Quantum 2000 Electrosurgical Generators Cooper Surgical
Medical device recall Z-0261-2022. Published in FDA’s Enforcement Report on Nov 24, 2021.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Nov 24, 2021. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Ongoing its status, in FDA’s file
- Nationwide FDA’s word for where it went
- 193 Units how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
Coagulate function may not operate with the use of the Hand Switch, failure of the device to coagulate tissue may lead to a prolonged procedure requiring medical intervention to control any intraoperative bleeding and complete the case.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
Quantum 2000 Electrosurgical Generators Cooper Surgical Part Number: 909075
193 Units, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
Serial Numbers: 0112010 0209001 0214002 0309031 0313007BFIX 0411002 0609010FIX 0710001 0711008 0713011C 0809009 0813032CFIX 091402 1013009C 1013015C 1109017 1214005 2015090019 2015110003 2016030001 2016040012 2016060016 2016090003…
Every code shown here, 1,589 characters
Serial Numbers: 0112010 0209001 0214002 0309031 0313007BFIX 0411002 0609010FIX 0710001 0711008 0713011C 0809009 0813032CFIX 091402 1013009C 1013015C 1109017 1214005 2015090019 2015110003 2016030001 2016040012 2016060016 2016090003 2016110002 2017010005 2017060010 2017080010 QTMAA-4-095 QTM-AB-10-149 QTM-M-4-102FIX W110954 W351106 W60413 0114011 0212004 0214007 0312021 0313010B 0413003B 0612001B 0710003 0711008FIX 0713018C 0811017 0813033C 1011010 1013010C 1014004 1111002 2015040009 2015090021 2015110004 2016030002 2016040015 2016060020 2016090004 2016110003 2017020004 2017060013 AFZFIX QTMAA4101 QTM-AB-12-164 QTMY-11-0201 W204996 W381180 W605041 0114013 0212008 0214008 0312024 0313011B 0511002 0612003B 0710008 0711009 0714007 0811021 0813034C 1011010FIX 1013011C 1109008 1213005C 2015060003 2015100004 2015110009 2016040001 2016040018 2016070003 2016090007 2016110004 2017030002 2017080002 ATM-AA-4-106 QTMAA-4-102 QTM-AB-12-165 QTMY-12-0208 W2A1022 W381192 W708311FIX 0115006 0212010 0215005 0312033 0409005 0511003 0612010B 0710008FIX 0712009B 0809004 0811021FIX 0814011FIX 1012006B 1013012C 1109014 1213006C 2015090012 2015100006FIX 2015110010 2016040002 2016060012 2016070007 2016090014 2016120002 2017030005 2017080003 GHILL QTMAA6116 QTMAB3067 QTMY70139 W2B1036 W604009 0115007 0214001 0215007 0313001B 0409006 0511003FIX 0612016B 0711006 0713001C 0809007 0813029C 0814012 1013006C 1013014C 1109015 1214003 2015090015 2015110001 2016020006 2016040007 2016060013 2016080002 2016110001 2017010002 2017030009 2017080006 IUY80207 QTMAA-6-116 QTM-AB-9135 W107893 W2C1054 W604013
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
US Nationwide Distribution Foreign: ARMENIA, BOLIVIA, CAMBODIA, CANADA, CHILE, CHINA, COSTA RICA, DOMINICAN REPUBLIC, ECUADOR, ENGLAND, GERMANY, HONG KONG, INDIA, INDONESIA, IRAN, ITALY, JORDAN, KENYA, MALAYSIA, MAURITIUS, MEXICO, NEW ZEALAND, NICARAGUA, NIGERIA, PHILIPPINES, PUERTO RICO, QATAR, RWANDA, SINGAPORE, SOUTH AFRICA, SRI LANKA, TAIWAN, TANZANIA, THAILAND, TRINIDAD, TURKEY, UGANDA, UNITED ARAB EMIRATES, VIETNAM
These words name no state.
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Ongoing, in FDA’s file.
“A recall which is currently in progress.”
openFDA, on “On-Going”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Sep 29, 2021.
- FDA classified it on Nov 18, 2021.
- FDA’s Enforcement Report published it on Nov 24, 2021.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- CooperSurgical, Inc., of Trumbull, CT.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.