The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and…
Medical device recall Z-0286-2014. Published in FDA’s Enforcement Report on Nov 20, 2013.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Nov 20, 2013. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Terminated its status, in FDA’s file
- Nationwide FDA’s word, and 52 states named
- 214 units how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Plum A+ infusers have the potential for the distal (occlusion) press sensor pin to break. The distal pressure sensor pin is a part of the overall subsystem that measures the pressure within the distal line of the administration set and indicates the presence of a full or partial distal occlusion. A broken distal pin can only be detected via a visual inspection of the distal pressure pin. A broken
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
The Plum A+3 Infusion System is a cassette based multifunction infusion system that allows two lines in and one line out and can be used for standard, piggyback, or concurrent delivery. The Plum A+3 Infusion System is designed to deliver parenteral, enteral, or epidural infusions over a broad range of infusion rates from multiple fluid container types. The Plum A+3 Infusion System includes an infuser, disposable IV sets (hereafter called a set), optional accessories, and this operator’s manual. The Plum A+ Infusion System with Hospira MedNet Software is indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.
214 units, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
1) List Number 12348; Serial Numbers: 13740156 , 13740256 , 13740266 , 13740351 , 13740380 , 13740451 , 13740508 , 13740612 , 13740632 , 13740731 , 13740732 , 13740733 , 13740734 , 13740735 , 13740736 , 13740737 , 13740738 , 13740739 ,…
Every code shown here, 1,996 characters
1) List Number 12348; Serial Numbers: 13740156 , 13740256 , 13740266 , 13740351 , 13740380 , 13740451 , 13740508 , 13740612 , 13740632 , 13740731 , 13740732 , 13740733 , 13740734 , 13740735 , 13740736 , 13740737 , 13740738 , 13740739 , 13740740 , 13740741 , 13740742 , 13740743 , 13740744 , 13740745 , 13740746 , 13740747 , 13740748 , 13740749 , 13740750 , 13740751 , 13740752 , 13740753 , 13740754 , 13740755 , 13740756 , 13740757 , 13740758 , 13740759 , 13740760 , 13740761 , 13740762 , 13740763 , 13740764 , 13740765 , 13740766 , 13740767 , 13740768 , 13740769 , 13740770 , 13740771 , 13740772 , 13740773 , 13740774 , 13740775 , 13740776 , 13740777 , 13740778 , 13740779 , 13740780 , 13740781 , 13740782 , 13740783 , 13740784 , 13740785 , 13740786 , 13740787 , 13740788 , 13740789 , 13740790 , 13740791 , 13740792 , 13740793 , 13740794 , 13740795 , 13740796 , 13740797 , 13740798 , 13740799 , 13740800 , 13740801 , 13740802 , 13740803 , 13740804 , 13740805 , 13740806 , 13740807 , 13740808 , 13740809 , 13740810 , 13740811 , 13740812 , 13740813 , 13740814 , 13740815 , 13740816 , 13740818 , 13740819 , 13740820 , 13740821 , 13740822 , 13740823 , 13740824 , 13740825 , 13740826 , 13740827 , 13740828 , 13740829 , 13740830 , 13740831 , 13740832 , 13740833 , 13740834 , 13740835 , 13740836 , 13740837 , 13740838 , 13740839 , 13740840 , 13740841 , 13740842 , 13740843 , 13740844 , 13740845 , 13740846 , 13740847 , 13740848 , 13740849 , 13740850 , 13740851 , 13740852 , 13740853 , 13740854 , 13740855 , 13740856 , 13740857 , 13740858 , 13740859 , 13740860 , 13740861 , 13740862 , 13740863 , 13740864 , 13740865 , 13740866 , 13740867 , 13740868 , 13740869 , 13740870 , 13740871 , 13740872 , 13740873 , 13740874 , 13740875 , 13740876 , 13740877 , 13740878 , 13740879 , 13740880 , 13740881 , 13740882 , 13740883 , 13740884 , 13740885 , 13740886 , 13740887 , 13740888 , 13740889 , 13740890 , 13740891 , 13740892 , 13740893 , 13740894 , 13740895 , 13740896 , 13740897 , 13740898 , 13740899 , 13740900 ,…
FDA’s list runs to 2,389 characters. The first part is shown; the rest is on FDA’s record.
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide Distribution - USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, District of Columbia, Puerto Rico and U.S. Virgin Islands; and countries of: Costa Rica and Dominican Republic.
The states and territories these words name
52 states and territories.
Every one of them
- Alaska
- Alabama
- Arkansas
- Arizona
- California
- Colorado
- Connecticut
- District of Columbia
- Delaware
- Florida
- Georgia
- Iowa
- Idaho
- Illinois
- Indiana
- Kansas
- Kentucky
- Louisiana
- Massachusetts
- Maryland
- Maine
- Michigan
- Minnesota
- Missouri
- Mississippi
- Montana
- North Carolina
- North Dakota
- Nebraska
- New Hampshire
- New Jersey
- New Mexico
- Nevada
- New York
- Ohio
- Oklahoma
- Oregon
- Pennsylvania
- Puerto Rico
- Rhode Island
- South Carolina
- South Dakota
- Tennessee
- Texas
- Utah
- Virginia
- US Virgin Islands
- Vermont
- Washington
- Wisconsin
- West Virginia
- Wyoming
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Terminated, in FDA’s file.
“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”
openFDA, on “Terminated”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Feb 1, 2013.
- FDA classified it on Nov 14, 2013.
- FDA’s Enforcement Report published it on Nov 20, 2013.
- FDA terminated it on Dec 6, 2016.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- Hospira Inc., of Lake Forest, IL.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- FDA’s list of codes runs to 2,389 characters; the first part is shown here, and the rest is on FDA’s record.
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.