Hearth Record

Hamilton-C6

Medical device recall Z-0267-2023. Published in FDA’s Enforcement Report on Nov 30, 2022.

FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Nov 30, 2022. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class II, the middle of FDA’s three classes.

“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”

openFDA, in plain words, its enforcement report fields.

“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”

FDA’s rule, 21 CFR 7.3(m)(2), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

Due to a malfunction related to the backlight on the ventilator screen. The screen goes black for approximately 2-3 seconds, but it brightens up again when the backlight switches on again.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

Hamilton-C6

8,229 total systems worldwide (U.S. 443 systems), in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Model Number: 160021 UDI-DI Code: 07630002808590 All Serial Numbers U.S. Serial Numbers 6093 6159 6274 6276 6309 6320 6327 6464 6512 6545 6674 6789 7136 7173 7191 7273 7279 7422 7432 7569 7572 7587 7644 7676 7782 7783 7784 7786 7788 7789…

Every code shown here, 1,999 characters

Model Number: 160021 UDI-DI Code: 07630002808590 All Serial Numbers U.S. Serial Numbers 6093 6159 6274 6276 6309 6320 6327 6464 6512 6545 6674 6789 7136 7173 7191 7273 7279 7422 7432 7569 7572 7587 7644 7676 7782 7783 7784 7786 7788 7789 7790 7792 7793 7794 7795 7796 7797 7798 7799 7801 7806 7807 7812 7813 7837 7901 7923 7969 8182 8343 8394 8395 8399 8420 8421 8422 8423 8424 8497 8519 8593 8661 8668 8670 8673 8675 8676 8678 8707 8709 8712 9028 9029 9032 9033 9042 9043 9044 9046 14784 4833 5663 5679 5681 5939 6075 6149 6157 6212 6219 6231 6280 6349 6350 6416 6480 6505 6516 6537 6562 6566 6579 6641 6658 6671 6723 6774 6790 6829 6858 6868 6871 6896 6907 6925 6932 6937 6944 6952 6953 6965 6995 7004 7033 7066 7113 7115 7117 7122 7130 7138 7150 7156 7165 7174 7177 7196 7201 7252 7265 7307 7349 7358 7382 7410 7420 7581 7586 7608 7614 7656 7666 7695 7719 7740 7754 7772 7777 7787 7802 7808 7809 7819 7821 7822 7826 7827 7831 7836 7845 7861 7884 7887 7888 7893 7894 7898 7938 7966 7980 7986 7993 7994 8003 8005 8049 8052 8063 8070 8082 8087 8097 8099 8102 8115 8121 8122 8148 8151 8178 8235 8249 8255 8256 8258 8261 8281 8297 8306 8310 8311 8313 8322 8325 8344 8346 8348 8356 8371 8401 8406 8409 8413 8477 8490 8517 8535 8568 8590 8594 8597 8599 8601 8603 8604 8616 8620 8658 8702 8706 8713 8714 8715 8717 8718 8719 8720 8722 8723 8725 8726 8727 8844 8845 8846 8847 8849 8850 8866 9036 9050 9051 9054 9062 9063 9064 9067 9069 9071 9073 9075 9076 9077 9080 9081 9091 9098 9099 9100 9101 9102 9103 9105 9106 9107 9108 9109 9110 9118 9120 9121 9124 9125 9126 9127 9128 9129 9130 9131 9132 9133 9134 9137 9138 9192 9224 9225 9227 9228 9229 9230 9231 9232 9233 9235 9236 9237 9239 9241 9242 9243 9244 9248 9257 9258 9259 9260 9290 9291 9292 9293 9294 9297 9298 9299 9300 9301 9303 9324 9325 9329 9337 9348 9352 9353 9354 9422 9423 9424 9425 9426 9427 9428 9429 9430 9432 9433 9452 9458 9459 9460 9462 9463 9464 9465 9466 9487 9488 9490 9491 9508 9527 9535 9543 9545 9553 9575 9576 9577 9578 9579 9582…

FDA’s list runs to 2,003 characters. The first part is shown; the rest is on FDA’s record.

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

U.S. Nationwide distribution including in the states of CA, FL, GA, ID, IN, KS, KY, MA, MI, MO, NV, NM, NY, NC, OH, PA, Puerto Rico, SC, TX, VA, WA, WV, WI, and WY.

The states and territories these words name

24 states and territories.

Every one of them
  • California
  • Florida
  • Georgia
  • Idaho
  • Indiana
  • Kansas
  • Kentucky
  • Massachusetts
  • Michigan
  • Missouri
  • North Carolina
  • New Mexico
  • Nevada
  • New York
  • Ohio
  • Pennsylvania
  • Puerto Rico
  • South Carolina
  • Texas
  • Virginia
  • Washington
  • Wisconsin
  • West Virginia
  • Wyoming

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Sep 22, 2022.
  • FDA classified it on Nov 18, 2022.
  • FDA’s Enforcement Report published it on Nov 30, 2022.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • Hamilton Medical AG, of Domat/Ems, Switzerland.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • FDA’s list of codes runs to 2,003 characters; the first part is shown here, and the rest is on FDA’s record.
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.