Hearth Record

ROSA One 3.1 Brain application The device is intended for the spatial positioning and orientation of…

Medical device recall Z-0118-2022. Published in FDA’s Enforcement Report on Nov 3, 2021.

  • FDA Class I
FDA’s own record, on accessdata.fda.gov

No Watch here, and no alerts: openFDA asks that its enforcement data “should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall” (openFDA). For FDA’s own recall emails, subscribe at FDA. FDA’s own record is on accessdata.fda.gov.

The short answer

Published in FDA’s Enforcement Report on Nov 3, 2021. Copied from openFDA on Oct 20, 2026.

Every line below is in FDA’s own words, with its source beside it. FDA’s terms are explained in FDA’s own words.

How serious is it, by FDA’s class?

Class I, the most serious of FDA’s three classes.

“Dangerous or defective products that predictably could cause serious health problems or death. Examples include: food found to contain botulinum toxin, food with undeclared allergens, a label mix-up on a lifesaving drug, or a defective artificial heart valve.”

openFDA, in plain words, its enforcement report fields.

“Class I is a situation in which there is a reasonable probability that the use of, or exposure to, a violative product will cause serious adverse health consequences or death.”

FDA’s rule, 21 CFR 7.3(m)(1), eCFR.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

What’s wrong with it?

The firm has become aware of a software anomaly affecting ROSA One 3.1 Brain application which led to the inaccurate placement of an electrode during surgery. The firm has received 3 global complaints related to the issue. An incorrect trajectory could result in serious injury or death if undetected during surgery.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Which ones are recalled?

ROSA One 3.1 Brain application The device is intended for the spatial positioning and orientation of instruments holders or tool guides to be used by trained surgeons to guide standard surgical instruments during brain and spine surgeries.

119 (US); 55 (OUS), in FDA’s words.

Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Codes to check against yours

Serial Numbers: US: BS16904 BS16905 BS16906 BS16909 BS16910 BS16911 BS16912 BS16913 BS16914 BS16915 BS16916 BS16917 BS16918 BS16921 BS16923 BS16925 BS17930 BS17934 BS17935 BS17936 BS17937 BS18900 BS18901 BS18924 BS18931 BS18938 BS18939…

Every code shown here, 1,416 characters

Serial Numbers: US: BS16904 BS16905 BS16906 BS16909 BS16910 BS16911 BS16912 BS16913 BS16914 BS16915 BS16916 BS16917 BS16918 BS16921 BS16923 BS16925 BS17930 BS17934 BS17935 BS17936 BS17937 BS18900 BS18901 BS18924 BS18931 BS18938 BS18939 BS18940 BS18942 BS18943 BS18945 BS18946 BS18947 BS18948 BS18949 BS18950 BS18951 BS18952 BS18953 BS18954 BS18955 BS18956 BS18957 BS18958 BS18959 BS18960 BS18961 BS18962 BS18963 BS18964 BS18965 BS18966 BS18967 BS18968 BS18969 BS18970 BS18971 BS18972 BS18973 BS18974 BS18975 BS18976 BS18977 BS18979 BS18980 BS18981 BS18982 BS18983 BS18984 BS18985 BS18986 BS18987 BS18988 BS18989 BS18990 BS18991 BS18992 BS18993 BS18994 BS18995 BS18996 BS18997 BS18998 BS18999 BS19037 BS19038 BS19039 BS19040 BS19041 BS19043 BS19047 BS19049 BS19050 BS19052 BS19053 BS19054 BS19057 BS19058 BS19059 BS19060 BS19061 BS19065 BS20066 BS20067 BS20072 BS20073 BS20079 BS20080 BS20086 BS20092 BS20100 BS20103 BS20107 BS21112 BS21115 BS21121 BS21129 BS21137 BS21138 OUS: BS15902 BS15903 BS16007 BS16919 BS16928 BS16929 BS17008 BS17010 BS17013 BS17021 BS17022 BS17023 BS18024 BS18026 BS18027 BS18028 BS18030 BS18031 BS19032 BS19033 BS19034 BS19035 BS19036 BS19042 BS19051 BS19056 BS19062 BS19064 BS20068 BS20069 BS20070 BS20071 BS20074 BS20075 BS20076 BS20077 BS20078 BS20082 BS20084 BS20085 BS20090 BS20095 BS20098 BS20099 BS20102 BS20105 BS20106 BS21108 BS21109 BS21114 BS21124 BS21128 BS21132 BS21133 BS21134

Source:

Copied exactly as FDA lists them: lot and serial numbers, dates and other codes on the product or its label.

