Delivery System Manufactured in China for Mindray DS USA Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430 This is a device used to administer to a patient continuously or intermittently a general inhalation anesthetic and to maintain a patient’s ventilation.

69 units (61 US - 8 Foreign), in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

0633F-01000-00 Various serial numbers 0633F-01000-00 FA-1A000033 0633F-01000-00 FA-1A000023 0633F-01000-00 FA-21000072 0633F-01000-00 FA-1A000029 0633F-01000-00 FA-21000085 0633F-01000-00 FA-19000011 0633F-01000-00 FA-19000008…

Every code shown here, 1,684 characters

0633F-01000-00 Various serial numbers 0633F-01000-00 FA-1A000033 0633F-01000-00 FA-1A000023 0633F-01000-00 FA-21000072 0633F-01000-00 FA-1A000029 0633F-01000-00 FA-21000085 0633F-01000-00 FA-19000011 0633F-01000-00 FA-19000008 0633F-01000-00 FA-19000009 0633F-01000-00 FA-1A000027 0633F-01000-00 FA-1A000047 0633F-01000-00 FA-1A000055 0633F-01000-00 FA-1A000041 0633F-01000-00 FA-1A000053 0633F-01000-00 FA-1B000056 0633F-01000-00 FA-21000080 0633F-01000-00 FA-19000007 0633F-01000-00 FA-19000013 0633F-01000-00 FA-21000066 0633F-01000-00 FA-21000078 0633F-01000-00 FA-19000012 0633F-01000-00 FA-19000017 0633F-01000-00 FA-1A000021 0633F-01000-00 FA-1A000028 0633F-01000-00 FA-1A000038 0633F-01000-00 FA-1A000045 0633F-01000-00 FA-1A000035 0633F-01000-00 FA-1A000037 0633F-01000-00 FA-1A000026 0633F-01000-00 FA-1A000030 0633F-01000-00 FA-1A000048 0633F-01000-00 FA-1A000040 0633F-01000-00 FA-1B000059 0633F-01000-00 FA-1A000034 0633F-01000-00 FA-1A000049 0633F-01000-00 FA-21000062 0633F-01000-00 FA-21000064 0633F-01000-00 FA-21000074 0633F-01000-00 FA-21000082 0633F-01000-00 FA-21000090 0633F-01000-00 FA-1A000022 0633F-01000-00 FA-1A000051 0633F-01000-00 FA-1A000052 0633F-01000-00 FA-1A000042 0633F-01000-00 FA-1A000050 0633F-01000-00 FA-1B000057 0633F-01000-00 FA-1B000061 0633F-01000-00 FA-1A000025 0633F-01000-00 FA-1A000031 0633F-01000-00 FA-1A000036 0633F-01000-00 FA-1A000044 0633F-01000-00 FA-21000076 0633F-01000-00 FA-21000077 0633F-01000-00 FA-21000088 0633F-01000-00 FA-1A000043 0633F-01000-00 FA-1A000032 0633F-01000-00 FA-1A000046 0633F-01000-00 FA-1B000058 0633F-01000-00 FA-1B000060 0633F-01000-00 FA-19000010 0633F-01000-00 FA-19000015 0633F-01000-00 FA-21000086

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Australia, Chile, Colombia, and Venezuela.

The states and territories these words name

  • Puerto Rico

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Terminated, in FDA’s file.

“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”

openFDA, on “Terminated”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Aug 13, 2012.
  • FDA classified it on Nov 16, 2012.
  • FDA’s Enforcement Report published it on Nov 28, 2012.
  • FDA terminated it on Jun 4, 2013.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Mindray DS USA, Inc. d.b.a. Mindray North America, of Mahwah, NJ.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.