GE Healthcare, SIGNA PET/MR 3.0T
Medical device recall Z-0260-2016. Published in FDA’s Enforcement Report on Nov 18, 2015.
No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.
The short answer
Published in FDA’s Enforcement Report on Nov 18, 2015. Copied from openFDA on Oct 20, 2026.
- Class II FDA’s class
- Terminated its status, in FDA’s file
- Nationwide FDA’s word, and 3 states named
- 17 how much, in FDA’s words
Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.
How serious is it, by FDA’s class?
Class II, the middle of FDA’s three classes.
“Products that might cause a temporary health problem, or pose only a slight threat of a serious nature. Example: a drug that is under-strength but that is not used to treat life-threatening situations.”
openFDA, in plain words, its enforcement report fields.
“Class II is a situation in which use of, or exposure to, a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.”
FDA’s rule, 21 CFR 7.3(m)(2), eCFR.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
What’s wrong with it?
When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Which ones are recalled?
GE Healthcare, SIGNA PET/MR 3.0T. Product Usage: The SIGNA PET/MR system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high-resolution physiologic and anatomic information, acquired simultaneously and isocentrically.
17, in FDA’s words.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Codes to check against yours
00000000UA0064 415723SHMRW 00000000UA0217 415CBMRP N/A 608262PETMR1 MAYOPETMR01 UABPETMR SID TBD SID TBD 262574PETMR1 310974MR1 MRUA0459 309452MR1 082427HWBAY1 M40345211 N/A SID TBD SID TBD SID TBD 00001309FMM076 C001MR01 00000309668MR2 AC001MR02 00000307417MR6 120810MR03
Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.
Where did it go?
FDA’s words for the “general area of initial distribution”: where the firm first shipped it.
Worldwide Distribution - US: Nationwide including DC, PR and GU and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BERMUDA, BOSNIA HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GERMANY, GREECE, HONG Kong, HUNGARY, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, KAZAKHSTAN, REPUBLIC OF KOREA, KUWAIT, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PERU, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, and VENEZUELA.
The states and territories these words name
- District of Columbia
- Guam
- Puerto Rico
What FDA means by nationwide
“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”
openFDA, its enforcement report fields.
The words say it went outside the United States too.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
Is it still going on?
Terminated, in FDA’s file.
“FDA has determined that all reasonable efforts have been made to remove or correct the violative product in accordance with the recall strategy, and proper disposition has been made according to the degree of hazard.”
openFDA, on “Terminated”, its enforcement report fields.
What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.
- The firm began the recall on Aug 31, 2015.
- FDA classified it on Nov 9, 2015.
- FDA’s Enforcement Report published it on Nov 18, 2015.
- FDA terminated it on Sep 7, 2018.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
Who is recalling it?
- GE Healthcare, of Waukesha, WI.
- Who began it, in FDA’s words: Voluntary: Firm initiated.
- How the firm first told its customers: Letter.
FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.
What this record can’t tell you
- Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
- Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
- How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
- Whether anyone was hurt or fell ill: the report doesn’t say.
- How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
- A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Hearth Record wrote these lines; the quotations are FDA’s.
Sources
-
U.S. Food and Drug Administration, Device recall enforcement reports.
Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.
Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.