Where did it go?

FDA’s words for the “general area of initial distribution”: where the firm first shipped it.

US Nationwide distribution and International Distribution to countries of: Australia, Taiwan, China, Thailand, Japan, India, South Korea, and the Netherlands.

These words name no state.

What FDA means by nationwide

“The term “nationwide” is defined to mean the fifty states or a significant portion. Note that subsequent distribution by the consignees to other parties may not be included.”

openFDA, its enforcement report fields.

The words say it went outside the United States too.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

Is it still going on?

Ongoing, in FDA’s file.

“A recall which is currently in progress.”

openFDA, on “On-Going”, its enforcement report fields.

Source:

What these words mean is quoted from openFDA’s field reference and 21 CFR 7.3.

  • The firm began the recall on Sep 22, 2021.
  • FDA classified it on Oct 25, 2021.
  • FDA’s Enforcement Report published it on Nov 3, 2021.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

Who is recalling it?

  • MEDTECH SAS, of Montpellier, France.
  • Who began it, in FDA’s words: Voluntary: Firm initiated.
  • How the firm first told its customers: Letter.
Source:

FDA, Device recall enforcement reports. Released Oct 14, 2026. Updated weekly.

What this record can’t tell you

  • Whether your own one is affected: compare the codes, dates and numbers above with yours. When unsure, ask the firm or the store.
  • Where it went after the firm’s first shipments. FDA’s words cover the firm’s initial distribution, and, in FDA’s words, “subsequent distribution by the consignees to other parties may not be included.”
  • How many have been returned, corrected or destroyed: FDA’s report doesn’t say.
  • Whether anyone was hurt or fell ill: the report doesn’t say.
  • How current it is: openFDA publishes FDA’s Enforcement Report weekly, after FDA does, and a firm can announce a recall weeks before it appears there. openFDA’s own caution: “This data should not be used as a method to collect data to issue alerts to the public, nor should it be used to track the lifecycle of a recall.”
  • A product with no recall here isn’t safe by that: FDA’s report lists the recalls FDA classified, and when.
Source:

Hearth Record wrote these lines; the quotations are FDA’s.

Sources

  • U.S. Food and Drug Administration, Device recall enforcement reports.

    Released Oct 14, 2026 and checked Oct 20, 2026. Updated weekly. Public domain (CC0 1.0, openFDA). FDA asks that nothing suggest it endorses this site.

Source:

Recalls of food, medicines, medical devices, pet food and animal drugs are from the U.S. Food and Drug Administration (FDA). FDA does not endorse Hearth Record. openFDA: “Do not rely on openFDA to make decisions regarding medical care.” Copied on Oct 20, 2026.

How to read a record

Four things every record does, shown on an example record made up for this guide.

An example record, made up for this guide

An example nursing home

Made up for this guide: no real place is shown

Not fully inspected in over two years (CMS flag)

11 health citations at the latest inspection cycle

The state average is 14.3.

Source:

CMS, State and US Averages and Health Deficiencies. Released Oct 28, 2026. Updated monthly.

From the homeNone yet. Replies and paid listings aren’t open, and when they open they will sit here, below the record.

1 of 4

Every fact names its source.

Who published it, when, and how often it changes sit beside each fact, with a link to the agency’s own file.

Sign in

A free account keeps your watch list, shortlists, comparisons and saved searches on every device, and can email you when something you watch changes. Browsing never needs one.